Peter Bramlage1, Cornelia Deutsch1, Ralf Krüger1, Andreas Wolf2, Peter Müller2, Thomas Zwingers3, Beate Beime1, Thomas Mengden4.
Abstract
OBJECTIVE: The aim of the present study was to validate the custo screen 400 ambulatory blood pressure-monitoring (ABPM) device according to the 2010 International Protocol revision of the European Society of Hypertension (ESH-IP). The device can be used for ABPM for up to 72 hours.Entities:
Keywords: ESH; ambulatory blood pressure monitoring; validation
Mesh:
Year: 2014 PMID: 24868162 PMCID: PMC4027883 DOI: 10.2147/VHRM.S63602
Source DB: PubMed Journal: Vasc Health Risk Manag ISSN: 1176-6344
Figure 1Custo screen 400 test device.
Screening and recruitment details
| Total screened | 37 | ||
| Total excluded | 4 | ||
| Ranges complete | 0 | ||
| Range adjustment | 2 | ||
| Arrhythmias | 0 | ||
| Device failure | 0 | ||
| Poor-quality sounds | 0 | ||
| Cuff size unavailable | 0 | ||
| Observer disagreement | 0 | ||
| Distribution | 2 | ||
| Other reasons | 0 | ||
| Total recruited | 33 | ||
|
| |||
| SBP | |||
| Low | <90 | 0 | 4 |
| Medium | 130–160 | 10 | 5 |
| High | >180 | 1 | 8 |
| DBP | |||
| Low | <40 | 0 | 5 |
| Medium | 80–100 | 11 | 5 |
| High | >130 | 0 | 7 |
Abbreviations: SBP, systolic blood pressure; DBP, diastolic blood pressure; Rx, antihypertensive drug prescription.
Subject characteristics at baseline
| Mean | SD | Min | Max | |
|---|---|---|---|---|
| Sex | ||||
| Male:female | 13:20 | |||
| Age (years) | 57.4 | 13 | 25 | 87 |
| Arm circumference (cm) | 28.9 | 2.9 | 24 | 36 |
| Cuff size for test device | ||||
| Standard (24–32 cm), n (%) | 31 (93.9) | |||
| Large (32–40 cm), n (%) | 2 (6.1) | |||
| SBP (mmHg) | 143.9 | 28.1 | 106 | 218 |
| DBP (mmHg) | 87.9 | 17.9 | 49 | 123 |
Abbreviations: SBP, systolic blood pressure; DBP, diastolic blood pressure; SD, standard deviation.
Observer measurements for subjects within each BP range
| SBP (mmHg) | DBP (mmHg) | |
|---|---|---|
| Overall range (low–high) | 100–220 | 49–123 |
| Low (SBP <130, DBP <80) | 43 | 40 |
| Medium (SBP 130–160, DBP 80–100) | 30 | 34 |
| High (SBP >160, DBP >100) | 26 | 25 |
| Maximum difference | 17 | 15 |
Abbreviations: SBP, systolic blood pressure; DBP, diastolic blood pressure.
Observer differences
| SBP (mmHg) | DBP (mmHg) | Repeated measurements | |
|---|---|---|---|
| Observer 2 – Observer 1 | |||
| Range | −2 to 4 | −4 to 4 | NA |
| Mean (SD) | 0.4 (1.4) | −0.4 (1.5) | NA |
Abbreviations: SBP, systolic blood pressure; DBP, diastolic blood pressure; SD, standard deviation; NA, not applicable.
Validation results
| Part 1 | ≤5 mmHg | ≤10 mmHg | ≤15 mmHg | Grade 1 | Mean (mmHg) | SD (mmHg) |
|---|---|---|---|---|---|---|
| Pass requirements | ||||||
| Two of | 73 | 87 | 96 | |||
| All of | 65 | 81 | 93 | |||
| Achieved | ||||||
| SBP | 84 | 98 | 99 | Passed | −0.5 | 4.5 |
| DBP | 93 | 99 | 99 | Passed | −0.1 | 3.3 |
|
| ||||||
| Pass requirements | ≥24 | ≤3 | ||||
| Achieved | ||||||
| SBP | 30 | 0 | Passed | |||
| DBP | 32 | 1 | Passed | |||
| Result | ||||||
|
| ||||||
Abbreviations: SBP, systolic blood pressure; DBP, diastolic blood pressure; SD, standard deviation.
Figure 2Bland–Altman plots demonstrating the systolic blood pressure (SBP) differences between the custo screen 400 device and reference sphygmomanometer and the average of device and observer pressure values.
Figure 3Bland–Altman plots demonstrating the diastolic blood pressure (DBP) differences between the custo screen 400 device and reference sphygmomanometer and the average of device and observer pressure values.