Electrical activity of the diaphragm during extubation readiness testing in critically ill children.

Gerhard K Wolf1, Brian K Walsh, Michael L Green, John H Arnold.   

Abstract

OBJECTIVES: To investigate the electrical activity of the diaphragm during extubation readiness testing.
DESIGN: Prospective observational trial.
SETTING: A 29-bed medical-surgical pediatric intensive care unit. PATIENTS: Mechanically ventilated children between 1 month and 18 yrs of age.
INTERVENTIONS: Twenty patients underwent a standardized extubation readiness test using a minimal pressure support ventilation strategy. A size-appropriate multiple-array esophageal electrode (electrical diaphragmatic activity catheter), which doubled as a feeding tube, was inserted. The electrical diaphragmatic activity, ventilatory parameters, and spirometry measurements were recorded with the Servo-i ventilator (Maquet, Solna, Sweden). Measurements were obtained before the extubation readiness test and 1 hr into the extubation readiness test.
MEASUREMENTS AND MAIN RESULTS: During extubation readiness testing, the ratio of tidal volume to delta electrical diaphragmatic activity was significantly lower in those patients who passed the extubation readiness test compared to those who failed the extubation readiness test (extubation readiness test, pass: 24.8 ± 20.9 mL/μV vs. extubation readiness test, fail: 67.2 ± 27 mL/μV, respectively; p = .02). Delta electrical diaphragmatic activity correlated significantly with neuromuscular drive assessed by airway opening pressure at 0.1 secs (before extubation readiness test: r = .591, p < .001; during extubation readiness test: r = .682, p < .001). Eight out of 20 patients had ventilator dys-synchrony identified with electrical diaphragmatic activity during extubation readiness testing.
CONCLUSIONS: Patients who generate higher diaphragmatic activity in relation to tidal volume may have better preserved diaphragmatic function and a better chance of passing the extubation readiness test as opposed to patients who generate lower diaphragmatic activity in relation to tidal volume, indicating diaphragmatic weakness. Electrical activity of the diaphragm also may be a useful adjunct to assess neuromuscular drive in ventilated children.

Entities:  

Mesh:

Year:  2011        PMID: 21057368     DOI: 10.1097/PCC.0b013e3181fe28fc

Source DB:  PubMed          Journal:  Pediatr Crit Care Med        ISSN: 1529-7535            Impact factor:   3.624


  16 in total

Review 1.  The role of computer-based clinical decision support systems to deliver protective mechanical ventilation.

Authors:  Robinder G Khemani; Justin C Hotz; Katherine A Sward; Christopher J L Newth
Journal:  Curr Opin Crit Care       Date:  2020-02       Impact factor: 3.687

2.  Evolution of inspiratory diaphragm activity in children over the course of the PICU stay.

Authors:  Guillaume Emeriaud; Alexandrine Larouche; Laurence Ducharme-Crevier; Erika Massicotte; Olivier Fléchelles; Amélie-Ann Pellerin-Leblanc; Sylvain Morneau; Jennifer Beck; Philippe Jouvet
Journal:  Intensive Care Med       Date:  2014-08-15       Impact factor: 17.440

3.  Diaphragm electromyographic activity as a predictor of weaning failure.

Authors:  Martin Dres; Matthieu Schmidt; Alexis Ferre; Julien Mayaux; Thomas Similowski; Alexandre Demoule
Journal:  Intensive Care Med       Date:  2012-09-26       Impact factor: 17.440

4.  Diaphragmatic electromyography during a spontaneous breathing trial to predict extubation failure in preterm infants.

Authors:  Emma E Williams; Fahad M S Arattu Thodika; Imogen Chappelow; Nicole Chapman-Hatchett; Theodore Dassios; Anne Greenough
Journal:  Pediatr Res       Date:  2022-05-06       Impact factor: 3.953

5.  Risk Factors for Pediatric Extubation Failure: The Importance of Respiratory Muscle Strength.

Authors:  Robinder G Khemani; Tro Sekayan; Justin Hotz; Rutger C Flink; Gerrard F Rafferty; Narayan Iyer; Christopher J L Newth
Journal:  Crit Care Med       Date:  2017-08       Impact factor: 7.598

6.  Pediatric extubation readiness tests should not use pressure support.

Authors:  Robinder G Khemani; Justin Hotz; Rica Morzov; Rutger C Flink; Asvari Kamerkar; Marie LaFortune; Gerrard F Rafferty; Patrick A Ross; Christopher J L Newth
Journal:  Intensive Care Med       Date:  2016-06-18       Impact factor: 17.440

7.  Evolution of inspiratory muscle function in children during mechanical ventilation.

Authors:  Benjamin Crulli; Atsushi Kawaguchi; Jean-Paul Praud; Basil J Petrof; Karen Harrington; Guillaume Emeriaud
Journal:  Crit Care       Date:  2021-06-30       Impact factor: 9.097

8.  Interest of monitoring diaphragmatic electrical activity in the pediatric intensive care unit.

Authors:  Laurence Ducharme-Crevier; Geneviève Du Pont-Thibodeau; Guillaume Emeriaud
Journal:  Crit Care Res Pract       Date:  2013-02-21

9.  Real-Time Effort Driven Ventilator Management: A Pilot Study.

Authors:  Justin C Hotz; Dinnel Bornstein; Kristen Kohler; Erin Smith; Anil Suresh; Margaret Klein; Anoopindar Bhalla; Christopher J Newth; Robinder G Khemani
Journal:  Pediatr Crit Care Med       Date:  2020-11       Impact factor: 3.971

10.  High Breath-by-Breath Variability Is Associated With Extubation Failure in Children.

Authors:  Kelby E Knox; Leonardo Nava-Guerra; Justin C Hotz; Christopher J L Newth; Michael C K Khoo; Robinder G Khemani
Journal:  Crit Care Med       Date:  2020-08       Impact factor: 9.296

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.