Lo J Bour1, M Fiorella Contarino, Elisabeth M J Foncke, Rob M A de Bie, Pepijn van den Munckhof, Johannes D Speelman, P Richard Schuurman.
Abstract
BACKGROUND: Intraoperative microelectrode recording (MER) for targeting during deep brain stimulation (DBS) procedures has been evaluated over a period of 4 years, in 57 consecutive patients with Parkinson's disease, who received DBS in the subthalamic nucleus (STN-DBS), and 28 consecutive patients with either dystonia (23) or Parkinson's disease (five), in whom the internal segment of the globus pallidus (GPi-DBS) was targeted.Entities:
Mesh:
Year: 2010 PMID: 20949292 PMCID: PMC2991152 DOI: 10.1007/s00701-010-0835-y
Source DB: PubMed Journal: Acta Neurochir (Wien) ISSN: 0001-6268 Impact factor: 2.216
Fig. 1Single/multi-unit activity of STN for various depths (−2 to +5 mm). Note that the negative values correspond with positions above the target. The superficial part of the STN (−1 to 0 mm) is clearly recognized by an increase in background noise and a sudden increase in discharge rate characterized by rhythmic bursts of activity in this case from 15 to 25 Hz. Deeper layers of the STN (+1.5 to 3.5 mm) show a more irregular high-frequency discharge pattern. SN activity (+5 mm) consists of a low-frequency tonic discharge
Fig. 2Single/multi-unit activity of GPi for various depths (−7.5 to 0 mm). Note that the negative values correspond with positions above the MRI-based target. At −4.5 and −4 mm clearly, the lamina between the external and internal segment of the GP can be observed. The bottom of the GPi is reached at 0 mm
Comparison between the channel with the best MER activity and the channel finally chosen for permanent stimulation in the STN
| Results STN | Right ( | Left ( | Total ( |
|---|---|---|---|
| Final electrode | |||
| Central | 47% (22) | 55% (28) | 50% |
| Medial | 8.5% (4) | 4% (2) | 6% |
| Lateral | 15% (7) | 6% (3) | 10% |
| Anterior | 25.5% (12) | 20% (10) | 24% |
| Posterior | 4% (2) | 15% (8) | 10% |
| Best electrode | |||
| Central | 47% (22) | 47% (24) | 47% |
| Medial | 13% (6) | 4% (2) | 8% |
| Lateral | 23% (11) | 16% (8) | 20% |
| Anterior | 13% (6) | 20% (10) | 16% |
| Posterior | 4% (2) | 13% (7) | 9% |
| Final = best | |||
| Central | 77% (17/22) | 79% (19/24) | 78% (36/46) |
| Medial | 67% (4/6) | 0% (0/2) | 50% (4/8) |
| Lateral | 36% (4/11) | 25% (2/8) | 32% (6/19) |
| Anterior | 100% (6/6) | 70% (7/10) | 81% (13/16) |
| Posterior | 50% (1/2) | 57% (4/7) | 55% (5/9) |
| Total | 66% (31/47) | 63% (32/51) | 64% (63/98) |
Comparison between the channel with the best MER activity and the channel finally chosen for permanent stimulation in the GPi
| Results GPi | Right ( | Left ( | Total ( |
|---|---|---|---|
| Final electrode | |||
| Central | 50% (12) | 64% (14) | 57% |
| Medial | 8% (2) | 18% (4) | 13% |
| Lateral | 34% (8) | 9% (2) | 22% |
| Anterior | 4% (1) | 0% (0) | 2% |
| Posterior | 4% (1) | 9% (2) | 6% |
| Best electrode | |||
| Central | 54% (13) | 54% (12) | 54% |
| Medial | 12.5% (3) | 18% (4) | 15% |
| Lateral | 12.5% (3) | 9% (2) | 11% |
| Anterior | 0% (0) | 0% (0) | 0% |
| Posterior | 21% (5) | 18% (4) | 20% |
| Final/best | |||
| Central | 85% (11/13) | 67% (8/12) | 76% (19/25) |
| Medial | 33% (1/3) | 50% (2/4) | 43% (3/7) |
| Lateral | 100% (3/3) | 50% (1/2) | 80% (4/5) |
| Anterior | 0% (0/0) | 0% (0/0) | 0% (0/0) |
| Posterior | 20% (1/5) | 25% (1/4) | 22% (2/9) |
| Total | 67% (16/24) | 54% (12/22) | 61% (28/46) |
Fig. 3a STN frequency distribution of final depth of the deepest contact point (mean = 2.1 mm; SD = 1.5 mm) with respect to the MRI-based target. Intended coordinates: 12 mm lateral to midplane; 2 mm posterior to MCP; 4 mm inferior ICL. b GPi frequency distribution of final depth of the deepest contact point (mean = −0.5 mm; SD = 1.2 mm) with respect to the MRI-based target. Intended coordinates: 21 mm lateral to midplane; 2 mm anterior to MCP; 5 mm inferior to ICL
Fig. 4a Frequency distribution of the STN-middle of the MER activity with respect to the MRI-based STN target. Mean = 0.0 mm; SD = 1.5 mm. b Scatter plot of the relationship between the STN-middle of the MER activity (horizontal) and the final depth of the deepest contact point (vertical) with respect to the MRI-based STN target. Data of right and left STN are pooled