Literature DB >> 21129339

Dose accuracy and injection force of disposable pens delivering pramlintide for the treatment of diabetes.

Douglas Merritt1, Barb Schreiner, Scott Harris, Mary Beth DeYoung, Susan Strobel, Joseph Lauinger.   

Abstract

BACKGROUND: The pen injection format, typically used for insulin administration, has been adapted for the injectable, noninsulin diabetes therapy pramlintide. Administered before major meals, pramlintide therapy requires two to four injections/day in addition to the patients' usual insulin injections. The dose accuracy and injection force was determined for the 60 and 120 µg pramlintide pens.
METHODS: Dose accuracy testing was conducted at two sites on multiple 60 µg (15, 30, and 60 µg doses) and 120 µg pens (60 and 120 µg doses) at prespecified temperatures (5-40 °C) and humidities (0-75%) using 29 G half-inch needles. All pens were stabilized under testing conditions for 4 h prior to testing. One site used a compression load cell (Zwick device) to test pens; one site performed tests manually. Injection-force testing was conducted at one site on multiple 60 and 120 µg pens at multiple temperatures (18-28 °C) and humidities (25-75%) using 29 and 31 G half-inch needles and an injection speed of 150 m/min. Injection-force testing was performed using a Zwick device.
RESULTS: Dose accuracy for all pens tested, regardless of location, reproducibly met/exceeded acceptance criteria. Mean percentage of dose accuracy was 96.04 to 102.45% [standard deviations (SDs) 0.3 to 1.4 µg] for the 60 µg pen and 98.16 to 101.83% (SDs 0.4 to 2.5 µg) for the 120 µg pen. The average injection force across both pens did not exceed 7 N regardless of needle size.
CONCLUSIONS: High dose accuracy and low injection force were observed for the 60 and 120 µg pens under a variety of conditions.
© 2010 Diabetes Technology Society.

Entities:  

Mesh:

Substances:

Year:  2010        PMID: 21129339      PMCID: PMC3005054          DOI: 10.1177/193229681000400618

Source DB:  PubMed          Journal:  J Diabetes Sci Technol        ISSN: 1932-2968


  15 in total

1.  [Fear of the injection must not be an argument. Every second type 2 diabetic patient needs insulin].

Authors: 
Journal:  MMW Fortschr Med       Date:  2002-12-05

2.  Dosage accuracy of self-mixed vs premixed insulin.

Authors:  D S Bell; R S Clements; G Perentesis; R Roddam; L Wagenknecht
Journal:  Arch Intern Med       Date:  1991-11

3.  Accuracy of pen injectors versus insulin syringes in children with type 1 diabetes.

Authors:  A N Lteif; W F Schwenk
Journal:  Diabetes Care       Date:  1999-01       Impact factor: 19.112

4.  Inadequate suspension of neutral protamine Hagendorn (NPH) insulin in pens.

Authors:  P M Jehle; C Micheler; D R Jehle; D Breitig; B O Boehm
Journal:  Lancet       Date:  1999-11-06       Impact factor: 79.321

5.  Invasiveness as a barrier to self-monitoring of blood glucose in diabetes.

Authors:  Julie Wagner; Carl Malchoff; Gina Abbott
Journal:  Diabetes Technol Ther       Date:  2005-08       Impact factor: 6.118

6.  Preference for insulin delivery systems among current insulin users and nonusers.

Authors:  Kent H Summers; Sheryl L Szeinbach; Sheila M Lenox
Journal:  Clin Ther       Date:  2004-09       Impact factor: 3.393

Review 7.  Dose accuracy and injection force dynamics of a novel disposable insulin pen.

Authors:  Alastair Clarke; Geralyn Spollett
Journal:  Expert Opin Drug Deliv       Date:  2007-03       Impact factor: 6.648

8.  Dosing accuracy of insulin pens versus conventional syringes and vials.

Authors:  Y M Luijf; J H DeVries
Journal:  Diabetes Technol Ther       Date:  2010-06       Impact factor: 6.118

9.  Accuracy and precision of low-dose insulin administration using syringes, pen injectors, and a pump.

Authors:  Katherine Keith; David Nicholson; Douglas Rogers
Journal:  Clin Pediatr (Phila)       Date:  2004 Jan-Feb       Impact factor: 1.168

10.  Accuracy and reproducibility of low dose insulin administration using pen-injectors and syringes.

Authors:  M G Gnanalingham; P Newland; C P Smith
Journal:  Arch Dis Child       Date:  1998-07       Impact factor: 3.791

View more
  1 in total

1.  FlexTouch: An Insulin Pen-Injector with a Low Activation Force Across Different Insulin Formulations, Needle Technologies, and Temperature Conditions.

Authors:  Niels Gudiksen; Thibaud Hofstätter; Birgitte B Rønn; Thomas Sparre
Journal:  Diabetes Technol Ther       Date:  2017-07-13       Impact factor: 6.118

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.