Literature DB >> 20887149

Accuracy of percutaneous lumbar pedicle screw placement using the oblique or "owl's-eye" view and novel guidance technology.

Cary Idler1, Kevin W Rolfe, Josef E Gorek.   

Abstract

OBJECT: This study was conducted to assess the in vivo safety and accuracy of percutaneous lumbar pedicle screw placement using the owl's-eye view of the pedicle axis and a new guidance technology system that facilitates orientation of the C-arm into the appropriate fluoroscopic view and the pedicle cannulation tool in the corresponding trajectory.
METHODS: A total of 326 percutaneous pedicle screws were placed from L-3 to S-1 in 85 consecutive adult patients. Placement was performed using simple coaxial imaging of the pedicle with the owl's-eye fluoroscopic view. NeuroVision, a new guidance system using accelerometer technology, helped align the C-arm trajectory into the owl's-eye view and the cannulation tool in the same trajectory. Postoperative fine-cut CT scans were acquired to assess screw position. Medical records were reviewed for complications.
RESULTS: Five of 326 screws breached a pedicle cortex—all breaches were less than 2 mm—for an accuracy rate of 98.47%. Five screws violated an adjacent facet joint. All were at the S-1 superior facet and included in a fusion. No screw violated an adjacent mobile facet or disc space. There were no cases of new or worsening neurological symptoms or deficits for an overall clinical accuracy of 100%.
CONCLUSIONS: The owl's-eye technique of coaxial pedicle imaging with the C-arm fluoroscopy, facilitated by NeuroVision, is a safe and accurate means by which to place percutaneous pedicle screws for degenerative conditions of the lumbar spine. This is the largest series reported to use the oblique or owl's-eye projection for percutaneous pedicle screw insertion. The accuracy of percutaneous screw insertion with this technique meets or exceeds that of other reported clinical series or techniques.

Entities:  

Mesh:

Year:  2010        PMID: 20887149     DOI: 10.3171/2010.4.SPINE09580

Source DB:  PubMed          Journal:  J Neurosurg Spine        ISSN: 1547-5646


  14 in total

1.  Fluoroscopy-guided pedicle screw accuracy with a mini-open approach: a tomographic evaluation of 470 screws in 125 patients.

Authors:  José Antonio Soriano-Sánchez; Luis Alberto Ortega-Porcayo; Carlos Francisco Gutiérrez-Partida; Luis Rodolfo Ramírez-Barrios; Ramses Uriel Ortíz-Leyva; Manuel Rodríguez-García; Oscar Sánchez-Escandón
Journal:  Int J Spine Surg       Date:  2015-10-23

2.  A low-cost tracked C-arm (TC-arm) upgrade system for versatile quantitative intraoperative imaging.

Authors:  Shahram Amiri; David R Wilson; Bassam A Masri; Carolyn Anglin
Journal:  Int J Comput Assist Radiol Surg       Date:  2013-12-10       Impact factor: 2.924

3.  CT-navigation versus fluoroscopy-guided placement of pedicle screws at the thoracolumbar spine: single center experience of 4,500 screws.

Authors:  Albrecht Waschke; Jan Walter; Pedro Duenisch; Rupert Reichart; Rolf Kalff; Christian Ewald
Journal:  Eur Spine J       Date:  2012-09-23       Impact factor: 3.134

4.  Biomechanical evaluation of the Facet Wedge: a refined technique for facet fixation.

Authors:  René Hartensuer; Oliver Riesenbeck; Martin Schulze; Dominic Gehweiler; Michael J Raschke; Paul W Pavlov; Thomas Vordemvenne
Journal:  Eur Spine J       Date:  2014-08-26       Impact factor: 3.134

Review 5.  Techniques and accuracy of thoracolumbar pedicle screw placement.

Authors:  Varun Puvanesarajah; Jason A Liauw; Sheng-Fu Lo; Ioan A Lina; Timothy F Witham
Journal:  World J Orthop       Date:  2014-04-18

Review 6.  Methods to determine pedicle screw placement accuracy in spine surgery: a systematic review.

Authors:  Ahmed A Aoude; Maryse Fortin; Rainer Figueiredo; Peter Jarzem; Jean Ouellet; Michael H Weber
Journal:  Eur Spine J       Date:  2015-03-07       Impact factor: 3.134

7.  Lumbosacral pedicle screw placement using a fluoroscopic pedicle axis view and a cannulated tapping device.

Authors:  Toshitaka Yoshii; Takashi Hirai; Tsuyoshi Yamada; Satoshi Sumiya; Renpei Mastumoto; Tsuyoshi Kato; Mitsuhiro Enomoto; Hiroyuki Inose; Shigenori Kawabata; Kenichi Shinomiya; Atsushi Okawa
Journal:  J Orthop Surg Res       Date:  2015-05-28       Impact factor: 2.359

8.  Use of a quantitative pedicle screw accuracy system to assess new technology: Initial studies on O-arm navigation and its effect on the learning curve of percutaneous pedicle screw insertion.

Authors:  Joseph A Sclafani; Gilad J Regev; Jonathan Webb; Steven R Garfin; Choll W Kim
Journal:  SAS J       Date:  2011-09-01

9.  Accuracy of Percutaneous Lumbosacral Pedicle Screw Placement Using the Oblique Fluoroscopic View Based on Computed Tomography Evaluations.

Authors:  Go Yoshida; Koji Sato; Tokumi Kanemura; Toshiki Iwase; Daisuke Togawa; Yukihiro Matsuyama
Journal:  Asian Spine J       Date:  2016-08-16

10.  A new free-hand pedicle screw placement technique with reference to the supraspinal ligament.

Authors:  Juming Li; Hong Zhao; Hao Xie; Lipeng Yu; Jifu Wei; Min Zong; Feng Chen; Ziqiang Zhu; Ning Zhang; Xiaojian Cao
Journal:  J Biomed Res       Date:  2013-10-25
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.