Alanine/EPR dosimetry applied to the verification of a total body irradiation protocol and treatment planning dose calculation using a humanoid phantom.

B Schaeken1, S Lelie, P Meijnders, D Van den Weyngaert, H Janssens, D Verellen.   

Abstract

PURPOSE: To avoid complications in total body irradiation (TBI), it is important to achieve a homogeneous dose distribution throughout the body and to deliver a correct dose to the lung which is an organ at risk. The purpose of this work was to validate the TBI dose protocol and to check the accuracy of the 3D dose calculations of the treatment planning system.
METHODS: Dosimetry based on alanine/electron paramagnetic resonance (EPR) was used to measure dose at numerous locations within an anthropomorphic phantom (Alderson) that was irradiated in a clinical TBI beam setup. The alanine EPR dosimetry system was calibrated against water calorimetry in a Co-60 beam and the absorbed dose was determined by the use of "dose-normalized amplitudes" A(D). The dose rate of the TBI beam was checked against a Farmer ionization chamber. The phantom measurements were compared to 3D dose calculations from a treatment planning system (Pinnacle) modeled for standard dose calculations.
RESULTS: Alanine dosimetry allowed accurate measurements which were in accordance with ionization chamber measurements. The combined relative standard measurement uncertainty in the Alderson phantom was U(r)(A(D))=0.6%. The humanoid phantom was irradiated to a reference dose of 10 Gy, limiting the lung dose to 7.5 Gy. The ratio of the average measured dose midplane in the craniocaudal direction to the reference dose was 1.001 with a spread of +/- 4.7% (1 sd). Dose to the lung was measured in 26 locations and found, in average, 1.8% lower than expected. Lung dose was homogeneous in the ventral-dorsal direction but a dose gradient of 0.10 Gy cm(-1) was observed in the craniocaudal direction midline within the lung lobe. 3D dose calculations (Pinnacle) were found, in average, 2% lower compared to dose measurements on the body axis and 3% lower for the lungs.
CONCLUSIONS: The alanine/EPR dosimetry system allowed accurate dose measurements which enabled the authors to validate their TBI dose protocol. Dose calculations based on a collapsed cone convolution dose algorithm modeled for regular treatments are accurate within 3% and can further be improved when the algorithm is modeled for TBI.

Entities:  

Mesh:

Substances:

Year:  2010        PMID: 21302785     DOI: 10.1118/1.3496355

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  6 in total

1.  Characterization of phenolic pellets for ESR dosimetry in photon beam radiotherapy.

Authors:  Salvatore Gallo; Giuseppina Iacoviello; Salvatore Panzeca; Ivan Veronese; Antonio Bartolotta; Daniele Dondi; Anna Maria Gueli; Gianfranco Loi; Anna Longo; Eleonora Mones; Maurizio Marrale
Journal:  Radiat Environ Biophys       Date:  2017-09-19       Impact factor: 1.925

Review 2.  Clinical applications of alanine/electron spin resonance dosimetry.

Authors:  Oswaldo Baffa; Angela Kinoshita
Journal:  Radiat Environ Biophys       Date:  2014-01-08       Impact factor: 1.925

3.  Time-gated scintillator imaging for real-time optical surface dosimetry in total skin electron therapy.

Authors:  Petr Bruza; Sarah L Gollub; Jacqueline M Andreozzi; Irwin I Tendler; Benjamin B Williams; Lesley A Jarvis; David J Gladstone; Brian W Pogue
Journal:  Phys Med Biol       Date:  2018-05-02       Impact factor: 3.609

4.  Total body irradiation with step translation and dynamic field matching.

Authors:  Ho-Hsing Chen; Jay Wu; Keh-Shih Chuang; Jia-Fu Lin; Jia-Cheng Lee; Jin-Ching Lin
Journal:  Biomed Res Int       Date:  2013-07-01       Impact factor: 3.411

5.  Fast Monte Carlo simulation for total body irradiation using a (60)Co teletherapy unit.

Authors:  Xiaodong Liu; Danielle Lack; Joseph T Rakowski; Cory Knill; Michael Snyder
Journal:  J Appl Clin Med Phys       Date:  2013-05-06       Impact factor: 2.102

6.  Positron emission tomography guided dose painting by numbers of lung cancer: Alanine dosimetry in an anthropomorphic phantom.

Authors:  Iosif Papoutsis; Ingerid Skjei Knudtsen; Erlend Peter Skaug Sande; Bernt Louni Rekstad; Michel Öllers; Wouter van Elmpt; Marius Røthe Arnesen; Eirik Malinen
Journal:  Phys Imaging Radiat Oncol       Date:  2022-02-26
  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.