Development of a computer assisted gantry system for gaining rapid and accurate calyceal access during percutaneous nephrolithotomy.

A D Zarrabi1, J P Conradie, C F Heyns, C Scheffer, K Schreve.   

Abstract

PURPOSE: To design a simple, cost-effective system for gaining rapid and accurate calyceal access during percutaneous nephrolithotomy (PCNL).
MATERIALS AND METHODS: The design consists of a low-cost, light-weight, portable mechanical gantry with a needle guiding device. Using C-arm fluoroscopy, two images of the contrast-filled renal collecting system are obtained: at 0-degrees (perpendicular to the kidney) and 20-degrees. These images are relayed to a laptop computer containing the software and graphic user interface for selecting the targeted calyx. The software provides numerical settings for the 3 axes of the gantry, which are used to position the needle guiding device. The needle is advanced through the guide to the depth calculated by the software, thus puncturing the targeted calyx. Testing of the system was performed on 2 target types: 1) radiolucent plastic tubes the approximate size of a renal calyx (5 or 10 mm in diameter, 30 mm in length); and 2) foam-occluded, contrast-filled porcine kidneys.
RESULTS: Tests using target type 1 with 10 mm diameter (n=14) and 5 mm diameter (n=7) tubes resulted in a 100% targeting success rate, with a mean procedure duration of 10 minutes. Tests using target type 2 (n=2) were both successful, with accurate puncturing of the selected renal calyx, and a mean procedure duration of 15 minutes.
CONCLUSIONS: The mechanical gantry system described in this paper is low-cost, portable, light-weight, and simple to set up and operate. C-arm fluoroscopy is limited to two images, thus reducing radiation exposure significantly. Testing of the system showed an extremely high degree of accuracy in gaining precise access to a targeted renal calyx.

Entities:  

Mesh:

Year:  2010        PMID: 21176281     DOI: 10.1590/s1677-55382010000600013

Source DB:  PubMed          Journal:  Int Braz J Urol        ISSN: 1677-5538            Impact factor:   1.541


  4 in total

Review 1.  Vision for the future on urolithiasis: research, management, education and training-some personal views.

Authors:  A Rodgers; A Trinchieri; M H Ather; N Buchholz
Journal:  Urolithiasis       Date:  2018-10-29       Impact factor: 3.436

Review 2.  Training the resident in percutaneous nephrolithotomy.

Authors:  M Hammad Ather; Chi-Fai Ng; Gholamraza Pourmand; Palle J Osther
Journal:  Arab J Urol       Date:  2013-09-16

3.  A Novel method of ensuring safe and accurate dilatation during percutaneous nephrolithotomy.

Authors:  Tarun Javali; Amey Pathade; H K Nagaraj
Journal:  Int Braz J Urol       Date:  2015 Sep-Oct       Impact factor: 1.541

4.  Kidney access device.

Authors:  Jasneet Singh Bhullar; Robert Scott; Mitesh Patel; Vijay K Mittal
Journal:  JSLS       Date:  2014 Oct-Dec       Impact factor: 2.172

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.