Fast, low-dose patient localization on TomoTherapy via topogram registration.

Kevin L Moore1, Geethpriya Palaniswaamy, Benjamin White, S Murty Goddu, Daniel A Low.   

Abstract

PURPOSE: To investigate a protocol which efficiently localizes TomoTherapy patients with a scout imaging (topogram) mode that can be used with or instead of 3D megavoltage computed tomography (MVCT) imaging.
METHODS: The process presented here is twofold: (a) The acquisition of the topogram using the TomoTherapy MV imaging system and (b) the generation of a digitally reconstructed topogram (DRT) derived from a standard kV CT simulation data set. The unique geometric characteristics of the current TomoTherapy imaging system were explored both theoretically and by acquiring topograms of anthropomorphic phantoms and comparing these images to DRT images. The performance of the MV topogram imaging system in terms of image quality, dose incurred to the patient, and acquisition time was investigated using ionization chamber and radiographic film measurements.
RESULTS: The time required to acquire a clinically usable topogram, limited by the maximum couch speed of 4.0 cm s(-1), was 12.5 s for a 50 cm long field. The patient dose was less than 1% of that delivered by a helical MVCT scan. Further refinements within the current TomoTherapy system, most notably decreasing the imaging beam repetition rate during MV topogram acquisition, would further reduce the topogram dose to less than 25 microGy per scan without compromising image quality.
CONCLUSIONS: Topogram localization on TomoTherapy is a fast and low-dose alternative to 3D MVCT localization. A protocol designed that exclusively utilized MV topograms would result in a 30-fold reduction in imaging time and a 100-fold reduction in dose from localization scans using the current TomoTherapy workflow.

Entities:  

Mesh:

Year:  2010        PMID: 20879568     DOI: 10.1118/1.3453577

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  5 in total

1.  Megavoltage CT imaging quality improvement on TomoTherapy via tensor framelet.

Authors:  Hao Gao; X Sharon Qi; Yu Gao; Daniel A Low
Journal:  Med Phys       Date:  2013-08       Impact factor: 4.071

2.  Technical Note: Iterative megavoltage CT (MVCT) reconstruction using block-matching 3D-transform (BM3D) regularization.

Authors:  Qihui Lyu; Chunlin Yang; Hao Gao; Yi Xue; Daniel O'Connor; Tianye Niu; Ke Sheng
Journal:  Med Phys       Date:  2018-04-30       Impact factor: 4.071

3.  Multi-institutional feasibility study of a fast patient localization method in total marrow irradiation with helical tomotherapy: a global health initiative by the international consortium of total marrow irradiation.

Authors:  Yutaka Takahashi; Stefano Vagge; Stefano Agostinelli; Eunyoung Han; Lukasz Matulewicz; Kai Schubert; Ravishankar Chityala; Vaneerat Ratanatharathorn; Koen Tournel; Jose A Penagaricano; Sterzing Florian; Marc-Andre Mahe; Michael R Verneris; Daniel J Weisdorf; Renzo Corvo; Kathryn E Dusenbery; Guy Storme; Susanta K Hui
Journal:  Int J Radiat Oncol Biol Phys       Date:  2014-10-30       Impact factor: 7.038

4.  Clinical Development and Evaluation of Megavoltage Topogram for Fast Patient Alignment on Helical Tomotherapy.

Authors:  X Sharon Qi; Fang-I Chu; Zhe Zhang; Robert K Chin; Ann Raldow; Amar U Kishan; Percy Lee; Albert Chang; Anusha Kalbasi; Mitchell Kamrava; Michael L Steinberg; Daniel A Low
Journal:  Adv Radiat Oncol       Date:  2020-06-17

5.  Fast Megavoltage Computed Tomography: A Rapid Imaging Method for Total Body or Marrow Irradiation in Helical Tomotherapy.

Authors:  Taiki Magome; Akihiro Haga; Yutaka Takahashi; Keiichi Nakagawa; Kathryn E Dusenbery; Susanta K Hui
Journal:  Int J Radiat Oncol Biol Phys       Date:  2016-07-06       Impact factor: 7.038

  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.