Literature DB >> 20862403

Influence of implant collar design on stress and strain distribution in the crestal compact bone: a three-dimensional finite element analysis.

Wan-Ling Shen1, Chen-Sheng Chen, Ming-Lun Hsu.   

Abstract

PURPOSE: To evaluate the influence of implant collar geometry on the distribution of stress and strain in the crestal compact bone contiguous to an implant collar for four types of bone under axial and oblique loads.
MATERIALS AND METHODS: Finite element models of threaded implants with three kinds of implant collar designs (divergent, straight, and convergent) with their corresponding suprastructures embedded in the posterior mandible were created with ANSYS software. Eight different test conditions incorporating four types of bone (orthotropic and effectively isotropic in part 1 and high and low densities in part 2) under separate 100-N axial and 35.6-degree oblique forces were created to investigate the stress and strain distributions in the crestal compact bone around the implant collars.
RESULTS: In all eight conditions, the divergent collar demonstrated the lowest maximum von Mises and principal stresses and strains in the crestal compact bone contiguous to the implant collar, followed by the straight and convergent collars. The oblique load induced higher peak values than the axial load. The orthotropic design amplified and increased the pathologic microstrains and tensile stresses in the crestal compact bone compared to the effectively isotropic design, especially in models with a convergent collar design. In part 2 of the study, the maximum von Mises stresses and strains increased with a decrease in the cancellous bone density. Under oblique loading, the convergent and straight collars showed pathologic microstrain values as well as excessive ultimate tensile stresses in the orthotropic bone model with low-density cancellous bone.
CONCLUSIONS: Within the limitations, it was concluded that stress and strain distributions in the adjacent compact bone are influenced by the implant collar design. The divergent implant collar design was associated with the lowest stress and strain concentrations in the crestal compact bone.

Entities:  

Mesh:

Substances:

Year:  2010        PMID: 20862403

Source DB:  PubMed          Journal:  Int J Oral Maxillofac Implants        ISSN: 0882-2786            Impact factor:   2.804


  4 in total

1.  Influence of orthotropy on biomechanics of peri-implant bone in complete mandible model with full dentition.

Authors:  Xi Ding; Sheng-Hui Liao; Xing-Hao Zhu; Hui-Ming Wang
Journal:  Biomed Res Int       Date:  2014-11-03       Impact factor: 3.411

Review 2.  Finite element analysis of dental implants with validation: to what extent can we expect the model to predict biological phenomena? A literature review and proposal for classification of a validation process.

Authors:  Yuanhan Chang; Abhijit Anil Tambe; Yoshinobu Maeda; Masahiro Wada; Tomoya Gonda
Journal:  Int J Implant Dent       Date:  2018-03-08

3.  A three-dimensional finite element analysis of the influence of varying implant crest module designs on the stress distribution to the bone.

Authors:  Shweta Maruti Patil; Abhijit Suresh Deshpande; Rahul Ramesh Bhalerao; Suryakant Bhanudas Metkari; Prithviraj Maruti Patil
Journal:  Dent Res J (Isfahan)       Date:  2019 May-Jun

4.  The effects of bone density and crestal cortical bone thickness on micromotion and peri-implant bone strain distribution in an immediately loaded implant: a nonlinear finite element analysis.

Authors:  Tsutomu Sugiura; Kazuhiko Yamamoto; Satoshi Horita; Kazuhiro Murakami; Sadami Tsutsumi; Tadaaki Kirita
Journal:  J Periodontal Implant Sci       Date:  2016-06-28       Impact factor: 2.614