Abstract

BACKGROUND: We report an evaluation of the Menarini/ARKRAY ADAMS A1c HA-8180V analyser (HA-8180V), the fifth generation Menarini/ARKRAY ion-exchange HPLC for the measurement of HbA(1c).
METHODS: We evaluated the analytical performance, the measurement of haemoglobin variants and the performance in comparison to major analytical methods.
RESULTS: Within-run, between-run and total CV were 0.2%, 0.4% and 0.7% at low HbA(1c) concentrations and 0.2%, 0.2% and 0.4% at high HbA(1c) concentrations, respectively. Trueness revealed a maximum deviation of 0.8 mmol/mol (IFCC units) or 0.1% (NGSP units) over the relevant analytical range. Linearity, carry-over and linear drift were excellent. Labile-HbA(1c), carbamylated haemoglobin, icteric samples and variation in hematocrit did not affect HbA(1c) outcome. Haemoglobin variants AS, AC and F do not affect HbA(1c) outcome and are explicitly identified and correctly quantified. HbA(1c) can not be measured in samples with AE and AD, but these variants are identified correctly. In comparison to other methods used at present, the HA-8180V shows excellent performance.
CONCLUSIONS: The HA-8180V performs at a high level and is fit for any clinical application.

Entities:  

Mesh:

Substances:

Year:  2011        PMID: 21323625     DOI: 10.1515/CCLM.2011.096

Source DB:  PubMed          Journal:  Clin Chem Lab Med        ISSN: 1434-6621            Impact factor:   3.694


  3 in total

1.  Comparison of Arkray/ELITech ADAMS HA-8180V with Bio-Rad Variant, II Turbo2.0 and Tosoh Bioscience HLC-723G8 for HbA1c determination.

Authors:  Stéphanie Badiou; Julie Guillot; Nils Kuster; Anne-Sophie Bargnoux; Patricia Aguilar-Martinez; Elodie Boissier; Jean-Paul Cristol; Anne-Marie Dupuy
Journal:  J Clin Lab Anal       Date:  2014-03-20       Impact factor: 2.352

Review 2.  Methods, units and quality requirements for the analysis of haemoglobin A1c in diabetes mellitus.

Authors:  Ilkka Penttilä; Karri Penttilä; Päivi Holm; Harri Laitinen; Päivi Ranta; Jukka Törrönen; Rainer Rauramaa
Journal:  World J Methodol       Date:  2016-06-26

3.  Evaluation of Bio-Rad D-100 HbA1c analyzer against Tosoh G8 and Menarini HA-8180V.

Authors:  José María Maesa; Patricia Fernández-Riejos; Catalina Sánchez Mora; María de Toro; Paloma Menéndez Valladares; Concepción González-Rodriguez
Journal:  Pract Lab Med       Date:  2016-05-14
  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.