Literature DB >> 20924265

A new dual port cutter system for vitrectomy surgery.

Luiz H Lima1, Charles Deboer, Matthew McCormick, Ralph Kerns, Prashant Bhadri, Mark S Humayun.   

Abstract

PURPOSE: To evaluate and compare the performance of several designs of 20-gauge dual port (DP) vitreous cutter tips with a standard 20-gauge single port tip.
METHODS: Custom 20-gauge pneumatic vitreous DP cutter tips with different sizes and port positions were evaluated through porcine vitreous and water flow rates. Five designed and fabricated DP cutter tips were compared with a normal single port control tip and evaluated by the measurement of water and porcine vitreous flow rates, and surgical examination in enucleated porcine eyes.
RESULTS: Some DP tips approached a maximum vitreous and water flow rates, removing water and vitreous faster than the normal control tip. With reference to surgical evaluation, some DP tips performed better than the single port tip for bulk vitrectomy, but none shaved the vitreous base more effectively.
CONCLUSION: The DP cutter system has the potential to increase the flow rates depending on the size and position of the extra port. In the future, the DP cutter may allow the surgeon to perform bulk vitrectomy more efficiently.

Entities:  

Mesh:

Substances:

Year:  2010        PMID: 20924265     DOI: 10.1097/IAE.0b013e3181ea48f9

Source DB:  PubMed          Journal:  Retina        ISSN: 0275-004X            Impact factor:   4.256


  4 in total

1.  Performance analysis of a new hypersonic vitrector system.

Authors:  Paulo Eduardo Stanga; Salvador Pastor-Idoate; Isaac Zambrano; Paul Carlin; David McLeod
Journal:  PLoS One       Date:  2017-06-06       Impact factor: 3.240

2.  Ultrastructural and histopathologic findings after pars plana vitrectomy with a new hypersonic vitrector system. Qualitative preliminary assessment.

Authors:  Salvador Pastor-Idoate; Richard Bonshek; Luciane Irion; Isaac Zambrano; Paul Carlin; Aleksandr Mironov; Paul Bishop; David McLeod; Paulo Eduardo Stanga
Journal:  PLoS One       Date:  2017-04-11       Impact factor: 3.240

3.  Two-Dimensional Cutting (TDC) Vitrectome: In Vitro Flow Assessment and Prospective Clinical Study Evaluating Core Vitrectomy Efficiency versus Standard Vitrectome.

Authors:  Mitrofanis Pavlidis
Journal:  J Ophthalmol       Date:  2016-04-12       Impact factor: 1.909

4.  Treatment of Dropped Nucleus with a 27-Gauge Twin Duty Cycle Vitreous Cutter.

Authors:  Akira Watanabe; Akane Tsuzuki; Kota Arai; Tamaki Gekka; Hiroshi Tsuneoka
Journal:  Case Rep Ophthalmol       Date:  2016-01-14
  4 in total