| Literature DB >> 21121072 |
Hong-ying Liu1, Gang Wang, Kang Wei, Xi-tian Pi, Lan Zhu, Xiao-lin Zheng, Zhi-yu Wen.
Abstract
In clinical practice, examination of the hemorrhagic spot (HS) remains difficult. In this paper, we describe a remote controlled capsule (RCC) micro-system with an automated, color-based sensor to identify and localize the HS of the gastrointestinal (GI) tract. In vitro testing of the detecting sensor demonstrated that it was capable of discriminating mimetic intestinal fluid (MIF) with and without the hemoglobin (Hb) when the concentration of Hb in MIF was above 0.05 g/ml. Therefore, this RCC system is able to detect the relatively accurate location of the HS in the GI tract.Entities:
Mesh:
Year: 2010 PMID: 21121072 PMCID: PMC2997402 DOI: 10.1631/jzus.B1000047
Source DB: PubMed Journal: J Zhejiang Univ Sci B ISSN: 1673-1581 Impact factor: 3.066