Michael S Gossman1, Morris I Bank.
Abstract
Dose-volume histograms provide key information to radiation oncologists when they assess the adequacy of a patient treatment plan in radiation therapy. It is important therefore that all clinically relevant data be accurate. In this article we present the first quality assurance routine involving a direct comparison of planning system results with the results obtained from independent hand calculations. Given a known three-dimensional (3-D) structure such as a parallelepiped, a simple beam arrangement, and known physics beam data, a time-efficient and reproducible method for verifying the accuracy of volumetric statistics (DVH) from a radiation therapy treatment planning system (TPS) can be employed rapidly, satisfying the QA requirements for (TPS) commissioning, upgrades, and annual checks. Using this method, the maximum disagreement was only 1.7% for 6 MV and 1.3% for 18 MV photon energies. The average accuracy was within 0.6% for 6 MV and 0.4% for 18 MV for all depth-dose results. A 2% disagreement was observed with the treatment planning system DVH from defined volume comparison to the known structure dimensions.Entities:
Keywords: Dose; Volume; dose–volume; dose–volume histogram; histogram; quality assurance; treatment planning system
Year: 2010 PMID: 21170183 PMCID: PMC2990113 DOI: 10.4103/0971-6203.71759
Source DB: PubMed Journal: J Med Phys ISSN: 0971-6203
Figure 1Axial view isodose distribution indicating the position of the rectangular volume within the water phantom. The isodose levels represent the percentage of dose, relative to 100 cGy at 10 cm phantom depth (5 cm structure depth), for an 18 MV beam having normal geometry and a field size of 30 × 30 cm2 at the SAD.
Figure 2As in the previous illustration [Figure 1], this is a zoom-in on the distribution of dose that is seen inside the structure in the axial view
Figure 3Dose-volume histogram statistical analysis from the treatment planning system for 6 MV (dashed) and 18 MV (solid)
Spreadsheet of independent dose–volume calculation correlation to treatment planning system DVH results. The independent calculations appear on the left of the spreadsheet, with DVH results on the right along with their overall agreement. Both 6 MV and 18 MV results are presented.
| 6 MV | ||||||||||||||||
| 6 | 30.0 | 28.8 | 0.917 | 1.001 | 1.047 | 1.025 | 1.085 | 112.2 | 119.8 | 10 | 10.01 | 119.8 | 120.5 | 9.9 | -1.7 | |
| 7 | 30.0 | 29.1 | 0.895 | 1.001 | 1.047 | 1.025 | 1.063 | 112.2 | 114.5 | 20 | 20.02 | 114.5 | 115.1 | 20.0 | -0.8 | |
| 8 | 30.0 | 29.4 | 0.873 | 1.001 | 1.047 | 1.025 | 1.041 | 112.2 | 109.4 | 30 | 30.04 | 109.4 | 110.1 | 29.9 | -0.9 | |
| 9 | 30.0 | 29.7 | 0.852 | 1.001 | 1.047 | 1.026 | 1.020 | 112.2 | 104.7 | 40 | 40.05 | 104.7 | 105.0 | 39.9 | -0.6 | |
| 10 | 30.0 | 30.0 | 0.830 | 1.001 | 1.047 | 1.026 | 1.000 | 112.2 | 100.0 | 50 | 50.06 | 100.0 | 100.2 | 50.0 | -0.4 | |
| 11 | 30.0 | 30.3 | 0.807 | 1.001 | 1.047 | 1.026 | 0.980 | 112.2 | 95.3 | 60 | 60.07 | 95.3 | 95.7 | 60.0 | -0.5 | |
| 12 | 30.0 | 30.6 | 0.785 | 1.001 | 1.047 | 1.026 | 0.961 | 112.2 | 90.9 | 70 | 70.08 | 90.9 | 91.1 | 70.0 | -0.4 | |
| 13 | 30.0 | 30.9 | 0.763 | 1.001 | 1.047 | 1.026 | 0.943 | 112.2 | 86.7 | 80 | 80.10 | 86.7 | 86.9 | 80.0 | -0.4 | |
| 14 | 30.0 | 31.2 | 0.741 | 1.001 | 1.047 | 1.026 | 0.925 | 112.2 | 82.5 | 90 | 90.11 | 82.5 | 82.8 | 90.0 | -0.5 | |
| 15 | 30.0 | 31.5 | 0.721 | 1.001 | 1.047 | 1.027 | 0.907 | 112.2 | 78.9 | 100 | 100.12 | 78.9 | 78.8 | 100.0 | -0.0 | |
| 18 MV | ||||||||||||||||
| 6 | 30.0 | 28.8 | 0.965 | 1.000 | 1.080 | 1.008 | 1.085 | 102.2 | 116.3 | 10 | 10.00 | 116.3 | 117.1 | 9.9 | -1.3 | |
| 7 | 30.0 | 29.1 | 0.949 | 1.000 | 1.080 | 1.008 | 1.063 | 102.2 | 112.0 | 20 | 20.00 | 112.0 | 112.6 | 20.1 | -0.2 | |
| 8 | 30.0 | 29.4 | 0.933 | 1.000 | 1.080 | 1.008 | 1.041 | 102.2 | 107.9 | 30 | 30.00 | 107.9 | 108.3 | 30.0 | -0.4 | |
| 9 | 30.0 | 29.7 | 0.918 | 1.000 | 1.080 | 1.009 | 1.020 | 102.2 | 104.1 | 40 | 40.00 | 104.1 | 104.2 | 40.0 | -0.2 | |
| 10 | 30.0 | 30.0 | 0.898 | 1.000 | 1.080 | 1.009 | 1.000 | 102.2 | 100.0 | 50 | 50.00 | 100.0 | 100.3 | 50.1 | -0.2 | |
| 11 | 30.0 | 30.3 | 0.882 | 1.000 | 1.080 | 1.009 | 0.980 | 102.2 | 96.1 | 60 | 60.00 | 96.1 | 96.6 | 60.0 | -0.4 | |
| 12 | 30.0 | 30.6 | 0.866 | 1.000 | 1.080 | 1.010 | 0.961 | 102.2 | 92.6 | 70 | 70.00 | 92.6 | 93.3 | 69.9 | -0.5 | |
| 13 | 30.0 | 30.9 | 0.849 | 1.000 | 1.080 | 1.010 | 0.943 | 102.2 | 89.0 | 80 | 80.00 | 89.0 | 89.5 | 80.0 | -0.5 | |
| 14 | 30.0 | 31.2 | 0.833 | 1.000 | 1.080 | 1.010 | 0.925 | 102.2 | 85.7 | 90 | 90.00 | 85.7 | 86.1 | 90.0 | -0.5 | |
| 15 | 30.0 | 31.5 | 0.818 | 1.000 | 1.080 | 1.011 | 0.907 | 102.2 | 82.6 | 100 | 100.00 | 82.6 | 82.7 | 99.9 | -0.2 | |