Bacterial screening by flow cytometry offers potential for extension of platelet storage: results of 14 months of active surveillance.

T Vollmer1, J Engemann, K Kleesiek, J Dreier.   

Abstract

BACKGROUND: Bacterial contamination is currently the major infectious hazard of platelet transfusion in developed countries. It has been demonstrated that a significant transfusion risk remains, in particular with older platelet concentrates (PCs). In 2009, the shelf life of PCs was therefore reduced in Germany to 4 days after the day of production according to Vote 38. The aim of the present study was the application and implementation of a recently developed flow cytometry-based rapid screening method (BactiFlow) for bacterial contamination at the end of PC shelf life as a routine in-process control. STUDY DESIGN AND METHODS: A total of 472 apheresis-derived PCs were tested using the BactiFlow flow cytometric assay to detect and count bacteria based on esterase activity in viable bacterial cells, while the BacT/Alert automated culture system served as the reference method. The automation potential of the flow cytometric assay was analysed by applying the semi-automated BactiFlow ALS system.
RESULTS: An algorithm was developed for use in routine blood bank operations to extend the storage period of PCs. Two of the 472 apheresis PCs tested were positive in culture and identified as Propionibacterium species. One PC was positive for Staphylococcus aureus by both methods. All remaining specimens were tested negative by both methods.
CONCLUSIONS: Our study demonstrates that routine bacterial testing of PCs was successfully implemented and the established algorithm proved efficient. The BactiFlow flow cytometric assay is the first rapid screening method which is suitable for a routine application combined with a high sensitivity.
© 2011 The Authors. Transfusion Medicine © 2011 British Blood Transfusion Society.

Entities:  

Mesh:

Year:  2011        PMID: 21272103     DOI: 10.1111/j.1365-3148.2011.01070.x

Source DB:  PubMed          Journal:  Transfus Med        ISSN: 0958-7578            Impact factor:   2.019


  5 in total

1.  Implementation of Bacterial Detection Methods into Blood Donor Screening - Overview of Different Technologies.

Authors:  Michael Schmidt; Walid Sireis; Erhard Seifried
Journal:  Transfus Med Hemother       Date:  2011-07-07       Impact factor: 3.747

2.  Extension of the Storage Period of Platelet Concentrates in Germany to 5 Days by Bacterial Testing: Is it Worth the Effort?

Authors:  Tanja Vollmer; Dennis Hinse; Jürgen Diekmann; Cornelius Knabbe; Jens Dreier
Journal:  Transfus Med Hemother       Date:  2019-03-14       Impact factor: 3.747

Review 3.  Diagnostic methods for platelet bacteria screening: current status and developments.

Authors:  Melanie Störmer; Tanja Vollmer
Journal:  Transfus Med Hemother       Date:  2013-12-30       Impact factor: 3.747

4.  Bacterial screening of platelet concentrates on day 2 and 3 with flow cytometry: the optimal sampling time point?

Authors:  Tanja Vollmer; Volkmar Schottstedt; Juergen Bux; Gabriele Walther-Wenke; Cornelius Knabbe; Jens Dreier
Journal:  Blood Transfus       Date:  2014-01-02       Impact factor: 3.443

5.  Bench Test for the Detection of Bacterial Contamination in Platelet Concentrates Using Rapid and Cultural Detection Methods with a Standardized Proficiency Panel.

Authors:  Tanja Vollmer; Cornelius Knabbe; Wolf-Jochen Geilenkeuser; Michael Schmidt; Jens Dreier
Journal:  Transfus Med Hemother       Date:  2015-07-16       Impact factor: 3.747

  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.