An in vitro evaluation of electromagnetic interference between implantable cardiac devices and ophthalmic laser systems.

Neal A Sher1, Mikhail P Golben, Kathleen Kresge, Larry Selznick, Selcuk Adabag.   

Abstract

AIMS: Ophthalmic laser treatments are discouraged in patients with implantable pulse generators (IPGs, pacemakers) and implantable cardioverter defibrillators (ICD) due to potential effects of the electromagnetic interference (EMI) emitted by ophthalmic laser systems. We assessed the effects of EMI generated by ophthalmic laser systems and laser discharge on IPG and ICD function. METHODS AND
RESULTS: Two implantable devices, one Victory dual-chamber IPG (St Jude Medical, Minneapolis, MN, USA) and one Atlas II + dual-chamber ICD (St Jude Medical), were consecutively placed in a simulated thoracic chamber and exposed to three ophthalmic laser systems: the VISX Star S4 Excimer Laser, Lumenis Selecta II 532 neodymium-doped yttrium aluminium garnet (Nd:YAG) laser, and Ellex Ultra Q 1064 nm Nd:YAG laser. For each laser system, the apparatus was placed in the relative position of a patient while common laser procedures were delivered to a plastic object. Device pacing parameters were programmed to the highest possible sensitivity settings. The pacing and defibrillation function of the implantable devices, including electrograms, were continuously monitored. The EMI emitted from ophthalmic lasers did not lead to oversensing, inappropriate therapy, or change in the programming of the implantable cardiac devices. Manufacturing electrical tests performed on both devices showed that the cardiac devices continued to meet all the specifications for proper device function.
CONCLUSION: The St Jude Medical Victory IPG and Atlas II + ICD were not affected by the EMI emitted by the ophthalmic laser systems.

Entities:  

Mesh:

Year:  2011        PMID: 21252191     DOI: 10.1093/europace/euq495

Source DB:  PubMed          Journal:  Europace        ISSN: 1099-5129            Impact factor:   5.214


  5 in total

Review 1.  Management of cardiac implantable electronic devices during interventional pulmonology procedures.

Authors:  Abhishek Kumar; Samjot Singh Dhillon; Spandan Patel; Matthias Grube; Amit Noheria
Journal:  J Thorac Dis       Date:  2017-09       Impact factor: 2.895

Review 2.  Management of Patients With Cardiovascular Implantable Electronic Devices in Dental, Oral, and Maxillofacial Surgery.

Authors:  James Tom
Journal:  Anesth Prog       Date:  2016

3.  Electromagnetic interference from lasers and intense light sources in the treatment of patients with artificial pacemakers and other implantable cardiac devices.

Authors:  Tom Lister; Lindsay Grant; Siu-Man Lee; Richard P Cole; Anthony Jones; Timothy Taylor; Angela Mayo; Philip A Wright
Journal:  Lasers Med Sci       Date:  2013-10-27       Impact factor: 3.161

4.  Safety of the Excimer Laser in LASIK and PRK for Patients with Implantable Cardiac Devices: Our Clinical Experience in the Past Two Decades.

Authors:  Tirth J Shah; Majid Moshirfar; Phillip C Hoopes
Journal:  J Ophthalmic Vis Res       Date:  2019-10-24

5.  Implantable cardioverter defibrillator during laser transurethral resection of the prostate.

Authors:  A F Deroee; B J Cohen; J F O'Hara
Journal:  Heart Lung Vessel       Date:  2014
  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.