Literature DB >> 20924806

Comparison of MALDI TOF with conventional identification of clinically relevant bacteria.

Martin Risch1, Darko Radjenovic, Jong Nam Han, Monica Wydler, Urs Nydegger, Lorenz Risch.   

Abstract

BACKGROUND: A completely new approach to diagnose microbial agents at least one day earlier based on mass spectrometric analysis becomes possible in the microbiology laboratory: MALDI TOF: matrix-assisted laser desorption/ionisation time-of-flight. Comparison between results of the new procedure with those obtained by conventional testing is mandatory.
METHODS: 204 clinical isolates grown on agar plates were analysed both, by the MALDI TOF Bruker microflex apparatus and by conventional identification using the VITEK II and API systems, both from bioMérieux.
RESULTS: Of the identified isolates, 72 were gram-positive and 130 gram-negative; 2 were yeasts (candida). Concordance was seen with 61/72 (85%) of the Gram-positive bacteria and with 115/130 (88%) of the Gram-negative bacteria. In 27 samples (13.2%), a discrepancy of the species and/or genus was obvious. The discrepancy appeared with 16 gram-negative (12.2%) and with 11 gram-positive germs (15.3%, n.s.). In the latter group, 6 samples showed discordance with Streptococcus pneumoniae (MALDI) and Streptococcus mitis/oralis (conventional identification) constellation. Among gram-negative samples, most differences occurred on the species level only, e.g. Enterobacter cloacae versus Enterobacter kobei. In 5 cases, discordance was major and appeared on the genus level: Enterobacter/Raoultella, Streptococcus/Gemella, Pseumdomonas/Burkholderia, Microbacter/Sphingomonas and Candida/Cryptococcus. The most outstanding difference was Microbacterium arborescens (MALDI TOF) and Sphingomonas paucimobilis (conventional). Molecular biological identification of two Streptococcus mitis group bacteria confirmed the erroneous diagnosis by MALDI TOF of Streptococcus pneumoniae.
CONCLUSION: Good comparability between MALDI TOF analysis and conventional identification procedures (86.8%) but special caution is needed when identifying streptococcal species.

Entities:  

Mesh:

Substances:

Year:  2010        PMID: 20924806     DOI: 10.4414/smw.2010.13095

Source DB:  PubMed          Journal:  Swiss Med Wkly        ISSN: 0036-7672            Impact factor:   2.193


  24 in total

1.  Clinical characteristics of Raoultella ornithinolytica bacteremia.

Authors:  S Chun; J W Yun; H J Huh; N Y Lee
Journal:  Infection       Date:  2014-11-01       Impact factor: 3.553

2.  Evaluation of species-specific score cutoff values of routinely isolated clinically relevant bacteria using a direct smear preparation for matrix-assisted laser desorption/ionization time-of-flight mass spectrometry-based bacterial identification.

Authors:  F Szabados; H Tix; A Anders; M Kaase; S G Gatermann; G Geis
Journal:  Eur J Clin Microbiol Infect Dis       Date:  2011-09-27       Impact factor: 3.267

3.  Direct Identification of Aerobic Bacteria by Matrix-Assisted Laser Desorption Ionization Time-of-Flight Mass Spectrometry Is Accurate and Robust.

Authors:  Javier Mestas; Teephany Quias; Jennifer Dien Bard
Journal:  J Clin Lab Anal       Date:  2015-12-14       Impact factor: 2.352

4.  Performances of the Vitek MS matrix-assisted laser desorption ionization-time of flight mass spectrometry system for rapid identification of bacteria in routine clinical microbiology.

Authors:  Damien Dubois; Marion Grare; Marie-Françoise Prere; Christine Segonds; Nicole Marty; Eric Oswald
Journal:  J Clin Microbiol       Date:  2012-05-16       Impact factor: 5.948

5.  Multi-center evaluation of the VITEK® MS system for mass spectrometric identification of non-Enterobacteriaceae Gram-negative bacilli.

Authors:  R Manji; M Bythrow; J A Branda; C-A D Burnham; M J Ferraro; O B Garner; R Jennemann; M A Lewinski; A B Mochon; G W Procop; S S Richter; J A Rychert; L Sercia; L F Westblade; C C Ginocchio
Journal:  Eur J Clin Microbiol Infect Dis       Date:  2013-09-10       Impact factor: 3.267

6.  Comparison of direct colony method versus extraction method for identification of gram-positive cocci by use of Bruker Biotyper matrix-assisted laser desorption ionization-time of flight mass spectrometry.

Authors:  Adnan A Alatoom; Scott A Cunningham; Sherry M Ihde; Jayawant Mandrekar; Robin Patel
Journal:  J Clin Microbiol       Date:  2011-05-25       Impact factor: 5.948

7.  Identification of blood culture isolates directly from positive blood cultures by use of matrix-assisted laser desorption ionization-time of flight mass spectrometry and a commercial extraction system: analysis of performance, cost, and turnaround time.

Authors:  Philippe R S Lagacé-Wiens; Heather J Adam; James A Karlowsky; Kimberly A Nichol; Paulette F Pang; Jodi Guenther; Amanda A Webb; Crystal Miller; Michelle J Alfa
Journal:  J Clin Microbiol       Date:  2012-08-08       Impact factor: 5.948

8.  Optimizing identification of clinically relevant Gram-positive organisms by use of the Bruker Biotyper matrix-assisted laser desorption ionization-time of flight mass spectrometry system.

Authors:  Erin McElvania Tekippe; Sunni Shuey; David W Winkler; Meghan A Butler; Carey-Ann D Burnham
Journal:  J Clin Microbiol       Date:  2013-02-20       Impact factor: 5.948

9.  Evaluation of the Bruker Biotyper and VITEK MS MALDI-TOF MS systems for the identification of unusual and/or difficult-to-identify microorganisms isolated from clinical specimens.

Authors:  E McElvania TeKippe; C-A D Burnham
Journal:  Eur J Clin Microbiol Infect Dis       Date:  2014-06-26       Impact factor: 3.267

10.  Identification of Rare Bacterial Pathogens by 16S rRNA Gene Sequencing and MALDI-TOF MS.

Authors:  Percy Schröttner; Florian Gunzer; Jana Schüppel; Wolfram W Rudolph
Journal:  J Vis Exp       Date:  2016-07-11       Impact factor: 1.355