A 'quad-phantom' film dosimeter for use as a multi-planar verification tool for PRESAGE/optical-CT.

L Stunja1, A Thomas, J Adamovics, J Deasy, M Oldham.   

Abstract

INTRODUCTION: To develop and characterize the accuracy and reproducibility of a quad-phantom dosimeter which will serve as an independent verification tool during commissioning of a PRESAGE/optical-CT 3D dosimetry system.
METHODS: A 16cm × 12cm cylindrical quad-phantom was constructed from four pieces of solid polyurethane mimicking the PRESAGE material. Films were placed and anchored in orthogonal planes and the quad-phantom was fastened tightly together and placed in a water-filled Styrofoam container for irradiation. A simple, two-field plan consisting of 6×6cm anterior-posterior and right-lateral 6MV photon beams (400cGy) was delivered three times (fresh films inserted for each) with a Varian Clinac 600C. Image registration was performed in the Computational Environment for Radiological Research (CERR) and dose profiles and gamma analysis was performed in CERR and MATLAB. RESULTS #ENTITYSTARTX00026; DISCUSSION: Excellent reproducibility was observed during the irradiations, with ~2.3% standard deviation between all pixels. Using a 3%, 3mm gamma criteria, excellent dosimetric accuracy was observed, with 98.8% and 96.3% passing rates in the sagittal and axial planes, respectively.
CONCLUSION: The preliminary results indicate that the quad-phantom can serve as a reproducible and accurate system for high resolution dosimetry in orthogonal planes and should serve as an effective verification tool for PRESAGE/optical-CT in more challenging clinical scenarios.

Entities:  

Year:  2010        PMID: 21218141      PMCID: PMC3017397          DOI: 10.1088/1742-6596/250/1/012097

Source DB:  PubMed          Journal:  J Phys Conf Ser        ISSN: 1742-6588


  5 in total

1.  Point/counterpoint. Radiochromic film is superior to ion chamber arrays for IMRT quality assurance.

Authors:  Slobodan Devic; Malcolm R McEwen; Colin G Orton
Journal:  Med Phys       Date:  2010-03       Impact factor: 4.071

2.  Precautions and strategies in using a commercial flatbed scanner for radiochromic film dosimetry.

Authors:  L Paelinck; W De Neve; C De Wagter
Journal:  Phys Med Biol       Date:  2006-12-18       Impact factor: 3.609

3.  On dose distribution comparison.

Authors:  Steve B Jiang; Greg C Sharp; Toni Neicu; Ross I Berbeco; Stella Flampouri; Thomas Bortfeld
Journal:  Phys Med Biol       Date:  2006-01-25       Impact factor: 3.609

4.  Analysis of uncertainties in Gafchromic EBT film dosimetry of photon beams.

Authors:  Mária Martisíková; Benjamin Ackermann; Oliver Jäkel
Journal:  Phys Med Biol       Date:  2008-11-18       Impact factor: 3.609

5.  An investigation of the accuracy of an IMRT dose distribution using two- and three-dimensional dosimetry techniques.

Authors:  Mark Oldham; Harshad Sakhalkar; Pengyi Guo; John Adamovics
Journal:  Med Phys       Date:  2008-05       Impact factor: 4.071

  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.