Active illumination microscopy (AIM) is a method of redistributing dynamic range in a scanning microscope using real-time feedback to control illumination power on a sub-pixel time scale. We describe and demonstrate a fully integrated instrument that performs both feedback and image reconstruction. The image is reconstructed on a logarithmic scale to accommodate the dynamic range benefits of AIM in a single output channel. A theoretical and computational analysis of the influence of noise on active illumination feedback is presented, along with imaging examples illustrating the benefits of AIM. While AIM is applicable to any type of scanning microscope, we apply it here specifically to two-photon microscopy.
Active illumination microscopy (AIM) is a method of redistributing dynamic range in a scanning microscope using real-time feedback to control illumination power on a sub-pixel time scale. We describe and demonstrate a fully integrated instrument that performs both feedback and image reconstruction. The image is reconstructed on a logarithmic scale to accommodate the dynamic range benefits of AIM in a single output channel. A theoretical and computational analysis of the influence of noise on active illumination feedback is presented, along with imaging examples illustrating the benefits of AIM. While AIM is applicable to any type of scanning microscope, we apply it here specifically to two-photon microscopy.
Scanning fluorescence microscopy techniques such as confocal [1] and two-photon [2]
microscopy have become indispensable tools in the biomedical community. In their
conventional implementations, a laser beam of constant power is scanned through a
sample, leading to a fixed illumination energy per pixel integration time. Recently,
it has been shown that a dynamic control of the illumination energy per pixel
integration time can lead to several benefits. For example, Controlled Light
Exposure Microscopy (CLEM) [3], wherein the
laser power is shut off once a prescribed amount of signal is detected from a given
pixel, minimizes exposure of the sample to unnecessary illumination, leading to
reduced photobleaching. Alternatively, a technique called Adaptive Illumination
Microscopy (AIM) [4] operates by adjusting
the laser power in an analog manner to maintain a constant signal level per pixel.
The benefits of AIM are primarily the elimination of detector saturation, thanks to
the ability of a feedback circuit to scale down the illumination power for bright
objects, and an enhanced sensitivity to weak signals resulting from an increased
illumination of dim objects. These two improvements can together be considered an
enhancement of dynamic range. A difficulty with our previous implementation of AIM
was that it required two detection channels, one for the signal and another for the
laser power. We describe here a much more practical implementation of AIM that
requires only a single detection channel, and can be operated as a simple drop-in
addition to any standard scanning microscope, provided this is equipped with an
analog laser power controller. Moreover, we equip our AIM microscope with direct log
output to accommodate our enhanced dynamic range. While AIM is a general technique
that can be applied to any scanning microscope, we apply it here specifically to
two-photon microscopy, and provide a detailed theoretical description of the
resultant benefits in signal to noise ratio (SNR) and dynamic range (DR). These
results are corroborated with two-photon microscopy images of GFP labeled neurons in
a mouse brain slice. Finally, we note that we have changed the name of our technique
to Active Illumination Microscopy (still abbreviated AIM) to avoid confusion with
adaptive optics, which is unrelated to this work.
2. Layout
The overall setup of the system is shown in Fig.
1. This is a conventional two-photon excited fluorescence (TPEF) microscope
equipped with an electro-optic modulator (EOM, Conoptics 350-80) to control the
laser power P. The amplifiers used in the feedback circuit are high
bandwidth op-amps, primarily the Maxim Integrated Products MAX477 (300 MHz
bandwidth) and the Texas Instruments OPA604 (20 MHz bandwidth). The bandwidth of the
system is limited therefore by the EOM and not the electronics. The typical pixel
rate of our TPEF system is approximately 100 kHz, while the EOM bandwidth approaches
1 MHz, allowing the circuit to rapidly control power on an intra-pixel time scale.
Fig. 1.
AIM layout for two-photon microscopy. Dashed line represents the
integrated AIM instrument. Fluorescent output
XP2 from a sample X is
detected by a photomultiplier tube (PMT), producing a signal
S that is maintained at a set point
S by analog feedback to an EOM that
controls the illumination power P. X
is reconstructed and output on a log scale.
We define the fluorescent sample strength X to be a variable that
includes all factors contributing to the local fluorescent emissivity, including
concentration, cross-section, and quantum yield, such that the detected two-photon
excited fluorescence is S = XP2. The
feedback system, encased in the dotted line in Fig.
1, uses negative feedback to hold the detected fluorescence to a constant
S by controlling the input illumination
power, up to a user-defined maximum power P. When
sufficient power is available to hold S, the system
is in “feedback-active” mode. When more than P would
be needed to reach S, the power is automatically set
to P and the system switches to “power-limited” mode.
In either case, the desired quantity of interest is the sample strength
X = S/P2. This is
evaluated directly in the electronic instrument and supplied as an output. We
emphasize that our feedback system is entirely self-contained here, and can be
implemented in conventional TPEF microscopy as a simple add-on. The output of our
instrument thus plays the role of the direct detector output of a conventional
microscope.The conversion to a single output instrument represents a practical improvement over
our previous implementation of AIM [4] in
which two output channels were recorded simultaneously (S and
P) and image reconstruction was performed in software.
Representing reconstructed AIM data in a single channel allows AIM to be more
readily applied as a drop-in enhancement to a wider selection of instruments, since
no software reconstruction is required to visualize the AIM image in real time, and
not all microscope systems have two available input channels. However, care must be
taken that the gain in dynamic range obtained by AIM not be undermined by a
limitation in the dynamic range of the process itself of analog-to-digital
conversion of our single output. In order to fully accommodate the enhanced dynamic
range of our system, while maintaining the practicality of a single output, we
decided to represent our reconstructed image (i.e. X) in
logarithmic scale, achieved using Texas Instruments logarithmic amplifier TL441
(bandwidth = 40 MHz). On the log scale, the output of our AIM is given byAIM layout for two-photon microscopy. Dashed line represents the
integrated AIM instrument. Fluorescent output
XP2 from a sample X is
detected by a photomultiplier tube (PMT), producing a signal
S that is maintained at a set point
S by analog feedback to an EOM that
controls the illumination power P. X
is reconstructed and output on a log scale.(Note: the factor of 2 is specific to two-photon excitation) On a linear scale, each
successive step on the ladder represents an equal change in signal, whereas on a
logarithmic scale each step corresponds to an equal change in decibels (or
percentage) from the previous rung on the ladder. Consequently, on the low end of
the range, the step size on the log-scale is much smaller than on the high end. The
price of the low-end precision is degraded high-end precision. However, this is not
a heavy cost. Though the signal-to-noise ratio of a shot noise-limited measurement
scales with √N (where N is the number of detected
photons for a given pixel), the overall noise increases at higher photon counts,
also scaling with √N. Equal precision at the high range is
therefore wasted, since the uncertainty in that range is higher anyway. A log-scale
output is thus well-suited to AIM images, since this matches the characteristic of
good low signal precision and high overall range. Additionally, from a practical
standpoint, the reconstruction equation is more easily implemented in the
electronics as a logarithm (subtraction) than on a linear scale (division).Using a log scale amplifier alone without performing AIM will not increase dynamic
range unless the limiting factor is the analog-to-digital converter, which is rarely
the case. The dynamic range of the detector, which is usually the bottleneck, cannot
be increased by processing the signal differently downstream. AIM is able to
circumvent dynamic range limitations of the detector by allowing the illumination
power to vary and carry information. We use the log amplifier to ensure that the new
enhanced dynamic range is not lost when digitized by the computer.It should be noted that whether data is represented in linear or log scale, it
remains the same data (arguments related to SNR and dynamic range notwithstanding).
Any subsequent conversion from log scale to linear scale, if desired, can therefore
readily be performed a posteriori in software (we will be seeing examples of this in
Section 5).
3. Principle of AIM
The shot-noise limited SNR’s for conventional and AIM systems are illustrated in
Figs. 2a and 2b, respectively. We represent a minimum acceptable SNR
(which may simply be 1, where signal begins to exceed noise) and a maximum SNR (due
to saturation of the detector) as dashed lines. When the SNR is between these lines,
we are within the dynamic range of our system.
Fig. 2.
Signal to noise ratio (SNR) vs. fluorescent sample strength
(X) under varying AIM settings. SNR follows standard
shot noise model in the power-limited regime but is capped to a constant
maximum once the system reaches S.
Signal to noise ratio (SNR) vs. fluorescent sample strength
(X) under varying AIM settings. SNR follows standard
shot noise model in the power-limited regime but is capped to a constant
maximum once the system reaches S.In a conventional TPEF system, the SNR rises proportionately with √X
due to the properties of shot noise. Figure
2a provides an illustration of the dynamic range of the system when using an
illumination power P0. Increasing power, up to the
maximum available laser power, enables the measurement of lower values of
X, but also lowers the saturation threshold. The dynamic range
X/X of a
conventional system remains constant regardless of power setting.In an AIM system, there are two regimes of operation, as illustrated in Fig. 2b. A feedback circuit is designed to
hold the detected signal at a user-defined set point, thus preventing the
possibility of detector saturation. When the sample is strong enough for the set
point to be attained, the system is said to be in feedback-active mode. On the other
hand, if the sample is so weak that the maximum available power is insufficient to
attain the set point, then the system switches to power-limited mode. When in
power-limited mode, the AIM essentially behaves as a conventional system, but with
an illumination power P rather than
P0, thereby enabling a measurement of the smallest
possible X.The question arises as to what values should be used for and
S and P. When in
feedback-active mode, the number of detected photons per pixel dwell time is held
constant, and hence the shot-noise limited SNR is also held constant. The user can
therefore select S based on a desired SNR
performance. Since SNR is constant, noise becomes a constant fraction of the signal,
which on the logarithmic scale is simply additive noise whose magnitude is
independent of X. However, aside from preventing saturation,
limiting signal output from a local region of the sample also limits phototoxic
effects. Thus, while higher S improves the SNR in
feedback-active mode, it also increases the potential for phototoxicity. Similarly,
in power-limited mode, while a higher P allows
measurement of smaller X, it also increases the potential for
photobleaching. P is therefore practically limited
not only by available laser power, but by the maximum permissible photobleaching.It should be noted that photobleaching and phototoxicity are related but disparate
phenomena. Photobleaching is the permanent transformation of a fluorophore into a
non-fluorescent state [5], and phototoxicity
refers to the adverse biological effects, usually caused by the generation of free
radicals or singlet oxygen during fluorescent imaging [6]. For a given fluorophore species and excitation wavelength,
photobleaching rate is determined primarily by illumination power (fluorophore
concentration independent), whereas phototoxicity effects depend on total
illumination absorption from a given volume (concentration and illumination
dependent). A high concentration of fluorophores illuminated at low intensity will
exhibit less photobleaching than a low concentration of fluorophores imaged under
high excitation power such that the total fluorescent output is equal in both cases.
However, both scenarios do result in equal phototoxicity.In order to reduce phototoxicity in a conventional scanning fluorescence system, the
only recourse is to reduce overall illumination power, the obvious cost of which is
reduced SNR in regions of weak sample strength despite low risk of phototoxicity in
these areas. On the other hand, with the AIM system, phototoxicity is constrained by
S, whereas photobleaching, which becomes
potentially problematic only when the sample strength is low, is constrained by
P. The proper choices of
S and P that
maximize image quality while abiding by the constraints of phototoxicity and
photobleaching thus ultimately depend on the sample in question.We also note that while AIM provides a mechanism for limiting photobleaching by
limiting P, the same could be done on an ordinary
TPEF microscope by simply turning down the power. If a sample can be imaged using
standard TPEF using a given power level without saturation, AIM will not provide an
improvement in image quality when P is set to the
same power level. Additionally, photobleaching in the focal plane may be somewhat
nonuniform due to the varying illumination power, which may complicate the use of
any photobleaching correction algorithms. However, AIM will still provide a
reduction in phototoxicity, since any trouble areas that generate high fluorescent
output per unit volume will receive reduced illumination.
4. Analysis of performance
In conventional imaging, the impact of noise sources such as electronic noise is well
understood. However, with AIM, we must evaluate noise in the context of a
closed-loop feedback system. We will assume that the detector introduces some amount
of noise, represented by the random variable δS, such that the
detected signal is Ŝ =
GXP2+δS, where G is
the detector gain. We also introduce a power-control noise δP that
represents a discrepancy between the electronically measured signal
P and the actual laser power leaving the EOM. Figure 3 shows the feedback system diagram with noise sources
included.
Fig. 3.
Block diagram of feedback circuit components with noise introduced.
Block diagram of feedback circuit components with noise introduced.The illumination power determined by the feedback system can be written aswhere H is the transfer function of the feedback system. Taking the
limit where H is very large, and using the approximation that
δP is small compared to P, we haveSolving for P and ignoring the nonphysical negative solution leads
toWe attempt to recover X using our noisy measurements of
S and P to yield the reconstructed
X̂:We can rearrange the above in terms of the desired signal X:Again using small noise assumptions while substituting in P from
Eq. (4):The SNR associated with our reconstruction of X is thuswhere the noise standard deviations are defined by and .From this equation, we see that the SNR from independent circuit noises scales with
the set point S. When the detector noise is dominant
over power-control noise (σ >
σ), the SNR scales directly with
S. When the reverse is true
(σ > σ) the SNR
scales with .Let us compare this result with the SNR for a conventional TPEF microscope that uses
a power P0 to image a fluorescence strength
X. Before proceeding, we note that the noise
σ can be broken down into two independent components:
laser intensity fluctuations σ and electronic
measurement error σ; σ
is common to both conventional TPEF and AIM, while σ
is unique to AIM. Using the same process as above, the SNR for conventional
two-photon imaging can be shown to beThe ratio SNRAI/SNR0 should be greater than unity in order for
AIM to yield an advantage. Using Eqs.
8 and 9, the ratio iswhereIn Eq. 10,
S/GXP02 is
the ratio of the signals, and ξ is the ratio of the noises. The
behavior of ξ depends on whether detector noise or power noise is
more dominant. The best-case scenario is when σ is
dominant, and ξ → 1. When the σ terms
exceed σ, ξ scales with
√X, reflecting the fact that AIM uses lower P for
higher X, making this term more susceptible to fluctuations and
noise.If the power signal is known with perfect accuracy (i.e.
σ = 0), and the set point is fixed to same signal
level recorded in conventional imaging (i.e. S =
GXP02), then the SNR of AIM is identical
to the conventional case, as expected.We now evaluate the effect of this system on dynamic range (DR). DR is typically
treated as the maximum recordable value divided by the minimum resolvable difference
in values (which is often defined by noise rather than bit-depth). In a conventional
configuration, the maximum measureable sample strength is
X =
S/P02
where S is the output intensity at which detector
saturation occurs. The minimum resolution is similarly
σ =
σ/P02. Thus,
for conventional imaging the DR, defined by
X/σ, is given byIn AIM, saturation only occurs when the feedback circuit has already lowered
P to the minimum resolvable value, which is
σ, in order to hold S =
S. Therefore X
= S/σ2. The
minimum σ is defined the same way as in the
conventional case, since at the low end of X the AIM system is
power-limited while using constant illumination power
P, so σ =
σ/P2.
The DR of AIM is thenThe dynamic range enhancement of AIM therefore lies in the term
(P/σ)2,
which is simply the dynamic range of the illumination power control system, with an
extra power of 2 as a result of two-photon imaging. The last term,
S/S, accounts for
the eventuality that the full dynamic range of the detector is not used. That is,
because of feedback, no value of S higher than
S can occur. In practice,
(P/σ)2
can be several orders of magnitude while
S/S can be set
fairly close to 1 (limited only by the possibility of phototoxicity). That is, in
practice the DR improvement obtained with AIM can be quite substantial, as we will
see below.
5. Results
Imaging results in Fig. 4 demonstrate the
advantages of AIM.
Fig. 4.
Image of mouse brain labeled with GFP (scale bar 20 μm).
Conventional TPEF image, linear scale (a) and log scale (c). AIM image,
linearized in software (b) and as acquired on log scale (d). Magnified
insets of (c) and (d) shown respectively in (e) and (f). Video online of
(a) to (d) showing a depth scan through 98 μm at a 2
μm spacing between slices.(Media 1)
Image of mouse brain labeled with GFP (scale bar 20 μm).
Conventional TPEF image, linear scale (a) and log scale (c). AIM image,
linearized in software (b) and as acquired on log scale (d). Magnified
insets of (c) and (d) shown respectively in (e) and (f). Video online of
(a) to (d) showing a depth scan through 98 μm at a 2
μm spacing between slices.(Media 1)The sample imaged is a mouse brain hippocampus labeled by cytoplasmic EGFP by
in utero electroporation [7]. Figs. 4a and 4d are the unaltered images obtained with TPEF
microscopy operated in standard mode (linear scale) and with AIM (logarithmic
scale). These are the images as they are acquired by the user in real time. The log
AIM image of course exhibits increased visibility in weakly fluorescent areas simply
due to the logarithmic scale, however the benefits of AIM go much beyond this simple
change of scale. The pixel rate on these images is 125 kHz, which is well within the
roughly 1 MHz feedback system bandwidth.To properly compare both images on the same scale, either a logarithm can be applied
numerically to the conventional image, or an anti-logarithm can be applied
numerically to the AIM image. For example, comparing both images on the logarithmic
scale (Figs. 4c and 4d) demonstrates that the improvement from AIM is not limited
to log-scale contrast. The application of a logarithm to the data from Fig. 4a produces Fig. 4c, but the simple emulation of the log lookup table
does not duplicate the improved SNR available to the AIM system resulting from
increased illumination power. While the total laser power available with or without
AIM is unchanged, the maximum power that can be used in a conventional image is
limited by saturation. For the conventional images in Fig. 4, the illumination power was set as high as possible
without experiencing detector saturation (90 mW at 835 nm entering the scan head).
AIM is able to use more power (220 mW) yet avoids saturation.Line profile of power spectral density of images from Fig. 4a (conventional image, blue
trace) and 4b (AIM image, black
trace). Power density shown on a log scale against
k on x-axis and
k = 0. The noise floor for AIM (black) is
lower than for normal imaging (blue), reflecting improved SNR.Alternatively, Fig. 4b shows the AIM image in
4d reconstructed on a linear scale. That is, once the data is collected on a
logarithmic scale, the AIM image may be readily converted to a linear scale in
software, with the same result that the effective dynamic range of the linearized
image produced from AIM is higher than in the conventional linear image (Fig. 4a). Though not as obvious when presented
on a linear scale, the SNR associated with dimmer regions appears superior when
using AIM for the same reasons described above. In both AIM images, the transition
between power-limited and feedback-active pixels is seamless and cannot be seen on
the images.The SNR improvement is also demonstrated in the power spectral density functions of
an arbitrary line through the linear images, shown in Fig. 5. The ratio of the conventional to AI noise floor
levels is 6.8, determined by averaging each power density function outside the
spatial frequency k = 1/6 pixels-1, where
all of the power present is due to noise. This approximately corresponds to the
ratio expected from the power increase when using AIM, which is
(P/P0)2 =
6.
Fig. 5.
Line profile of power spectral density of images from Fig. 4a (conventional image, blue
trace) and 4b (AIM image, black
trace). Power density shown on a log scale against
k on x-axis and
k = 0. The noise floor for AIM (black) is
lower than for normal imaging (blue), reflecting improved SNR.
Figure 4 primarily showcases the low-signal
advantages of AIM when the conventional two-photon image is driven to the brink of
saturation. To showcase the high-signal benefits, conventional two-photon microscopy
is taken beyond the brink of saturation in Fig.
6 (top). The problem of saturation is completely remedied in Fig. 6 (bottom) where we observe that the AIM
image not only avoids saturation in the strong sample regions but does so without
sacrificing SNR in the weak sample regions.
Fig. 6.
Image of GFP-labeled mouse neurons; conventional TPEF (top) cannot
properly quantify what should be a bright neuron body indicated by arrow
due to saturation. Saturation is avoided and the fluorescence of the
neuron is properly captured in the linearized AIM image (bottom) without
sacrificing SNR for dim objects. Arbitrary units of fluorescence are
represented by a color look-up table (right) to highlight this effect;
the bottom of the bar is zero fluorescence. Scale bar is 20
μm.
Image of GFP-labeled mouse neurons; conventional TPEF (top) cannot
properly quantify what should be a bright neuron body indicated by arrow
due to saturation. Saturation is avoided and the fluorescence of the
neuron is properly captured in the linearized AIM image (bottom) without
sacrificing SNR for dim objects. Arbitrary units of fluorescence are
represented by a color look-up table (right) to highlight this effect;
the bottom of the bar is zero fluorescence. Scale bar is 20
μm.In summary, we have shown that our log-scale AIM system is able to confer dynamic
range and exposure management improvements while using a single A/D channel. We
demonstrate imaging results using TPEF microscopy. Additionally, since all of the
components are integrated into a self-contained instrument, our AIM device can be
easily applied to any scanning imaging system equipped with a fast modulator, such
as a scanning confocal microscope equipped with acousto-optic modulator controlled
laser illumination.
Authors: R A Hoebe; C H Van Oven; T W J Gadella; P B Dhonukshe; C J F Van Noorden; E M M Manders Journal: Nat Biotechnol Date: 2007-01-21 Impact factor: 54.908
Authors: C Vinegoni; C Leon Swisher; P Fumene Feruglio; R J Giedt; D L Rousso; S Stapleton; R Weissleder Journal: Nat Commun Date: 2016-04-01 Impact factor: 14.919