MRI-guided prostate radiation therapy planning: Investigation of dosimetric accuracy of MRI-based dose planning.

Jonathan Lambert1, Peter B Greer, Fred Menk, Jackie Patterson, Joel Parker, Kara Dahl, Sanjiv Gupta, Anne Capp, Chris Wratten, Colin Tang, Mahesh Kumar, Jason Dowling, Sarah Hauville, Cynthia Hughes, Kristen Fisher, Peter Lau, James W Denham, Olivier Salvado.   

Abstract

BACKGROUND AND
PURPOSE: Dose planning requires a CT scan which provides the electron density distribution for dose calculation. MR provides superior soft tissue contrast compared to CT and the use of MR-alone for prostate planning would provide further benefits such as lower cost to the patient. This study compares the accuracy of MR-alone based dose calculations with bulk electron density assignment to CT-based dose calculations for prostate radiotherapy.
MATERIALS AND METHODS: CT and whole pelvis MR images were contoured for 39 prostate patients. Plans with uniform density and plans with bulk density values assigned to bone and tissue were compared to the patient's gold standard full density CT plan. The optimal bulk density for bone was calculated using effective depth measurements. The plans were evaluated using ICRU point doses, dose volume histograms, and Chi comparisons. Differences in spatial uniformity were investigated for the CT and MR scans.
RESULTS: The calculated dose for CT bulk bone and tissue density plans was 0.1±0.6% (mean±1 SD) higher than the corresponding full density CT plan. MR bulk bone and tissue density plans were 1.3±0.8% lower than the full density CT plan. CT uniform density plans and MR uniform density plans were 1.4±0.9% and 2.6±0.9% lower, respectively. Paired t-tests performed on specific points on the DVH graphs showed that points on DVHs for all bulk electron density plans were equivalent with two exceptions. There was no significant difference between doses calculated on Pinnacle and Eclipse. The dose distributions of six patients produced Chi values outside the acceptable range of values when MR-based plans were compared to the full density plan.
CONCLUSIONS: MR-alone bulk density planning is feasible provided bone is assigned a density, however, manual segmentation of bone on MR images will have to be replaced with automatic methods. The major dose differences for MR bulk density plans are due to differences in patient external contours introduced by the MR couch-top and pelvic coil.
Copyright © 2011 Elsevier Ireland Ltd. All rights reserved.

Entities:  

Mesh:

Year:  2011        PMID: 21339009     DOI: 10.1016/j.radonc.2011.01.012

Source DB:  PubMed          Journal:  Radiother Oncol        ISSN: 0167-8140            Impact factor:   6.280


  44 in total

Review 1.  Magnetic resonance image guidance in external beam radiation therapy planning and delivery.

Authors:  Ilamurugu Arivarasan; Chandrasekaran Anuradha; Shanmuga Subramanian; Ayyalusamy Anantharaman; Velayudham Ramasubramanian
Journal:  Jpn J Radiol       Date:  2017-06-13       Impact factor: 2.374

2.  Multiatlas approach with local registration goodness weighting for MRI-based electron density mapping of head and neck anatomy.

Authors:  Reza Farjam; Neelam Tyagi; Harini Veeraraghavan; Aditya Apte; Kristen Zakian; Margie A Hunt; Joseph O Deasy
Journal:  Med Phys       Date:  2017-06-01       Impact factor: 4.071

3.  MRI-based treatment planning for brain stereotactic radiosurgery: Dosimetric validation of a learning-based pseudo-CT generation method.

Authors:  Tonghe Wang; Nivedh Manohar; Yang Lei; Anees Dhabaan; Hui-Kuo Shu; Tian Liu; Walter J Curran; Xiaofeng Yang
Journal:  Med Dosim       Date:  2018-08-14       Impact factor: 1.482

4.  Dose evaluation of MRI-based synthetic CT generated using a machine learning method for prostate cancer radiotherapy.

Authors:  Ghazal Shafai-Erfani; Tonghe Wang; Yang Lei; Sibo Tian; Pretesh Patel; Ashesh B Jani; Walter J Curran; Tian Liu; Xiaofeng Yang
Journal:  Med Dosim       Date:  2019-02-01       Impact factor: 1.482

5.  MR simulation for prostate radiation therapy: effect of coil mounting position on image quality.

Authors:  J Sun; P Pichler; J Dowling; F Menk; P Stanwell; J Arm; P B Greer
Journal:  Br J Radiol       Date:  2014-07-25       Impact factor: 3.039

6.  k-space sampling optimization for ultrashort TE imaging of cortical bone: applications in radiation therapy planning and MR-based PET attenuation correction.

Authors:  Lingzhi Hu; Kuan-Hao Su; Gisele C Pereira; Anu Grover; Bryan Traughber; Melanie Traughber; Raymond F Muzic
Journal:  Med Phys       Date:  2014-10       Impact factor: 4.071

7.  Dosimetric and workflow evaluation of first commercial synthetic CT software for clinical use in pelvis.

Authors:  Neelam Tyagi; Sandra Fontenla; Jing Zhang; Michelle Cloutier; Mo Kadbi; Jim Mechalakos; Michael Zelefsky; Joe Deasy; Margie Hunt
Journal:  Phys Med Biol       Date:  2016-12-16       Impact factor: 3.609

8.  Development of a hybrid magnetic resonance/computed tomography-compatible phantom for magnetic resonance guided radiotherapy.

Authors:  Min-Joo Kim; Seu-Ran Lee; Kyu-Ho Song; Hyeon-Man Baek; Bo-Young Choe; Tae Suk Suh
Journal:  J Radiat Res       Date:  2020-03-23       Impact factor: 2.724

Review 9.  MRI-only treatment planning: benefits and challenges.

Authors:  Amir M Owrangi; Peter B Greer; Carri K Glide-Hurst
Journal:  Phys Med Biol       Date:  2018-02-26       Impact factor: 3.609

10.  Investigation of a method for generating synthetic CT models from MRI scans of the head and neck for radiation therapy.

Authors:  Shu-Hui Hsu; Yue Cao; Ke Huang; Mary Feng; James M Balter
Journal:  Phys Med Biol       Date:  2013-11-11       Impact factor: 3.609

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.