AIM: Measurement of interface pressure between the skin and a compression device has gained practical importance not only for characterizing the efficacy of different compression products in physiological and clinical studies but also for the training of medical staff.
METHODS: A newly developed portable pneumatic pressure transducer (Picopress®) was compared with two established systems (Kikuhime® and SIGaT tester®) measuring linearity, variability and accuracy on a cylindrical model using a stepwise inflated sphygmomanometer as the reference. In addition the variation coefficients were measured by applying the transducers repeatedly under a blood pressure cuff on the distal lower leg of a healthy human subject with stepwise inflation.
RESULTS: In the pressure range between 10 and 80 mmHg all three devices showed a linear association compared with the sphygmomanometer values (Pearson r>0.99). The best reproducibility (variation coefficients between 1.05-7.4%) and the highest degree of accuracy demonstrated by Bland-Altman plots was achieved with the Picopress® transducer. Repeated measurements of pressure in a human leg revealed average variation coefficients for the three devices of 4.17% (Kikuhime®), 8.52% (SIGaT®) and 2.79% (Picopress®).
CONCLUSION: The results suggest that the Picopress® transducer, which also allows dynamic pressure tracing in connection with a software program and which may be left under a bandage for several days, is a reliable instrument for measuring the pressure under a compression device.

Entities:  

Mesh:

Year:  2010        PMID: 20924346

Source DB:  PubMed          Journal:  Int Angiol        ISSN: 0392-9590            Impact factor:   2.789


  11 in total

1.  Compression therapy - cross-sectional observational survey about knowledge and practical treatment of specialised and non-specialised nurses and therapists.

Authors:  Kristina Heyer; Kerstin Protz; Matthias Augustin
Journal:  Int Wound J       Date:  2017-07-25       Impact factor: 3.315

2.  An investigation of the ability to produce a defined 'target pressure' using the PressCise compression bandage.

Authors:  Kerstin Wiklander; Annette Erichsen Andersson; Ulrika Källman
Journal:  Int Wound J       Date:  2015-10-28       Impact factor: 3.315

Review 3.  Muscle Oximetry in Sports Science: A Systematic Review.

Authors:  Stephane Perrey; Marco Ferrari
Journal:  Sports Med       Date:  2018-03       Impact factor: 11.136

4.  Use of customised pressure-guided elastic bandages to improve efficacy of compression bandaging for venous ulcers.

Authors:  Nuttawut Sermsathanasawadi; Choedpong Chatjaturapat; Rattana Pianchareonsin; Nattawut Puangpunngam; Chumpol Wongwanit; Khamin Chinsakchai; Chanean Ruangsetakit; Pramook Mutirangura
Journal:  Int Wound J       Date:  2016-08-09       Impact factor: 3.315

5.  Effects of Compression Stockings on Elevation of Leg Lymph Pumping Pressure and Improvement of Quality of Life in Healthy Female Volunteers: A Randomized Controlled Trial.

Authors:  Ryota Sugisawa; Naoki Unno; Takaaki Saito; Naoto Yamamoto; Kazunori Inuzuka; Hiroki Tanaka; Masaki Sano; Kazuto Katahashi; Hironori Uranaka; Tomohiko Marumo; Hiroyuki Konno
Journal:  Lymphat Res Biol       Date:  2016-01-29       Impact factor: 2.589

6.  Kikuhime Device in the Management of Venous Leg Ulcers.

Authors:  Marian Karafa; Anna Karafova
Journal:  Clin Interv Aging       Date:  2020-09-03       Impact factor: 4.458

7.  Exploring textile-based electrode materials for electromyography smart garments.

Authors:  Emily Lam; Milad Alizadeh-Meghrazi; Alessandra Schlums; Ladan Eskandarian; Amin Mahnam; Bastien Moineau; Milos R Popovic
Journal:  J Rehabil Assist Technol Eng       Date:  2022-02-01

8.  Interim pressure garment therapy (4-6 mmHg) and its effect on donor site healing in burn patients: study protocol for a randomised controlled trial.

Authors:  Michelle L Donovan; Michael J Muller; Claire Simpson; Michael Rudd; Jennifer Paratz
Journal:  Trials       Date:  2016-04-26       Impact factor: 2.279

9.  Study of Three Interface Pressure Measurement Systems Used in the Treatment of Venous Disease.

Authors:  Gayani K Nandasiri; Arash M Shahidi; Tilak Dias
Journal:  Sensors (Basel)       Date:  2020-10-12       Impact factor: 3.576

10.  Fibre Bragg Grating Based Interface Pressure Sensor for Compression Therapy.

Authors:  James A Bradbury; Qimei Zhang; Francisco U Hernandez Ledezma; Ricardo Correia; Serhiy Korposh; Barrie R Hayes-Gill; Ferdinand Tamoué; Alison Parnham; Simon A McMaster; Stephen P Morgan
Journal:  Sensors (Basel)       Date:  2022-02-24       Impact factor: 3.576

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.