Literature DB >> 20878382

[Automatic registration of patients with A-mode ultrasound for computer-assisted surgery. Laboratory proof of concept].

G Diakov1, F Kral, O Güler, W Freysinger.   

Abstract

BACKGROUND: The main source of error in 3D navigation is the patient-to-image registration process. Anatomical landmarks or adhesive markers perform sub-optimally. Bone-anchored invasive markers significantly change the clinical workflow of navigated ENT surgery, are invasive and cause patient discomfort. In order to minimize registration errors and to further streamline the clinical use of intraoperative 3D navigation we demonstrate that A-mode ultrasound allows an accurate 3D surface profile of the os occipitale to be created which can be reliably registered on preoperative patient CT data.
METHODS: The transducer is mechanically positioned with sub-millimeter accuracy on the patient's occiput. From the sound echos a 3D surface is generated and registered to the preoperative CT images with the iterative closest point (ICP) algorithm. The evaluation of our setup was performed on three anatomic specimens and one bony skull.
RESULTS: The ultrasound echoes from the occiput allowed the creation of an adequate 3D surface which could be registered to a segmentation of the CT image with an accuracy greater than 1.5 mm. The experiments were evaluated by an intuitive representation of the spatial deviation between CT and ultrasound data as a color-coded map.
CONCLUSION: The approach to scan the posterior skull with A-mode ultrasound enables automatic intraoperative registration and can be integrated into the intraoperative setup.

Entities:  

Mesh:

Year:  2010        PMID: 20878382     DOI: 10.1007/s00106-010-2171-1

Source DB:  PubMed          Journal:  HNO        ISSN: 0017-6192            Impact factor:   1.284


  12 in total

Review 1.  [Computer-aided 3D-navigation systems. Survey and location determination].

Authors:  A R Gunkel; W F Thumfart; W Freysinger
Journal:  HNO       Date:  2000-02       Impact factor: 1.284

2.  Analytic expressions for fiducial and surface target registration error.

Authors:  Burton Ma; Randy E Ellis
Journal:  Med Image Comput Comput Assist Interv       Date:  2006

3.  [A-mode ultrasonic pointer for navigated pelvic surgery].

Authors:  M Oszwald; M Citak; J Kowal; C Amstutz; D Kendoff; T Kirchhoff; L P Nolte; C Krettek; T Hüfner
Journal:  Unfallchirurg       Date:  2008-03       Impact factor: 1.000

4.  [Navigation-assisted sonography for soft tissues in the head and neck region].

Authors:  M Helbig; K Krysztoforski; J Kucharski; M Popek; T Kroll; S Helbig; A May; W Gstoettner; J Kozak
Journal:  HNO       Date:  2009-10       Impact factor: 1.284

5.  Predicting error in rigid-body point-based registration.

Authors:  J M Fitzpatrick; J B West; C R Maurer
Journal:  IEEE Trans Med Imaging       Date:  1998-10       Impact factor: 10.048

6.  An ultrasonic approach to localization of fiducial markers for interactive, image-guided neurosurgery--Part I: Principles.

Authors:  J T Lewis; R L Galloway; S Schreiner
Journal:  IEEE Trans Biomed Eng       Date:  1998-05       Impact factor: 4.538

7.  Laser surface scanning for patient registration in intracranial image-guided surgery.

Authors:  Andreas Raabe; René Krishnan; Robert Wolff; Elvis Hermann; Michael Zimmermann; Volker Seifert
Journal:  Neurosurgery       Date:  2002-04       Impact factor: 4.654

8.  A-mode ultrasound-based registration in computer-aided surgery of the skull.

Authors:  Christoph Amstutz; Marco Caversaccio; Jens Kowal; Richard Bächler; Lutz-Peter Nolte; Rudolf Häusler; Martin Styner
Journal:  Arch Otolaryngol Head Neck Surg       Date:  2003-12

9.  Soft tissue scanning for patient registration in image-guided surgery.

Authors:  Rüdiger Marmulla; Stefan Hassfeld; Tim Lüth; Ulrich Mende; Joachim Mühling
Journal:  Comput Aided Surg       Date:  2003

10.  Image-guided endoscopic ENT surgery.

Authors:  W Freysinger; A R Gunkel; W F Thumfart
Journal:  Eur Arch Otorhinolaryngol       Date:  1997       Impact factor: 3.236

View more
  2 in total

1.  Basic study for ultrasound-based navigation for pedicle screw insertion using transmission and backscattered methods.

Authors:  Ziqiang Chen; Bing Wu; Xiao Zhai; Yushu Bai; Xiaodong Zhu; Beier Luo; Xiao Chen; Chao Li; Mingyuan Yang; Kailiang Xu; Chengcheng Liu; Chuanfeng Wang; Yingchuan Zhao; Xianzhao Wei; Kai Chen; Wu Yang; Dean Ta; Ming Li
Journal:  PLoS One       Date:  2015-04-10       Impact factor: 3.240

2.  Vector field analysis for surface registration in computer-assisted ENT surgery.

Authors:  Georgi Diakov; Wolfgang Freysinger
Journal:  Int J Med Robot       Date:  2019-01-07       Impact factor: 2.547

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.