Literature DB >> 21216551

Local in vivo shimming using adaptive passive shim positioning.

Sejung Yang1, Hahnsung Kim, Min-Oh Ghim, Byung-Uk Lee, Dong-Hyun Kim.   

Abstract

A subject-specific local in vivo passive shimming method, focusing on the prefrontal and temporal regions, is proposed. The aim of the investigation is to show that subject variability exists in optimal passive shimming and that the proposed method can be effectively used to overcome these differences. A shimming structure capable of adjusting the position of the passive shims to within a millimeter resolution is built. The optimal shim positions for each individual subject are computed from obtained field map using a convex optimization algorithm. Passive shim experiments at predetermined fixed shim positions vs. individually adjusted shim positions were performed and compared. The results show that intersubject variability exists in the optimal shim positions and that the location-sensitive method proposed can be useful for improving main field homogeneity in vivo.
Copyright © 2011 Elsevier Inc. All rights reserved.

Mesh:

Year:  2011        PMID: 21216551     DOI: 10.1016/j.mri.2010.10.004

Source DB:  PubMed          Journal:  Magn Reson Imaging        ISSN: 0730-725X            Impact factor:   2.546


  3 in total

Review 1.  The future of acquisition speed, coverage, sensitivity, and resolution.

Authors:  Lawrence L Wald
Journal:  Neuroimage       Date:  2012-03-06       Impact factor: 6.556

2.  Automated slice-specific simultaneous z-shim method for reducing B1 inhomogeneity and susceptibility-induced signal loss with parallel transmission at 3T.

Authors:  Rainer Schneider; Fernando Boada; Jens Haueisen; Josef Pfeuffer
Journal:  Magn Reson Med       Date:  2014-10-07       Impact factor: 4.668

Review 3.  Spinal cord MRI at 7T.

Authors:  Robert L Barry; S Johanna Vannesjo; Samantha By; John C Gore; Seth A Smith
Journal:  Neuroimage       Date:  2017-07-03       Impact factor: 6.556

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.