Literature DB >> 21219621

Revision of tibial TKA components: bone loss is independent of cementing type and technique: an in vitro cadaver study.

Turgay Efe1, Jens Figiel, David Sibbert, Susanne Fuchs-Winkelmann, Carsten O Tibesku, Nina Timmesfeld, Jürgen R J R Paletta, Adrian Skwara.   

Abstract

BACKGROUND: Different bone cements and various cementation techniques can lead to different bone loss in revision surgery. We investigated the degree of tibial bone loss depending on different cements and techniques.
METHODS: 30 tibia specimens were matched into three groups (10 each). In all cases Genesis II tibia component were implanted. In two groups, the tibia base plate alone was cemented with Palacos® R+G and Refobacin® Bone Cement R. In the third group, both tibial base plate and tibial stem were cemented with Palacos® R+G. Afterwards, the specimens were axial loaded with 2000 N for 10,000 cycles. Tibial components were explanted and the required time to explantation was recorded. Bone loss after explantation was measured by CT.
RESULTS: On CT, there was no significant difference in bone loss between cementing techniques (p = 0.077; 95% CI -1.14 - 21.03) or the cements themselves (p = 0.345; 95% CI -6.05 - 16.70). The required time to explantation was 170.6 ± 54.89, 228.7 ± 84.5, and 145.7 ± 73.0 seconds in the first, second, and third groups, respectively.
CONCLUSIONS: Cement technique and type do not influence tibial bone loss in simulated revision surgery of the tibial component in knee arthroplasty.

Entities:  

Mesh:

Substances:

Year:  2011        PMID: 21219621      PMCID: PMC3224366          DOI: 10.1186/1471-2474-12-6

Source DB:  PubMed          Journal:  BMC Musculoskelet Disord        ISSN: 1471-2474            Impact factor:   2.362


  24 in total

1.  The Genesis total knee prosthesis: a 10-year followup study.

Authors:  R S Laskin
Journal:  Clin Orthop Relat Res       Date:  2001-07       Impact factor: 4.176

2.  Tibial component fixation with cement: full- versus surface-cementation techniques.

Authors:  Christopher L Peters; Marcis A Craig; R Alexander Mohr; Kent N Bachus
Journal:  Clin Orthop Relat Res       Date:  2003-04       Impact factor: 4.176

3.  Fixation of the tibial component using CMW-1 or Palacos bone cement with gentamicin: similar outcome in a randomized radiostereometric study of 51 total knee arthroplasties.

Authors:  Gunnar Adalberth; Kjell G Nilsson; Johan Kärrholm; Helen Hassander
Journal:  Acta Orthop Scand       Date:  2002-10

4.  The modular segmental kinematic rotating hinge for nonneoplastic limb salvage.

Authors:  Bryan D Springer; Franklin H Sim; Arlen D Hanssen; David G Lewallen
Journal:  Clin Orthop Relat Res       Date:  2004-04       Impact factor: 4.176

Review 5.  Survival and clinical function of cemented and uncemented prostheses in total knee replacement: a meta-analysis.

Authors:  R Gandhi; D Tsvetkov; J R Davey; N N Mahomed
Journal:  J Bone Joint Surg Br       Date:  2009-07

6.  The mechanics of the knee joint in relation to normal walking.

Authors:  J B Morrison
Journal:  J Biomech       Date:  1970-01       Impact factor: 2.712

7.  Radiolucency at the bone-cement interface in total knee replacement. The effects of bone-surface preparation and cement technique.

Authors:  M A Ritter; S A Herbst; E M Keating; P M Faris
Journal:  J Bone Joint Surg Am       Date:  1994-01       Impact factor: 5.284

Review 8.  [Bone cements based on polymethylmethacrylate].

Authors:  S J Breusch; K-D Kühn
Journal:  Orthopade       Date:  2003-01       Impact factor: 1.087

9.  The rates of osteolysis and loosening associated with a modular posterior stabilized knee replacement. Results at five to fourteen years.

Authors:  Paul F Lachiewicz; Elizabeth S Soileau
Journal:  J Bone Joint Surg Am       Date:  2004-03       Impact factor: 5.284

10.  Tensile bonding strength of the cement-prosthesis interface.

Authors:  J P Davies; W H Harris
Journal:  Orthopedics       Date:  1994-02       Impact factor: 1.390

View more
  3 in total

1.  Comparison of different cement application techniques for tibial component fixation in TKA.

Authors:  Ulf J Schlegel; Nicholas E Bishop; Klaus Püschel; Michael M Morlock; Katrin Nagel
Journal:  Int Orthop       Date:  2014-08-01       Impact factor: 3.075

2.  Validation of a measuring technique with computed tomography for cement penetration into trabecular bone underneath the tibial tray in total knee arthroplasty on a cadaver model.

Authors:  Hennie Verburg; Laurens C van de Ridder; Vincent W J Verhoeven; Peter Pilot
Journal:  BMC Med Imaging       Date:  2014-08-27       Impact factor: 1.930

3.  Accuracy of a non-invasive CT-based measuring technique for cement penetration depth in human tibial UKA.

Authors:  Christian B Scheele; Peter E Müller; Christian Schröder; Thomas Grupp; Volkmar Jansson; Matthias F Pietschmann
Journal:  BMC Med Imaging       Date:  2019-01-21       Impact factor: 1.930

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.