| Literature DB >> 21343216 |
Kenneth St John1, Minakshi Gupta.
Abstract
Acetabular hip joint components manufactured from gamma-sterilized ultra high molecular weight polyethylene (UHMWPE), gamma cross-linked UHMWPE, or polycarbonate-urethane (PCU) polymers were evaluated in a hip joint simulator, using cobalt alloy femoral components, for at least 5 million cycles. The volume of material losses due to wear was calculated for each type of sample, based upon mass loss measurements, every 500,000 cycles. The loss of material for the conventional UHMWPE was much higher than for the cross-linked UHMWPE, showing about a 70% reduction in wear due to cross-linking. The material loss for the PCU samples appears to have been at least 24% lower than for the cross-linked UHMWPE. Based upon these results, the PCU material seems to have potential for use as an alternative bearing material to UHMWPE for total hip replacement surgeries.Entities:
Mesh:
Substances:
Year: 2011 PMID: 21343216 DOI: 10.1177/0885328210394471
Source DB: PubMed Journal: J Biomater Appl ISSN: 0885-3282 Impact factor: 2.646