Comparison of image quality and lens dose in helical and sequentially acquired head CT.

N Abdeen1, S Chakraborty, T Nguyen, M P dos Santos, M Donaldson, G Heddon, B A Schwarz.   

Abstract

AIM: To prospectively compare helical and traditional sequential acquisition of computed tomography (CT) images of the head with respect to lens dose, thyroid dose, and image quality.
MATERIALS AND METHODS: Patients undergoing non-emergency unenhanced CT examinations of the head were prospectively randomized to sequential or helical multidetector CT (MDCT) acquisition. Lens dose and thyroid dose were measured directly using a MOSFET (metal oxide field effect transistor) system. The quality of images at the level of the basal ganglia and the posterior fossa were assessed by two neuroradiologists on a scale of 1 (poor) to 5 (very good), with respect to grey-white matter conspicuity and posterior fossa artefact respectively.
RESULTS: The lens dose was significantly higher for the sequential technique than the helical technique (61.5±12.9 mGy versus 43.5±4.3 mGy; p<0.0001). The thyroid dose was not significantly different (1.6±0. 6 mGy versus 2.1±0.6 mGy; p=0.8). The sequential technique was ranked higher than the helical technique with respect to the basal ganglia image for overall quality and grey-white matter conspicuity, but the majority of images in both groups were rated good or excellent. Posterior fossa images showed more artefacts on the sequential images than the helical technique images. Signal-to-noise ratios for grey and white matter were better using the sequential technique.
CONCLUSIONS: Helically acquired head CT images using 16 and 64 section MDCT are of comparable quality to sequential acquisition, have reduced posterior fossa artefacts and approximately two-thirds of the lens dose.
Copyright © 2010 The Royal College of Radiologists. Published by Elsevier Ltd. All rights reserved.

Entities:  

Mesh:

Year:  2010        PMID: 20933640     DOI: 10.1016/j.crad.2010.02.016

Source DB:  PubMed          Journal:  Clin Radiol        ISSN: 0009-9260            Impact factor:   2.350


  5 in total

1.  Clinical evaluation of a dose monitoring software tool based on Monte Carlo Simulation in assessment of eye lens doses for cranial CT scans.

Authors:  Nika Guberina; Saravanabavaan Suntharalingam; Kai Naßenstein; Michael Forsting; Jens Theysohn; Axel Wetter; Adrian Ringelstein
Journal:  Neuroradiology       Date:  2016-07-20       Impact factor: 2.804

2.  Evaluation of patient dose and operator dose in swallowing CT studies performed with a 320-detector-row multislice CT scanner.

Authors:  Masanao Kobayashi; Kichiro Koshida; Shoichi Suzuki; Kazuhiro Katada
Journal:  Radiol Phys Technol       Date:  2012-03-04

3.  Three-dimensional skull models as a problem-solving tool in suspected child abuse.

Authors:  Sanjay P Prabhu; Alice W Newton; Jeannette M Perez-Rossello; Paul K Kleinman
Journal:  Pediatr Radiol       Date:  2012-11-25

4.  Can leaded glasses protect the eye lens in patients undergoing neck computed tomography?

Authors:  Reza Abedi; Naser Ghaemian; Ali Shabestani Monfared; Mohammad Kiapour; Razzagh Abedi-Firouzjah; Fatemeh Niksirat; Alaba Tolulope Agbele; Kourosh Ebrahimnejad Gorji
Journal:  J Clin Transl Res       Date:  2021-07-16

5.  Deep learning-based reconstruction may improve non-contrast cerebral CT imaging compared to other current reconstruction algorithms.

Authors:  Luuk J Oostveen; Frederick J A Meijer; Frank de Lange; Ewoud J Smit; Sjoert A Pegge; Stefan C A Steens; Martin J van Amerongen; Mathias Prokop; Ioannis Sechopoulos
Journal:  Eur Radiol       Date:  2021-03-10       Impact factor: 5.315

  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.