| Literature DB >> 21118757 |
David M Pearce1, Daniel P Shenton, Jeffrey Holden, Charlotte A Gaydos.
Abstract
Atlas Genetics has developed a point-of-care device for Chlamydia trachomatis utilizing a novel electrochemical detection principle. The assay has a time-to-result of less than 25 min. An independent preclinical validation study using 306 pretyped clinical samples determined a clinical sensitivity of 98.1% and specificity of 98.0%.Entities:
Mesh:
Year: 2010 PMID: 21118757 PMCID: PMC3187616 DOI: 10.1109/TBME.2010.2095851
Source DB: PubMed Journal: IEEE Trans Biomed Eng ISSN: 0018-9294 Impact factor: 4.538