Abstract
A simple and fast reversed phase liquid chromatographic method was developed for estimation of paclitaxel in commercially available parenteral formulation and nanoparticles. Separations were carried out using mobile phase consisting of acetonitrile and 20 mM potassium dihydrogen phosphate (45:55, v/v) on Lichrocart(®) C(18) analytical column at a flow rate of 1 ml/min and detection wavelength of 230 nm. The developed method exhibited linearity over an analytical range of 50-2000 ng/ml with regression equation, mean peak area= 137.58 concentration (ng/ml)+1765.94, (R(2) =0.9999). The method demonstrated selectivity with no interfering peaks eluting near the vicinity of drug peak. The method was found to be sensitive with detection and quantification limits of 7.57 ng/ml and 22.94 ng/ml. The method has shown consistent and good recoveries from parenteral formulation (100.06±0.86%) and nanoparticles (100.43±0.91%). The method was successfully employed for the analysis of in vitro release study samples of nanoparticle formulation. The method was also applied for determination of paclitaxel content in various pharmaceutical formulations.Entities:
Keywords: RP-HPLC; nanoparticles; paclitaxel; parenteral formulation
Year: 2010 PMID: 21218057 PMCID: PMC3013583 DOI: 10.4103/0250-474X.73914
Source DB: PubMed Journal: Indian J Pharm Sci ISSN: 0250-474X Impact factor: 0.975
Fig. 1Overlay of calibration standards (50-2000 ng/ml).
Fig. 2Chromatograms of standards (a) placebo standard for parenteral formulation, (b) placebo standard for nanoparticles, (c) sample standard (100%) for parenteral formulation, (d) sample standard (100%) for nanoparticles
CALIBRATION DATA FOR PACLITAXEL
| Concentration (ng/ml) | Average peak area | %CV | Predicted concentration (ng/ml) |
|---|---|---|---|
| 50 | 8498.00 ± 62.11 | 0.73 | 48.93 |
| 100 | 15456.67 ± 190.50 | 1.23 | 99.51 |
| 250 | 36462.67 ± 282.83 | 0.78 | 252.20 |
| 500 | 70308.89 ± 518.60 | 0.74 | 498.22 |
| 1000 | 139592.56 ± 1515.72 | 1.09 | 1001.82 |
| 2000 | 276822.67 ± 2618.24 | 0.95 | 1999.31 |
each value represents the average of nine independent determinations
RECOVERY STUDY BY PLACEBO SPIKING TECHNIQUE
| Product | Amount of drug added (% of label claim)a | Mean absolute recovery (%) | %CV | % Bias |
|---|---|---|---|---|
| Parenteral Formulation | 25 | 100.27±0.67 | 0.66 | 0.27 |
| 50 | 100.60±1.31 | 1.30 | 0.60 | |
| 100 | 99.85±0.50 | 0.50 | 0.15 | |
| 150 | 99.61±0.82 | 0.82 | 0.39 | |
| 200 | 99.96±0.57 | 0.57 | 0.04 | |
| Overall recovery = 100.06 ± 0.86 | ||||
| Nanoparticles | 25 | 100.54±0.71 | 0.71 | 0.71 |
| 50 | 100.41±1.15 | 1.15 | 1.15 | |
| 100 | 100.23±1.01 | 1.01 | 1.01 | |
| 150 | 100.24±0.82 | 0.82 | 0.82 | |
| 200 | 100.75±0.91 | 0.91 | 0.91 | |
| Overall recovery = 100.43±0.91 | ||||
RESULTS FOR REPEATABILITY AND INTERMEDIATE PRECISION
| QC Level | Repeatability (intra-batch) | Intermediate precision | ||||||
|---|---|---|---|---|---|---|---|---|
| Batch-I | Batch-II | Batch-III | (inter-batch) | |||||
| Mean | %CV | Mean | %CV | Mean | %CV | Mean | %CV | |
| LQC | 49.91 | 1.82 | 50.27 | 1.87 | 50.38 | 2.06 | 50.19 | 1.83 |
| MQC | 501.62 | 0.97 | 499.28 | 1.04 | 499.90 | 1.13 | 500.27 | 0.99 |
| HQC | 2022.48 | 0.23 | 2014.32 | 0.27 | 2011.28 | 0.16 | 2016.03 | 0.32 |
each value represents average of five independent determinations
Fig. 3In vitro release study of nanoparticles CAR is cumulative amount released, mean±SD, n=3