Assessment of flow waves and colorimetric CO2 detector for endotracheal tube placement during neonatal resuscitation.

Georg M Schmölzer1, David A Poulton, Jennifer A Dawson, C Omar F Kamlin, Colin J Morley, Peter G Davis.   

Abstract

AIM: Clinical assessment and end-tidal CO(2) (ETCO(2)) detectors are routinely used to verify endotracheal tube (ETT) placement. However, ETCO(2) detectors may mislead clinicians by failing to identify correct placement under a variety of conditions. A flow sensor measures gas flow in and out of an ETT. We reviewed video recordings of neonatal resuscitations to compare a colorimetric CO(2) detector (Pedi-Cap®) with flow sensor recordings for assessing ETT placement.
METHODS: We reviewed recordings of infants <32 weeks gestation born between February 2007 and January 2010. Airway pressures and gas flow were recorded with a respiratory function monitor. Video recording were used (i) to identify infants who were intubated in the delivery room and (ii) to observe colour change of the ETCO(2) detector. Flow sensor recordings were used to confirm whether the tube was in the trachea or not.
RESULTS: Of the 210 infants recorded, 44 infants were intubated in the delivery room. Data from 77 intubation attempts were analysed. In 35 intubations of 20 infants both a PediCap® and flow sensor were available for analysis. In 21 (60%) intubations, both methods correctly identified successful ETT placement and in 3 (9%) both indicated the ETT was not in the trachea. In the remaining 11 (31%) intubations the PediCap® failed to change colour despite the flow wave indicating correct ETT placement.
CONCLUSION: Colorimetric CO(2) detectors may mislead clinicians intubating very preterm infants in the delivery room. They may fail to change colour in spite of correct tube placement in up to one third of the cases.
Copyright © 2010 Elsevier Ireland Ltd. All rights reserved.

Entities:  

Mesh:

Substances:

Year:  2010        PMID: 21167628     DOI: 10.1016/j.resuscitation.2010.11.008

Source DB:  PubMed          Journal:  Resuscitation        ISSN: 0300-9572            Impact factor:   5.262


  11 in total

1.  A novel airway device with tactile sensing capabilities for verifying correct endotracheal tube placement.

Authors:  Pauwel Goethals; Harshu Chaobal; Dominiek Reynaerts; David Schaner
Journal:  J Clin Monit Comput       Date:  2013-09-25       Impact factor: 2.502

2.  Efficacy and user preference of two CO2 detectors in an infant mannequin randomized crossover trial.

Authors:  G A Hawkes; B J O'Connell; V Livingstone; C P Hawkes; C A Ryan; E M Dempsey
Journal:  Eur J Pediatr       Date:  2013-06-12       Impact factor: 3.183

3.  Assessment of endotracheal tube placement in newborn infants: a randomized controlled trial.

Authors:  S van Os; P-Y Cheung; K Kushniruk; M O'Reilly; K Aziz; G M Schmölzer
Journal:  J Perinatol       Date:  2016-01-14       Impact factor: 2.521

Review 4.  WITHDRAWN: Techniques to ascertain correct endotracheal tube placement in neonates.

Authors:  Georg M Schmölzer; Charles Christoph C Roehr
Journal:  Cochrane Database Syst Rev       Date:  2018-07-05

5.  A novel training simulator for portable ultrasound identification of incorrect newborn endotracheal tube placement - observational diagnostic accuracy study protocol.

Authors:  Hasan S Merali; Mark O Tessaro; Khushboo Q Ali; Shaun K Morris; Sajid B Soofi; Shabina Ariff
Journal:  BMC Pediatr       Date:  2019-11-13       Impact factor: 2.125

6.  Noninvasive Monitoring during Interhospital Transport of Newborn Infants.

Authors:  Georg M Schmölzer; Megan O'Reilly; Po-Yin Cheung
Journal:  Crit Care Res Pract       Date:  2013-02-24

7.  Short- and intermediate-term outcomes of preterm infants receiving positive pressure ventilation in the delivery room.

Authors:  Megan O'Reilly; Po-Yin Cheung; Khalid Aziz; Georg M Schmölzer
Journal:  Crit Care Res Pract       Date:  2013-01-17

Review 8.  Enhanced Monitoring of the Preterm Infant during Stabilization in the Delivery Room.

Authors:  Daragh Finn; Geraldine B Boylan; C Anthony Ryan; Eugene M Dempsey
Journal:  Front Pediatr       Date:  2016-03-31       Impact factor: 3.418

Review 9.  Non-invasive carbon dioxide monitoring in neonates: methods, benefits, and pitfalls.

Authors:  Deepika Sankaran; Lida Zeinali; Sameeia Iqbal; Praveen Chandrasekharan; Satyan Lakshminrusimha
Journal:  J Perinatol       Date:  2021-06-19       Impact factor: 2.521

10.  Canadian Airway Focus Group updated consensus-based recommendations for management of the difficult airway: part 1. Difficult airway management encountered in an unconscious patient.

Authors:  J Adam Law; Laura V Duggan; Mathieu Asselin; Paul Baker; Edward Crosby; Andrew Downey; Orlando R Hung; Philip M Jones; François Lemay; Rudiger Noppens; Matteo Parotto; Roanne Preston; Nick Sowers; Kathryn Sparrow; Timothy P Turkstra; David T Wong; George Kovacs
Journal:  Can J Anaesth       Date:  2021-06-18       Impact factor: 5.063

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.