Literature DB >> 21258237

Impact of applying the more stringent validation criteria of the revised European Society of Hypertension International Protocol 2010 on earlier validation studies.

George S Stergiou1, Nikos Karpettas, Neil Atkins, Eoin O'Brien.   

Abstract

OBJECTIVE: Since 2002 when the European Society of Hypertension International Protocol (ESH-IP) was published it has become the preferred protocol for validating blood pressure monitors worldwide. In 2010, a revised version of the ESH-IP with more stringent criteria was published. This study assesses the impact of applying the revised ESH-IP criteria.
METHODS: A systematic literature review of ESH-IP studies reported between 2002 and 2010 was conducted. The impact of applying the ESH-IP 2010 criteria retrospectively on the data reported in these studies was investigated. The performance of the oscillometric devices in the last decade was also investigated on the basis of the ESH-IP criteria.
RESULTS: Among 119 published studies, 112 with sufficient data were analyzed. According to ESH-IP 2002, the test device failed in 19 studies, whereas by applying the ESH-IP 2010 criteria in 28 additional studies increased the failure rate from 17 to 42%. Of these 28 studies, in 20 (71%) the test device failed at part 1 (accuracy per measurement) and in 22 (79%) at part 2 (accuracy per subject). Most of the failures involved the '5 mmHg or less' criterion. In the last decade there has been a consistent trend toward improved performance of oscillometric devices assessed on the basis of the ESH-IP criteria.
CONCLUSION: This retrospective analysis shows that the stricter revised ESH-IP 2010 criteria will noticeably increase the failure rate of devices being validated. Oscillometric devices are becoming more accurate, and the revised ESH-IP by acknowledging this trend will allow more accurate devices to enter the market.

Entities:  

Mesh:

Year:  2011        PMID: 21258237     DOI: 10.1097/MBP.0b013e32834331e7

Source DB:  PubMed          Journal:  Blood Press Monit        ISSN: 1359-5237            Impact factor:   1.444


  6 in total

Review 1.  Policy statement of the world hypertension league on noninvasive blood pressure measurement devices and blood pressure measurement in the clinical or community setting.

Authors:  Norm R C Campbell; Adel E Berbari; Lyne Cloutier; Mark Gelfer; John G Kenerson; Tej K Khalsa; Daniel T Lackland; Daniel Lemogoum; Birinder K Mangat; Sailesh Mohan; Martin G Myers; Mark L Niebylski; Eoin O'Brien; George S Stergiou; Eugenia Velludo VeIga; Xin-Hua Zhang
Journal:  J Clin Hypertens (Greenwich)       Date:  2014-05-08       Impact factor: 3.738

2.  Validation of the custo screen 400 ambulatory blood pressure-monitoring device according to the European Society of Hypertension International Protocol revision 2010.

Authors:  Peter Bramlage; Cornelia Deutsch; Ralf Krüger; Andreas Wolf; Peter Müller; Thomas Zwingers; Beate Beime; Thomas Mengden
Journal:  Vasc Health Risk Manag       Date:  2014-05-13

3.  A universal standard for the validation of blood pressure measuring devices: Association for the Advancement of Medical Instrumentation/European Society of Hypertension/International Organization for Standardization (AAMI/ESH/ISO) Collaboration Statement.

Authors:  George S Stergiou; Bruce Alpert; Stephan Mieke; Roland Asmar; Neil Atkins; Siegfried Eckert; Gerhard Frick; Bruce Friedman; Thomas Graßl; Tsutomu Ichikawa; John P Ioannidis; Peter Lacy; Richard McManus; Alan Murray; Martin Myers; Paolo Palatini; Gianfranco Parati; David Quinn; Josh Sarkis; Andrew Shennan; Takashi Usuda; Jiguang Wang; Colin O Wu; Eoin O'Brien
Journal:  J Hypertens       Date:  2018-03       Impact factor: 4.844

4.  Validation of the Omron HEM-7201 upper arm blood pressure monitor, for self-measurement in a high-altitude environment, according to the European Society of Hypertension International Protocol revision 2010.

Authors:  K Cho; M Tian; Y Lan; X Zhao; L L Yan
Journal:  J Hum Hypertens       Date:  2013-03-07       Impact factor: 3.012

5.  The prevalences of impaired fasting glucose and diabetes mellitus in working age men of North China: Anshan Worker Health Survey.

Authors:  Lei Liu; Chuang Zhou; Hang Du; Kai Zhang; Desheng Huang; Jingyang Wu
Journal:  Sci Rep       Date:  2014-04-29       Impact factor: 4.379

6.  Quality Control and Validation of Oscillometric Blood Pressure Measurements Taken During an Epidemiological Investigation.

Authors:  Xiaoqin Cao; Chunhua Song; Leilei Guo; Junxia Yang; Songyuan Deng; Yajuan Xu; Xiaolin Chen; Wolde Bekalo Sapa; Kaijuan Wang
Journal:  Medicine (Baltimore)       Date:  2015-09       Impact factor: 1.817

  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.