Biomechanical evaluation of an expandable meshed bag augmented with pedicle or facet screws for percutaneous lumbar interbody fusion.

Xiujun Zheng1, Rahul Chaudhari, Chunhui Wu, Amir A Mehbod, Serkan Erkan, Ensor E Transfeldt.   

Abstract

OBJECTIVE: To evaluate the biomechanics of lumbar motion segments instrumented with stand-alone OptiMesh system augmented with posterior fixation using facet or pedicle screws and the efficacy of discectomy and disc distraction. BACKGROUND CONTEXT: OptiMesh bone graft containment system has been used for vertebral compression fractures and percutaneous lumbar interbody fusion. The filled mesh bag serves as the interbody device providing structural support to the motion segment being fused. No biomechanical data of this new device are available in the literature.
METHODS: Twenty-four fresh human cadaveric lumbar motion segments were divided into two groups. In the control group, multidirectional flexibility testing was conducted after an intact condition and standard transforaminal lumbar interbody fusion (TLIF) procedure. In the OptiMesh group, testing was performed following intact, stand-alone OptiMesh procedure, OptiMesh with facet screws (placed using the transfacet approach), and OptiMesh with pedicle screws and rods. Range of motion (ROM) was calculated for each surgical treatment. The lordosis and disc height change of intact and instrumented specimens were measured in the lateral radiographs to evaluate the disc space distraction. In the OptiMesh group, cyclic loading in flexion extension (FE) was applied to measure cage subsidence or collapse (10,000 cycles at 6 Nm). After biomechanical testing, all the specimens were dissected to inspect the discectomy and end plate preparation. The area of discectomy was measured.
RESULTS: The mean ROM of the intact specimens was 2.7°, 7.4°, and 7.2° in axial torsion (AT), lateral bending (LB), and FE, respectively. There was no difference between the control group and OptiMesh group. The mean ROM of the stand-alone OptiMesh system decreased to 2.4°, 5.1°, and 4.3° in AT, LB, and FE. The ROM decreased to 0.9° in AT, 2.2° in LB, and 0.9° in FE with OptiMesh system and facet screws. On average, OptiMesh system with pedicle screws and rods reduced the ROM to 1.3° in AT, 1.6° in LB, and 1.1° in FE. Compared with the intact condition and stand-alone OptiMesh system, both posterior fixation options had significant statistical difference (p<.001). In AT, ROM of facet screws was lower than that of pedicle screws (p < .05). There was no statistical difference between the facet and pedicle screws in LB and FE (p > .05). The mean volume of bone graft packed into each bag was 8.3 ± 1.5 cc. The average increase of lordosis was 0.6° ± 1.0° after meshed bag was deployed. The average distraction achieved by the OptiMesh system was 1.0 ± 0.6 mm. The average prepared area of discectomy was 42% of the total disc. The disc height change after cyclic loading was 0.2 mm. No subsidence or collapse was noticed.
CONCLUSIONS: The OptiMesh system offers large volume of bone graft in the disc space with small access portals. The OptiMesh system had similar construct stability to that of standard TLIF procedure when posterior fixation was applied. However, the amount of distraction was limited without additional distraction tools. With the anterior support provided by the expandable meshed bag, facet screws had comparable construct stability to that of pedicle screws. Slightly higher stability was observed in facet screws in AT.
Copyright © 2010 Elsevier Inc. All rights reserved.

Entities:  

Mesh:

Year:  2010        PMID: 20970738     DOI: 10.1016/j.spinee.2010.08.016

Source DB:  PubMed          Journal:  Spine J        ISSN: 1529-9430            Impact factor:   4.166


  8 in total

1.  The History of Endoscopic Posterior Lumbar Surgery.

Authors:  Choll W Kim; Frank Phillips
Journal:  Int J Spine Surg       Date:  2021-12

2.  Comparative Effectiveness of Expandable Versus Static Interbody Spacers via MIS LLIF: A 2-Year Radiographic and Clinical Outcomes Study.

Authors:  Yan Michael Li; Richard F Frisch; Zheng Huang; James Towner; Yan Icy Li; Samantha L Greeley; Charles Ledonio
Journal:  Global Spine J       Date:  2019-10-29

3.  Can prevention of a reherniation be investigated? Establishment of a herniation model and experiments with an anular closure device.

Authors:  Hans-Joachim Wilke; Lena Ressel; Frank Heuer; Nicolas Graf; Stefan Rath
Journal:  Spine (Phila Pa 1976)       Date:  2013-05-01       Impact factor: 3.468

4.  The current testing protocols for biomechanical evaluation of lumbar spinal implants in laboratory setting: a review of the literature.

Authors:  Sabrina A Gonzalez-Blohm; James J Doulgeris; William E Lee; Thomas M Shea; Kamran Aghayev; Frank D Vrionis
Journal:  Biomed Res Int       Date:  2015-02-15       Impact factor: 3.411

5.  Minimally invasive treatment of a painful osteolytic lumbar lesion secondary to epithelioid hemangioendothelioma.

Authors:  Arjun S Sebastian; Marcus J Adair; Jonathan M Morris; Mustafa H Khan; Carola A S Arndt; Ahmad Nassr
Journal:  Global Spine J       Date:  2014-08-05

6.  Biomechanical Analysis of Cortical Versus Pedicle Screw Fixation Stability in TLIF, PLIF, and XLIF Applications.

Authors:  Edward K Nomoto; Guy R Fogel; Alexandre Rasouli; Justin V Bundy; Alexander W Turner
Journal:  Global Spine J       Date:  2018-07-31

7.  Minimally Invasive Lumbar Interbody Fusion With an Expandable Meshed Allograft Containment Device: Analysis of Subsidence With 12-Month Minimum Follow-Up.

Authors:  John Paul G Kolcun; George M Ghobrial; Kenneth M Crandall; Ken Hsuan-Kan Chang; Giacomo Pacchiorotti; Michael Y Wang
Journal:  Int J Spine Surg       Date:  2019-08-31

Review 8.  Recent Trends, Technical Concepts and Components of Computer-Assisted Orthopedic Surgery Systems: A Comprehensive Review.

Authors:  Jan Kubicek; Filip Tomanec; Martin Cerny; Dominik Vilimek; Martina Kalova; David Oczka
Journal:  Sensors (Basel)       Date:  2019-11-27       Impact factor: 3.576

  8 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.