Literature DB >> 21084910

In vivo assessment of coronary stents with 64-row multidetector computed tomography: analysis of metal artifacts.

Giacomo D E Papini1, Filippo Casolo, Giovanni Di Leo, Silvia Briganti, Cecilia Fantoni, Massimo Medda, Luigi Inglese, Luigi De Ambroggi, Francesco Sardanelli.   

Abstract

OBJECTIVE: To evaluate stent-induced artifacts by 64-row multidetector computed tomography (MDCT).
METHODS: We studied 26 stented patients with MDCT before conventional coronary angiography (CCA). The CT values were measured. Stents were classified as occluded, with significant stenosis, with nonsignificant stenosis, or patent. For the patent stents, mean in-stent and out-stent CT values were compared; stents 3 mm or smaller were compared with stents larger than 3 mm. Multidetector CT was compared with CCA.
RESULTS: We analyzed 42 stents. At CCA, 34 stents were patent, 5 were nonsignificantly stenosed, 1 was significantly stenosed, and 2 were occluded. At MDCT, 33 of 34 patent stents, 2 occluded stents, and 1 stent with significant stenosis were correctly diagnosed; nonsignificant stenoses were undetected, 1 patent stent was misdiagnosed as occluded (κ = 0.727). The out-stent CT value was lower than in-stent CT value both in stents 3 mm or smaller (P = 0.001) and stents larger than 3 mm (P < 0.001). The in-stent CT value of stents 3 mm or smaller was higher (P = 0.011) than that of stents larger than 3 mm.
CONCLUSIONS: Metal artifacts cause overlooking of nonsignificant stenosis.

Entities:  

Mesh:

Substances:

Year:  2010        PMID: 21084910     DOI: 10.1097/RCT.0b013e3181ddbbc0

Source DB:  PubMed          Journal:  J Comput Assist Tomogr        ISSN: 0363-8715            Impact factor:   1.826


  3 in total

1.  CT Imaging of Coronary Stents: Past, Present, and Future.

Authors:  Andreas H Mahnken
Journal:  ISRN Cardiol       Date:  2012-09-11

2.  Comparison of different CT metal artifact reduction strategies for standard titanium and carbon-fiber reinforced polymer implants in sheep cadavers.

Authors:  Florian A Huber; Kai Sprengel; Lydia Müller; Laura C Graf; Georg Osterhoff; Roman Guggenberger
Journal:  BMC Med Imaging       Date:  2021-02-15       Impact factor: 1.930

Review 3.  Health Care Monitoring and Treatment for Coronary Artery Diseases: Challenges and Issues.

Authors:  Mokhalad Alghrairi; Nasri Sulaiman; Saad Mutashar
Journal:  Sensors (Basel)       Date:  2020-08-01       Impact factor: 3.576

  3 in total