Assessment and monitoring tumor vascularity with contrast-enhanced ultrasound maximum intensity persistence imaging.

Marybeth A Pysz1, Kira Foygel, Cedric M Panje, Andrew Needles, Lu Tian, Jürgen K Willmann.   

Abstract

OBJECTIVES: Contrast-enhanced ultrasound imaging is increasingly being used in the clinic for assessment of tissue vascularity. The purpose of our study was to evaluate the effect of different contrast administration parameters on the in vivo ultrasound imaging signal in tumor-bearing mice using a maximum intensity persistence (MIP) algorithm and to evaluate the reliability of in vivo MIP imaging in assessing tumor vascularity. The potential of in vivo MIP imaging for monitoring tumor vascularity during antiangiogenic cancer treatment was further evaluated.
MATERIALS AND METHODS: In intraindividual experiments, varying contrast microbubble concentrations (5 × 10⁵, 5 × 10⁶, 5 × 10⁷, 5 × 10⁸ microbubbles in 100 μL saline) and contrast injection rates (0.6, 1.2, and 2.4 mL/min) in subcutaneous tumor-bearing mice were applied and their effects on in vivo contrast-enhanced ultrasound MIP imaging plateau values were obtained using a dedicated small animal ultrasound imaging system (40 MHz). Reliability of MIP ultrasound imaging was tested following 2 injections of the same microbubble concentration (5 × 10⁷ microbubbles at 1.2 mL/min) in the same tumors. In mice with subcutaneous human colon cancer xenografts, longitudinal contrast-enhanced ultrasound MIP imaging plateau values (baseline and at 48 hours) were compared between mice with and without antiangiogenic treatment (antivascular endothelial growth factor antibody). Ex vivo CD31 immunostaining of tumor tissue was used to correlate in vivo MIP imaging plateau values with microvessel density analysis.
RESULTS: In vivo MIP imaging plateau values correlated significantly (P = 0.001) with contrast microbubble doses. At 3 different injection rates of 0.6, 1.2, and 2.4 mL/min, MIP imaging plateau values did not change significantly (P = 0.61). Following 2 injections with the same microbubble dose and injection rate, MIP imaging plateau values were obtained with high reliability with an intraclass correlation coefficient of 0.82 (95% confidence interval: 0.64, 0.94). In addition, in vivo MIP imaging plateau values significantly correlated (P = 0.01; R² = 0.77) with ex vivo microvessel density analysis. Tumor volumes in treated and nontreated mice did not change significantly (P = 0.22) within 48 hours. In contrast, the change of in vivo MIP imaging plateau values from baseline to 48 hours was significantly different (P = 0.01) in treated versus nontreated mice.
CONCLUSIONS: Contrast-enhanced ultrasound MIP imaging allows reliable assessment of tumor vascularity and monitoring of antiangiogenic cancer therapy in vivo, provided that a constant microbubble dose is administered.

Entities:  

Mesh:

Substances:

Year:  2011        PMID: 21150790      PMCID: PMC4457398          DOI: 10.1097/RLI.0b013e3181f9202d

Source DB:  PubMed          Journal:  Invest Radiol        ISSN: 0020-9996            Impact factor:   6.016


  32 in total

Review 1.  Normalizing tumor vasculature with anti-angiogenic therapy: a new paradigm for combination therapy.

Authors:  R K Jain
Journal:  Nat Med       Date:  2001-09       Impact factor: 53.440

2.  Structure-function studies of two synthetic anti-vascular endothelial growth factor Fabs and comparison with the Avastin Fab.

Authors:  Germaine Fuh; Ping Wu; Wei-Ching Liang; Mark Ultsch; Chingwei V Lee; Barbara Moffat; Christian Wiesmann
Journal:  J Biol Chem       Date:  2005-12-22       Impact factor: 5.157

3.  Early quantitative evaluation of a tumor vasculature disruptive agent AVE8062 using dynamic contrast-enhanced ultrasonography.

Authors:  Sonia Lavisse; Pascale Lejeune; Valérie Rouffiac; Nicolas Elie; Estelle Bribes; Brigitte Demers; Patricia Vrignaud; Marie-Christine Bissery; Aude Brulé; Serge Koscielny; Pierre Péronneau; Nathalie Lassau
Journal:  Invest Radiol       Date:  2008-02       Impact factor: 6.016

Review 4.  Dynamic contrast-enhanced ultrasonography (DCE-US) with quantification of tumor perfusion: a new diagnostic tool to evaluate the early effects of antiangiogenic treatment.

Authors:  Nathalie Lassau; Linda Chami; Baya Benatsou; Pierre Peronneau; Alain Roche
Journal:  Eur Radiol       Date:  2007-12       Impact factor: 5.315

Review 5.  Tumour angiogenesis and its relation to contrast enhancement on computed tomography: a review.

Authors:  K A Miles
Journal:  Eur J Radiol       Date:  1999-06       Impact factor: 3.528

6.  Quantification of myocardial blood flow with ultrasound-induced destruction of microbubbles administered as a constant venous infusion.

Authors:  K Wei; A R Jayaweera; S Firoozan; A Linka; D M Skyba; S Kaul
Journal:  Circulation       Date:  1998-02-10       Impact factor: 29.690

7.  Comparison of conventional time-intensity curves vs. maximum intensity over time for post-processing of dynamic contrast-enhanced ultrasound.

Authors:  Moritz Palmowski; Wiltrud Lederle; Jessica Gaetjens; Michaela Socher; Peter Hauff; Jessica Bzyl; Wolfhard Semmler; Rolf W Günther; Fabian Kiessling
Journal:  Eur J Radiol       Date:  2009-11-27       Impact factor: 3.528

Review 8.  Non-invasive assessment of tumor neovasculature: techniques and clinical applications.

Authors:  Rodolfo Perini; Regine Choe; Arjun G Yodh; Chandra Sehgal; Chaitanya R Divgi; Mark A Rosen
Journal:  Cancer Metastasis Rev       Date:  2008-12       Impact factor: 9.264

9.  Chapter 14. Measuring intratumoral microvessel density.

Authors:  Noel Weidner
Journal:  Methods Enzymol       Date:  2008       Impact factor: 1.600

10.  Dose-response relationship of ultrasound contrast agent in an in vivo murine melanoma model.

Authors:  Gabriela S Seiler; Lisa S Ziemer; Susan Schultz; William M F Lee; Chandra M Sehgal
Journal:  Cancer Imaging       Date:  2007-12-17       Impact factor: 3.909

View more
  25 in total

1.  Quantitative assessment of tumor angiogenesis using real-time motion-compensated contrast-enhanced ultrasound imaging.

Authors:  Marybeth A Pysz; Ismayil Guracar; Kira Foygel; Lu Tian; Jürgen K Willmann
Journal:  Angiogenesis       Date:  2012-04-26       Impact factor: 9.596

2.  Role of contrast-enhanced ultrasound to define prognosis and predict response to biotherapy in pancreatic neuroendocrine tumors.

Authors:  M Del Prete; A Di Sarno; R Modica; F Lassandro; A Giorgio; A Bianco; M Muto; M Gasperi; F Del Prete; A Colao; V Montesarchio; A Faggiano
Journal:  J Endocrinol Invest       Date:  2017-06-30       Impact factor: 4.256

3.  Intra-Animal Comparison between Three-dimensional Molecularly Targeted US and Three-dimensional Dynamic Contrast-enhanced US for Early Antiangiogenic Treatment Assessment in Colon Cancer.

Authors:  Huaijun Wang; Amelie M Lutz; Dimitre Hristov; Lu Tian; Jürgen K Willmann
Journal:  Radiology       Date:  2016-08-04       Impact factor: 11.105

4.  Quantitative mapping of tumor vascularity using volumetric contrast-enhanced ultrasound.

Authors:  Kenneth Hoyt; Anna Sorace; Reshu Saini
Journal:  Invest Radiol       Date:  2012-03       Impact factor: 6.016

5.  2-tier in-plane motion correction and out-of-plane motion filtering for contrast-enhanced ultrasound.

Authors:  Casey N Ta; Mohammad Eghtedari; Robert F Mattrey; Yuko Kono; Andrew C Kummel
Journal:  Invest Radiol       Date:  2014-11       Impact factor: 6.016

6.  Unenhanced areas revealed by contrast-enhanced abdominal ultrasonography with Sonazoid™ potentially correspond to colorectal cancer.

Authors:  Minoru Tomizawa; Mizuki Togashi; Fuminobu Shinozaki; Rumiko Hasegawa; Yoshinori Shirai; Midori Noritake; Yukie Matsuoka; Hiroaki Kainuma; Yasuji Iwasaki; Kazunori Fugo; Yasufumi Motoyoshi; Takao Sugiyama; Shigenori Yamamoto; Takashi Kishimoto; Naoki Ishige
Journal:  Exp Ther Med       Date:  2016-11-03       Impact factor: 2.447

7.  Three-dimensional Dynamic Contrast-enhanced US Imaging for Early Antiangiogenic Treatment Assessment in a Mouse Colon Cancer Model.

Authors:  Huaijun Wang; Dimitre Hristov; Jiale Qin; Lu Tian; Jürgen K Willmann
Journal:  Radiology       Date:  2015-05-22       Impact factor: 11.105

Review 8.  Targeted contrast-enhanced ultrasound: an emerging technology in abdominal and pelvic imaging.

Authors:  Marybeth A Pysz; Jürgen K Willmann
Journal:  Gastroenterology       Date:  2011-01-18       Impact factor: 22.682

9.  Automating tumor classification with pixel-by-pixel contrast-enhanced ultrasound perfusion kinetics.

Authors:  Casey N Ta; Yuko Kono; Christopher V Barback; Robert F Mattrey; Andrew C Kummel
Journal:  J Vac Sci Technol B Nanotechnol Microelectron       Date:  2012-03-22

10.  Molecular Acoustic Angiography: A New Technique for High-resolution Superharmonic Ultrasound Molecular Imaging.

Authors:  Sarah E Shelton; Brooks D Lindsey; James K Tsuruta; F Stuart Foster; Paul A Dayton
Journal:  Ultrasound Med Biol       Date:  2015-12-08       Impact factor: 2.998

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.