Literature DB >> 21132561

Dynamic behavior of BIS, M-entropy and neuroSENSE brain function monitors.

Stéphane Bibian1, Guy A Dumont, Tatjana Zikov.   

Abstract

OBJECTIVES: The objective of this paper is to assess the suitability of brain function monitors for use in closed-loop anesthesia or sedation delivery. In such systems, monitors used as feedback sensors should preferably be Linear and Time Invariant (LTI) in order to limit sensor-induced uncertainty which can cause degraded performance. In this paper, we evaluate the suitability of the BIS A2000 (Aspect Medical Systems, MA), the M-Entropy Monitor (GE HealthCare), and the NeuroSENSE Monitor (NeuroWave Systems Inc, OH), by verifying whether their dynamic behavior conforms to the LTI hypothesis.
METHODS: We subjected each monitor to two different composite EEG signals containing step-wise changes in cortical activity. The first signal was used to identify Linear Time-Invariant (LTI) models that mathematically capture the dynamic behavior of each monitor. The identification of the model parameters was carried out using standard Recursive Least Squares (RLS) estimation. The second signal was used to assess the performance of the model, by comparing the output of the monitor to the simulated output predicted by the model.
RESULTS: While a LTI model was successfully derived for each monitor using the first signal, only the model derived for NeuroSENSE was capable to reliably predict the monitor output for the second input signals. This indicates that some algorithmic processes within the BIS A2000 and M-Entropy are non-linear and/or time variant.
CONCLUSION: While both BIS and M-Entropy monitors have been successfully used in closed-loop systems, we were unable to obtain a unique LTI model that could capture their dynamic behavior during step-wise changes in cortical activity. The uncertainty in their output during rapid changes in cortical activity impose limitations in the ability of the controller to compensate for rapid changes in patients' cortical state, and pose additional difficulties in being able to provide mathematically proof for the stability of the overall closed-loop system. Conversely, the NeuroSENSE dynamic behavior can be fully captured by a linear and time invariant transfer function, which makes it better suited for closed-loop applications.

Entities:  

Mesh:

Year:  2010        PMID: 21132561     DOI: 10.1007/s10877-010-9266-9

Source DB:  PubMed          Journal:  J Clin Monit Comput        ISSN: 1387-1307            Impact factor:   2.502


  13 in total

1.  Closed-loop control of anesthesia using Bispectral index: performance assessment in patients undergoing major orthopedic surgery under combined general and regional anesthesia.

Authors:  Anthony R Absalom; Nicholas Sutcliffe; Gavin N Kenny
Journal:  Anesthesiology       Date:  2002-01       Impact factor: 7.892

Review 2.  Closed loops in anaesthesia.

Authors:  Michel M R F Struys; Eric P Mortier; Tom De Smet
Journal:  Best Pract Res Clin Anaesthesiol       Date:  2006-03

3.  Quantifying uncertainty bounds in anesthetic PKPD models.

Authors:  Stéphane Bibian; Guy A Dumont; Mihai Huzmezan; Craig R Ries
Journal:  Conf Proc IEEE Eng Med Biol Soc       Date:  2004

4.  Bi-spectral index guided closed-loop anaesthesia delivery system (CLADS) in pheochromocytoma.

Authors:  Harihar V Hegde; G D Puri; Bhupesh Kumar; Arunanshu Behera
Journal:  J Clin Monit Comput       Date:  2009-06-11       Impact factor: 2.502

5.  The accuracy and clinical feasibility of a new bayesian-based closed-loop control system for propofol administration using the bispectral index as a controlled variable.

Authors:  Tom De Smet; Michel M R F Struys; Martine M Neckebroek; Kristof Van den Hauwe; Sjoert Bonte; Eric P Mortier
Journal:  Anesth Analg       Date:  2008-10       Impact factor: 5.108

6.  Comparison of closed-loop controlled administration of propofol using Bispectral Index as the controlled variable versus "standard practice" controlled administration.

Authors:  M M Struys; T De Smet; L F Versichelen; S Van De Velde; R Van den Broecke; E P Mortier
Journal:  Anesthesiology       Date:  2001-07       Impact factor: 7.892

7.  Quantifying cortical activity during general anesthesia using wavelet analysis.

Authors:  Tatjana Zikov; Stéphane Bibian; Guy A Dumont; Mihai Huzmezan; Craig R Ries
Journal:  IEEE Trans Biomed Eng       Date:  2006-04       Impact factor: 4.538

8.  Closed-loop control of consciousness during lung transplantation: an observational study.

Authors:  Ngai Liu; Thierry Chazot; Bernard Trillat; Mireille Michel-Cherqui; Jean Yves Marandon; Jean-Dominique Law-Koune; Bertrand Rives; Marc Fischler
Journal:  J Cardiothorac Vasc Anesth       Date:  2008-06-25       Impact factor: 2.628

Review 9.  Closed-loop strategies for patient care systems.

Authors:  Ronald Pauldine; George Beck; Jose Salinas; David W Kaczka
Journal:  J Trauma       Date:  2008-04

10.  A new closed-loop control system for isoflurane using bispectral index outperforms manual control.

Authors:  Stephan Locher; Konrad S Stadler; Thomas Boehlen; Thomas Bouillon; Daniel Leibundgut; Peter M Schumacher; Rolf Wymann; Alex M Zbinden
Journal:  Anesthesiology       Date:  2004-09       Impact factor: 7.892

View more
  8 in total

1.  Feasibility of continuous sedation monitoring in critically ill intensive care unit patients using the NeuroSENSE WAVCNS index.

Authors:  Nicholas West; Paul B McBeth; Sonia M Brodie; Klaske van Heusden; Sarah Sunderland; Guy A Dumont; Donald E G Griesdale; J Mark Ansermino; Matthias Görges
Journal:  J Clin Monit Comput       Date:  2018-02-20       Impact factor: 2.502

2.  Predicting postoperative delirium and postoperative cognitive decline with combined intraoperative electroencephalogram monitoring and cerebral near-infrared spectroscopy in patients undergoing cardiac interventions.

Authors:  Mona Momeni; Sabrina Meyer; Marie-Agnès Docquier; Guillaume Lemaire; David Kahn; Céline Khalifa; Maria Rosal Martins; Michel Van Dyck; Luc-Marie Jacquet; André Peeters; Christine Watremez
Journal:  J Clin Monit Comput       Date:  2019-01-11       Impact factor: 2.502

3.  Offline comparison of processed electroencephalogram monitors for anaesthetic-induced electroencephalogram changes in older adults.

Authors:  Sarah L Eagleman; Caitlin M Drover; Xi Li; M Bruce MacIver; David R Drover
Journal:  Br J Anaesth       Date:  2021-02-25       Impact factor: 9.166

4.  A Comparison of Multiscale Permutation Entropy Measures in On-Line Depth of Anesthesia Monitoring.

Authors:  Cui Su; Zhenhu Liang; Xiaoli Li; Duan Li; Yongwang Li; Mauro Ursino
Journal:  PLoS One       Date:  2016-10-10       Impact factor: 3.240

Review 5.  A Review of Modern Control Strategies for Clinical Evaluation of Propofol Anesthesia Administration Employing Hypnosis Level Regulation.

Authors:  Muhammad Ilyas; Muhammad Fasih Uddin Butt; Muhammad Bilal; Khalid Mahmood; Ali Khaqan; Raja Ali Riaz
Journal:  Biomed Res Int       Date:  2017-03-30       Impact factor: 3.411

Review 6.  Multiparametric Monitoring of Hypnosis and Nociception-Antinociception Balance during General Anesthesia-A New Era in Patient Safety Standards and Healthcare Management.

Authors:  Alexandru Florin Rogobete; Ovidiu Horea Bedreag; Marius Papurica; Sonia Elena Popovici; Lavinia Melania Bratu; Andreea Rata; Claudiu Rafael Barsac; Andra Maghiar; Dragos Nicolae Garofil; Mihai Negrea; Laura Bostangiu Petcu; Daiana Toma; Corina Maria Dumbuleu; Samir Rimawi; Dorel Sandesc
Journal:  Medicina (Kaunas)       Date:  2021-02-02       Impact factor: 2.430

7.  Predicting Deep Hypnotic State From Sleep Brain Rhythms Using Deep Learning: A Data-Repurposing Approach.

Authors:  Sunil Belur Nagaraj; Sowmya M Ramaswamy; Maud A S Weerink; Michel M R F Struys
Journal:  Anesth Analg       Date:  2020-05       Impact factor: 5.108

8.  Hypnosis regulation in propofol anaesthesia employing super-twisting sliding mode control to compensate variability dynamics.

Authors:  Muhammad Ilyas; Jamshed Iqbal; Sayyar Ahmad; Ali Arshad Uppal; Waqas Ahmad Imtiaz; Raja Ali Riaz
Journal:  IET Syst Biol       Date:  2020-04       Impact factor: 1.615

  8 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.