Effect of working length measurement by electronic apex locator or radiography on the adequacy of final working length: a randomized clinical trial.

Shohreh Ravanshad1, Alireza Adl, Javad Anvar.   

Abstract

INTRODUCTION: Obtaining a correct working length is critical to the success of endodontic therapy. The aim of this clinical study was to compare the effect of working length determination using electronic apex locator or working length radiograph on the length adequacy of final working length as well as the final obturation.
METHODS: A total of 84 patients with 188 canals were randomized into two groups; in group 1, the working length was determined by working length radiograph, whereas in group 2, it was determined by the Raypex5 electronic apex locator (VDW, Munich, Germany). Length adequacy was assessed in each group for master cone and final obturation radiography and categorized into short, acceptable, and over cases.
RESULTS: There was no statistically significant difference between the rates of acceptable (master cone radiography: group 1 = 82.1% and group 2 = 90.4%; final radiography: group 1 = 85.7% and group 2 = 90.4%) and short cases (master cone radiography: group 1 = 7.1% and group 2 = 8.7%; final radiography: group 1 = 1.2% and group 2 = 1%) between the two groups. Over cases in master cone radiography were significantly more in group 1 (10.7%) than group 2 (1%) (χ², p = 0.00). However, this category did not show a significant difference for final obturation between group 1 (13.1%) and group 2 (8.7%).
CONCLUSION: The results of endodontic treatment using the Raypex5 electronic apex locator are quite comparable, if not superior, to radiographic length measurement regarding the rates of acceptable and short cases. Furthermore, in addition to reducing the radiographic exposure, electronic apex locators are superior in reducing overestimation of the root canal length.
Copyright © 2010 American Association of Endodontists. Published by Elsevier Inc. All rights reserved.

Entities:  

Mesh:

Substances:

Year:  2010        PMID: 20951282     DOI: 10.1016/j.joen.2010.08.017

Source DB:  PubMed          Journal:  J Endod        ISSN: 0099-2399            Impact factor:   4.171


  14 in total

1.  Is endodontic treatment performed under general anaesthesia technically acceptable?

Authors:  Iyad Alsaleh; Pierre-Yves Cousson; Emmanuel Nicolas; Martine Hennequin
Journal:  Clin Oral Investig       Date:  2011-12-24       Impact factor: 3.573

2.  The influence of periapical lesions on the repeatability of two electronic apex locators in vivo.

Authors:  Carlos G Adorno; Sandra M Solaeche; Ileana E Ferreira; Andrea Pedrozo; Patricia M Escobar; Vicente R Fretes
Journal:  Clin Oral Investig       Date:  2021-02-13       Impact factor: 3.573

3.  An ex vivo comparison of working length determination by three electronic root canal length measurement devices integrated into endodontic rotary motors.

Authors:  Manal Maree Ali; Ronald Wigler; Shaul Lin; Arieh Y Kaufman
Journal:  Clin Oral Investig       Date:  2016-07-09       Impact factor: 3.573

4.  Review of High Level Endodontic Research in PubMed Index Journals from Iran.

Authors:  Mohammad Jafar Eghbal; Hassan Torabzadeh
Journal:  Iran Endod J       Date:  2012-08-01

5.  Efficiency of 2 electronic apex locators on working length determination: A clinical study.

Authors:  Sibel Koçak; Mustafa Murat Koçak; Baran Can Sağlam
Journal:  J Conserv Dent       Date:  2013-05

6.  Accuracy of an Electronic Apex Locator for Working Length Determination in Primary Anterior Teeth.

Authors:  Zahra Bahrololoomi; Ali Asghar Soleymani; Jalil Modaresi; Mahsa Imanian; Malihe Lotfian
Journal:  J Dent (Tehran)       Date:  2015-04

7.  Working Length Determination Using Cone-Beam Computed Tomography, Periapical Radiography and Electronic Apex Locator in Teeth with Apical Periodontitis: A Clinical Study.

Authors:  André Luiz Gomide de Morais; Ana Helena Gonçalves de Alencar; Cyntia Rodrigues de Araújo Estrela; Daniel Almeida Decurcio; Carlos Estrela
Journal:  Iran Endod J       Date:  2016-05-01

8.  Root Morphology and Canal Configuration of First and Second Maxillary Molars in a Selected Iranian Population: A Cone-Beam Computed Tomography Evaluation.

Authors:  Abbasali Khademi; Asieh Zamani Naser; Zahra Bahreinian; Mojdeh Mehdizadeh; Mojtaba Najarian; Saber Khazaei
Journal:  Iran Endod J       Date:  2017

9.  An In Vitro Comparison of Propex II Apex Locator to Standard Radiographic Method.

Authors:  Kalyan Vinayak Chakravarthy Pishipati
Journal:  Iran Endod J       Date:  2013-08-01

10.  An in vivo comparison of accuracy of two electronic apex locators in determining working length using stainless steel and nickel titanium files.

Authors:  Sanjeeb Chaudhary; Archana Gharti; Bhawana Adhikari
Journal:  Clin Cosmet Investig Dent       Date:  2018-05-22
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.