Nonlinear three-dimensional noise filter with low-dose CT angiography: effect on the detection of small high-contrast objects in a phantom model.

Zsolt Szucs-Farkas1, Susanne Bensler, Jaled Charimo Torrente, Jennifer L Cullmann, Peter Vock, Sebastian T Schindera.   

Abstract

PURPOSE: To study the effect of a nonlinear noise filter on the detection of simulated endoleaks in a phantom with 80- and 100-kVp multidetector computed tomographic (CT) angiography.
MATERIALS AND METHODS: An aortic aneurysm phantom, including iodinated endoleaks, was constructed. Multidetector CT angiography with use of 80-, 100-, and 120-kVp tube voltages was performed for simulated intermediate-sized and large patients (estimated body weights, 72-85 kg and 118-142 kg, respectively). Images obtained with 80 and 100 kVp were postprocessed by using a nonlinear noise filter. CT images containing 1152 endoleaks and images with no endoleaks were randomized and independently analyzed by three radiologists blinded to the location of the endoleaks. Diagnostic confidence and image quality were rated by using subjective scales. Analysis of variance was used for statistical assessment.
RESULTS: In simulated intermediate-sized patients, energy reduction from 120 to 100 kVp and from100 to 80 kVp did not decrease image quality when images with reduced kilovoltage were filtered (P = .2692 and P > .99, respectively). Readers detected more endoleaks on the filtered 100-kVp images than on the nonfiltered images in simulated large patients (83 vs 75 lesions, P = .041). The number of detected endoleaks and the confidence rate were similar at 100 kVp with a filter and at 120 kVp in simulated large patients (P = .339 and P = .211, respectively).
CONCLUSION: In a phantom, the nonlinear noise filter can prevent decreased image quality with use of 80- and 100-kVp abdominal multidetector CT angiography at a wide range of simulated body weights and may facilitate a better detection rate of endoleaks in heavy patients. © RSNA, 2010

Entities:  

Mesh:

Year:  2010        PMID: 21062921     DOI: 10.1148/radiol.10100760

Source DB:  PubMed          Journal:  Radiology        ISSN: 0033-8419            Impact factor:   11.105


  3 in total

Review 1.  Raise the bar and lower the dose: current and future strategies for radiation dose reduction in head and neck imaging.

Authors:  M Ibrahim; H Parmar; E Christodoulou; S Mukherji
Journal:  AJNR Am J Neuroradiol       Date:  2013-02-28       Impact factor: 3.825

2.  Cerebral CTA with Low Tube Voltage and Low Contrast Material Volume for Detection of Intracranial Aneurysms.

Authors:  Q Q Ni; G Z Chen; U J Schoepf; M A J Klitsie; C N De Cecco; C S Zhou; S Luo; G M Lu; L J Zhang
Journal:  AJNR Am J Neuroradiol       Date:  2016-05-05       Impact factor: 3.825

3.  The Impact of Combining a Low-Tube Voltage Acquisition with Iterative Reconstruction on Total Iodine Dose in Coronary CT Angiography.

Authors:  Toon Van Cauteren; Gert Van Gompel; Kaoru Tanaka; Douwe E Verdries; Dries Belsack; Koenraad H Nieboer; Inneke Willekens; Paul Evans; Sven Macholl; Guy Verfaillie; Steven Droogmans; Johan de Mey; Nico Buls
Journal:  Biomed Res Int       Date:  2017-05-23       Impact factor: 3.411

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.