Comparison of chemistry analytes between 2 portable, commercially available analyzers and a conventional laboratory analyzer in reptiles.

Stephanie L McCain1, Bente Flatland, Juergen P Schumacher, Elsburgh O Clarke Iii, Michael M Fry.   

Abstract

BACKGROUND: Advantages of handheld and small bench-top biochemical analyzers include requirements for smaller sample volume and practicality for use in the field or in practices, but little has been published on the performance of these instruments compared with standard reference methods in analysis of reptilian blood.
OBJECTIVE: The aim of this study was to compare reptilian blood biochemical values obtained using the Abaxis VetScan Classic bench-top analyzer and a Heska i-STAT handheld analyzer with values obtained using a Roche Hitachi 911 chemical analyzer.
METHODS: Reptiles, including 14 bearded dragons (Pogona vitticeps), 4 blue-tongued skinks (Tiliqua gigas), 8 Burmese star tortoises (Geochelone platynota), 10 Indian star tortoises (Geochelone elegans), 5 red-tailed boas (Boa constrictor), and 5 Northern pine snakes (Pituophis melanoleucus melanoleucus), were manually restrained, and a single blood sample was obtained and divided for analysis. Results for concentrations of albumin, bile acids, calcium, glucose, phosphates, potassium, sodium, total protein, and uric acid and activities of aspartate aminotransferase and creatine kinase obtained from the VetScan Classic and Hitachi 911 were compared. Results for concentrations of chloride, glucose, potassium, and sodium obtained from the i-STAT and Hitachi 911 were compared.
RESULTS: Compared with results from the Hitachi 911, those from the VetScan Classic and i-STAT had variable correlations, and constant or proportional bias was found for many analytes. Bile acid data could not be evaluated because results for 44 of 45 samples fell below the lower linearity limit of the VetScan Classic.
CONCLUSIONS: Although the 2 portable instruments might provide measurements with clinical utility, there were significant differences compared with the reference analyzer, and development of analyzer-specific reference intervals is recommended. ©2010 American Society for Veterinary Clinical Pathology.

Entities:  

Mesh:

Substances:

Year:  2010        PMID: 21198734     DOI: 10.1111/j.1939-165X.2010.00258.x

Source DB:  PubMed          Journal:  Vet Clin Pathol        ISSN: 0275-6382            Impact factor:   1.180


  8 in total

1.  Venous blood gases, plasma biochemistry, and hematology of wild-caught common chameleons (Chamaeleo chamaeleon).

Authors:  David Eshar; Melanie Ammersbach; Boaz Shacham; Gad Katzir; Hugues Beaufrère
Journal:  Can J Vet Res       Date:  2018-04       Impact factor: 1.310

2.  Reference Intervals for Plasma Biochemical Variables by Point-of-Care Testing in Captive Black-tailed Prairie Dogs (Cynomys ludovicianus).

Authors:  David Eshar; Sara M Gardhouse; Diana Schwartz; Hugues Beaufrere
Journal:  J Am Assoc Lab Anim Sci       Date:  2018-09-12       Impact factor: 1.232

3.  Reference Intervals for Selected Hematology and Clinical Chemistry Measurands in Temminck's Pangolin (Smutsia temminckii).

Authors:  Emma H Hooijberg; Karin Lourens; Leith C R Meyer
Journal:  Front Vet Sci       Date:  2021-07-08

4.  Energy Expenditure and Metabolic Changes of Free-Flying Migrating Northern Bald Ibis.

Authors:  Franz Bairlein; Johannes Fritz; Alexandra Scope; Ilse Schwendenwein; Gabriela Stanclova; Gertjan van Dijk; Harro A J Meijer; Simon Verhulst; John Dittami
Journal:  PLoS One       Date:  2015-09-16       Impact factor: 3.240

5.  Use of portable blood physiology point-of-care devices for basic and applied research on vertebrates: a review.

Authors:  Lauren J Stoot; Nicholas A Cairns; Felicia Cull; Jessica J Taylor; Jennifer D Jeffrey; Félix Morin; John W Mandelman; Timothy D Clark; Steven J Cooke
Journal:  Conserv Physiol       Date:  2014-04-04       Impact factor: 3.079

6.  Predictive Blood Chemistry Parameters for Pansteatitis-Affected Mozambique Tilapia (Oreochromis mossambicus).

Authors:  John A Bowden; Theresa M Cantu; Robert W Chapman; Stephen E Somerville; Matthew P Guillette; Hannes Botha; Andre Hoffman; Wilmien J Luus-Powell; Willem J Smit; Jeffrey Lebepe; Jan Myburgh; Danny Govender; Jonathan Tucker; Ashley S P Boggs; Louis J Guillette
Journal:  PLoS One       Date:  2016-04-26       Impact factor: 3.240

7.  Validation of the i-STAT and HemoCue systems for the analysis of blood parameters in the bar-headed goose, Anser indicus.

Authors:  T S Harter; M Reichert; C J Brauner; W K Milsom
Journal:  Conserv Physiol       Date:  2015-05-19       Impact factor: 3.079

8.  Venous blood gas in free-living eastern box turtles (Terrapene carolina carolina) and effects of physiologic, demographic and environmental factors.

Authors:  Laura Adamovicz; Katie Leister; John Byrd; Christopher A Phillips; Matthew C Allender
Journal:  Conserv Physiol       Date:  2018-07-25       Impact factor: 3.079

  8 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.