Failure of the tibial insert in a rotating hinge total knee arthroplasty.

Ran Schwarzkopf1, Sonia Chaudhry, Frederick J Kummer, Scott E Marwin.   

Abstract

There has been a steady increase in the number of revision total knee arthroplasties being performed in the United States. Hinge knee prostheses are used predominantly in complex primary or revision arthroplasties, often as salvage procedures. Significant improvement of the articulation between the rotating hinge mechanism and tibial component has decreased the stresses that contributed to earlier failures in previous generation designs. Two cases of fracture of the tibial metal post in the rotating hinge of a revision total knee arthroplasty were evaluated and treated at our institution within a 1-month period. We present our experience with this construct failure and subsequent patient management.
Copyright © 2011 Elsevier Inc. All rights reserved.

Entities:  

Mesh:

Substances:

Year:  2010        PMID: 20870381     DOI: 10.1016/j.arth.2010.08.008

Source DB:  PubMed          Journal:  J Arthroplasty        ISSN: 0883-5403            Impact factor:   4.757


  8 in total

1.  Case report: Failure of rotating-hinge total knee prosthesis by disengagement of the hinge-post extension.

Authors:  Debdut Biswas; Bryan Haughom; Robert E Mayle; Craig J Della Valle
Journal:  Clin Orthop Relat Res       Date:  2012-12-11       Impact factor: 4.176

Review 2.  Failure rate of a rotating hinge knee design due to yoke fracture of the hinged tibial insert: a retrospective data analysis and review of the literature.

Authors:  Joerg Friesenbichler; Ran Schwarzkopf; Patrick Sadoghi; Scott E Marwin; Mathias Glehr; Werner Maurer-Ertl; Andreas Leithner
Journal:  Int Orthop       Date:  2011-12-28       Impact factor: 3.075

3.  Fracture of the insert cone of a polyethylene liner in a failed posterior-stabilized, rotating-platform total knee arthroplasty.

Authors:  Marc R Angerame; Jason M Jennings; Douglas A Dennis
Journal:  Arthroplast Today       Date:  2017-11-07

Review 4.  Dislocation of rotating-hinge total knee arthroplasty.

Authors:  Carlos A Encinas-Ullán; Primitivo Gómez-Cardero; Juan S Ruiz-Pérez; E Carlos Rodríguez-Merchán
Journal:  EFORT Open Rev       Date:  2021-02-01

5.  Posterior dislocation of the hinge-post extension in a rotating hinge total knee prosthesis.

Authors:  Givenchy Manzano; Ran Schwarzkopf
Journal:  Case Rep Orthop       Date:  2013-09-23

6.  Evaluation of stability of rotating hinge knee prostheses: a biomechanical study.

Authors:  Joerg Friesenbichler; Andreas Leithner; Mathias Glehr; Patrick Sadoghi; Werner Maurer-Ertl; Alexander Avian; Reinhard Windhager
Journal:  ISRN Orthop       Date:  2013-07-18

7.  Biomechanical assessment of a patient-specific knee implant design using finite element method.

Authors:  Qiguo Rong; Jianfeng Bai; Yongling Huang; Jianhao Lin
Journal:  Biomed Res Int       Date:  2014-06-30       Impact factor: 3.411

Review 8.  Total knee arthroplasty using hinge joints: Indications and results.

Authors:  E Carlos Rodríguez-Merchán
Journal:  EFORT Open Rev       Date:  2019-04-25