A comprehensive study of the Gafchromic EBT2 radiochromic film. A comparison with EBT.

C Andrés1, A del Castillo, R Tortosa, D Alonso, R Barquero.   

Abstract

PURPOSE: Radiochromic film has become an important tool to assess complex dose distributions. In particular, EBT was accepted by the scientific community as a reference two-dimensional detector. Recently, Gafchromic EBT2 has replaced old film, providing new improvements in both accuracy and handling.
METHODS: This work presents a dosimetric study of the new Gafchromic EBT2 using an Epson 10000XL flatbed scanner, also comparing the results with EBT film as reference when necessary. The most important film characteristics have been studied, such as ambient light sensitivity, different possibilities of the three RGB color channels, postirradiation development, high dose behavior, exposition at temperatures similar to the human body, and dependence on orientation during the scanning process.
RESULTS: The results obtained confirm a considerably lower sensitivity to ambient light of EBT2, as well as a fast stabilization of the film within 2 h. It has also been found that the green channel has a better behavior at high dose levels up to 35 Gy, in addition to good behavior of the red channel at doses below 10 Gy. Other features, such as temperature independence and scanning orientation dependence, have also been shown.
CONCLUSIONS: Gafchromic EBT2 can be used for clinical practice in the same way as the old EBT film. However, a much easier handling as the result of all new enhancements improves film behavior, expanding in this way the potential applications of radiochromic film dosimetry.

Entities:  

Mesh:

Year:  2010        PMID: 21302783     DOI: 10.1118/1.3512792

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  31 in total

1.  Commissioning a small-field biological irradiator using point, 2D, and 3D dosimetry techniques.

Authors:  Joseph Newton; Mark Oldham; Andrew Thomas; Yifan Li; John Adamovics; David G Kirsch; Shiva Das
Journal:  Med Phys       Date:  2011-12       Impact factor: 4.071

2.  Evaluation of beam modeling for small fields using a flattening filter-free beam.

Authors:  Daisuke Kawahara; Shuichi Ozawa; Takeo Nakashima; Masamichi Aita; Shintaro Tsuda; Yusuke Ochi; Takuro Okumura; Hirokazu Masuda; Yoshimi Ohno; Yuji Murakami; Yasushi Nagata
Journal:  Radiol Phys Technol       Date:  2016-06-21

3.  A novel high-throughput irradiator for in vitro radiation sensitivity bioassays.

Authors:  Tyler L Fowler; Regina K Fulkerson; John A Micka; Randall J Kimple; Bryan P Bednarz
Journal:  Phys Med Biol       Date:  2014-02-28       Impact factor: 3.609

4.  Image-guided microbeam irradiation to brain tumour bearing mice using a carbon nanotube x-ray source array.

Authors:  Lei Zhang; Hong Yuan; Laurel M Burk; Christy R Inscoe; Michael J Hadsell; Pavel Chtcheprov; Yueh Z Lee; Jianping Lu; Sha Chang; Otto Zhou
Journal:  Phys Med Biol       Date:  2014-02-20       Impact factor: 3.609

5.  A study on comparison of Gafchromic EBT2 film response under single and cumulative exposure conditions.

Authors:  K Ganapathy; P G G Kurup; V Murali; M Muthukumaran; J Velmurugan
Journal:  J Med Phys       Date:  2013-10

6.  Towards Establishing Capacity for Biological Dosimetry at Ghana Atomic Energy Commission.

Authors:  Daniel Gyingiri Achel; Elom Achoribo; Sandra Agbenyegah; Rudolph M Adaboro; Shadrack Donkor; Nana A K Adu-Bobi; Akwasi A Agyekum; Felicia Akuamoa; Samuel N Tagoe; Kofi A Kyei; Joel Yarney; Antonio Serafin; John M Akudugu
Journal:  Genome Integr       Date:  2016-12-30

7.  The use of radiochromic EBT2 film for the quality assurance and dosimetric verification of 3D conformal radiotherapy using Microtek ScanMaker 9800XL flatbed scanner.

Authors:  G S Sim; J H D Wong; K H Ng
Journal:  J Appl Clin Med Phys       Date:  2013-07-08       Impact factor: 2.102

8.  Dosimetric characterization and use of GAFCHROMIC EBT3 film for IMRT dose verification.

Authors:  Valeria Casanova Borca; Massimo Pasquino; Giuliana Russo; Pierangelo Grosso; Domenico Cante; Piera Sciacero; Giuseppe Girelli; Maria Rosa La Porta; Santi Tofani
Journal:  J Appl Clin Med Phys       Date:  2013-03-04       Impact factor: 2.102

9.  Evaluation of triple channel correction acquisition method for radiochromic film dosimetry.

Authors:  Naoki Hayashi; Yoichi Watanabe; Ryan Malmin; Hideki Kato
Journal:  J Radiat Res       Date:  2012-08-21       Impact factor: 2.724

10.  Skin Dosimetry in Radiotherapy of Breast Cancer: a Comparison between EBT and EBT3 Radiochromic Films.

Authors:  M T Bahreyni Toosi; N Mohamadian; M Ghorbani; F Khorshidi; F Akbari; C Knaup
Journal:  J Biomed Phys Eng       Date:  2016-06-01
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.