Mechanical failure of the long gamma nail in two proximal femur fractures.

Soheil Najibi1, Lemos Mark, David Fehnel.   

Abstract

Mechanical failure of the long gamma nail was encountered in two elderly patients with proximal femur fractures over a 6-month period. One of the patients had a known history of lymphoma. The other patient had a history of rheumatoid arthritis but no history of cancer or other metabolic bone disease. Both nails failed at the junction of the compression screw and the nail. The angle of failure of the nail was the same in both cases. The index of suspicion for imminent mechanical failure of the gamma nail should be higher in pathologic fractures and fractures which are malreduced during nailing.

Entities:  

Mesh:

Year:  2010        PMID: 21046000      PMCID: PMC2958299     

Source DB:  PubMed          Journal:  Iowa Orthop J        ISSN: 1541-5457


  6 in total

1.  Mechanical failure of a gamma nail in a patient with an impending pathologic subtrochanteric fracture.

Authors:  J A Randle; B Meisami-Fard; M D McKee
Journal:  Can J Surg       Date:  1999-10       Impact factor: 2.089

2.  Use of the Gamma nail in the treatment of fractures of the proximal femur.

Authors:  J A Valverde; M G Alonso; J G Porro; D Rueda; P M Larrauri; J J Soler
Journal:  Clin Orthop Relat Res       Date:  1998-05       Impact factor: 4.176

3.  Failure of the gamma nail in a highly unstable proximal femur fracture: report of four cases encountered in The Netherlands.

Authors:  W A van den Brink; I M Janssen
Journal:  J Orthop Trauma       Date:  1995-02       Impact factor: 2.512

4.  Fractured Gamma nail.

Authors:  G Zafiropoulos; D J Pratt
Journal:  Injury       Date:  1994-07       Impact factor: 2.586

5.  Multicenter trial of modified Gamma nail in East Asia.

Authors:  K S Leung; C M Chen; W S So; K Sato; C H Lai; B Machaisavariya; S Suntharalingam
Journal:  Clin Orthop Relat Res       Date:  1996-02       Impact factor: 4.176

6.  Treatment of impending and actual pathological femoral fractures with the long Gamma nail in The Netherlands.

Authors:  R van Doorn; J W Stapert
Journal:  Eur J Surg       Date:  2000-03
  6 in total
  3 in total

1.  Breakage of intramedullary femoral nailing or femoral plating: how to prevent implant failure.

Authors:  Henrik C Bäcker; Mark Heyland; Chia H Wu; Carsten Perka; Ulrich Stöckle; Karl F Braun
Journal:  Eur J Med Res       Date:  2022-01-13       Impact factor: 2.175

2.  Breakage of a third generation gamma nail: a case report and review of the literature.

Authors:  Takashi Iwakura; Takahiro Niikura; Sang Yang Lee; Yoshitada Sakai; Kotaro Nishida; Ryosuke Kuroda; Masahiro Kurosaka
Journal:  Case Rep Orthop       Date:  2013-05-20

3.  Strength⁻Durability Correlation of Osteosynthesis Devices Made by 3D Layer Manufacturing.

Authors:  Yoshimitsu Okazaki; Emiko Gotoh; Jun Mori
Journal:  Materials (Basel)       Date:  2019-01-31       Impact factor: 3.623

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.