Huanfen Yao1, Angela J Shum, Melissa Cowan, Ilkka Lähdesmäki, Babak A Parviz.
Abstract
We report the design, construction, and testing of a contact lens with an integrated amperometric glucose sensor, proposing the possibility of in situ human health monitoring simply by wearing a contact lens. The glucose sensor was constructed by creating microstructures on a polymer substrate, which was subsequently shaped into a contact lens. Titania sol-gel film was applied to immobilize glucose oxidase, and Nafion® was used to decrease several potential interferences (ascorbic acid, lactate, and urea) present in the tear film. The sensor exhibits a fast response (20s), a high sensitivity (240 μA cm(-2) mM(-1)) and a good reproducibility after testing a number of sensors. It shows good linearity for the typical range of glucose concentrations in the tear film (0.1-0.6 mM), and acceptable accuracy in the presence of interfering agents. The sensor can attain a minimum detection of less than 0.01 mM glucose. Published by Elsevier B.V.Entities:
Mesh:
Substances:
Year: 2010 PMID: 21257302 PMCID: PMC3043144 DOI: 10.1016/j.bios.2010.12.042
Source DB: PubMed Journal: Biosens Bioelectron ISSN: 0956-5663 Impact factor: 10.618