Literature DB >> 20849796

Preliminary use of the Leo stent in the endovascular treatment of wide-necked cerebral aneurysms.

Junsheng Luo1, Xianli Lv, Chuhan Jiang, Zhongxue Wu.   

Abstract

BACKGROUND: Currently available stents for intracranial use usually are Neuroform (Boston Scientific/Target, Fremont, CA) and Leo (Balt, Montmorency, France) stents. We present the results of our initial experience in using the Leo stent to treat patients with wide-necked cerebral aneurysms.
METHODS: Fifteen consecutive patients with wide-necked intracranial aneurysm were treated with a combined approach that consisted of delivery of a flexible self-expending neurovascular stent through a microcatheter to cover the neck of the aneurysm and subsequent filling of the aneurym with coils through the stent interstices. We assessed the clinical history, aneurysm dimensions, and technical details of the procedures, including any difficulties with stent deployment, degree of aneurysm occlusion, and complications.
RESULTS: During a 1-year period, 15 patients with 17 broad-necked aneurysms (n = 13; average neck length, 5.1 mm; average aneruysm size, 9 mm) were treated with the Leo stent. Fifteen stents were deployed successfully; two failed. Of the coiled aneurysms, complete or subtotal (more than 95%) occlusion was achieved in 11 aneurysms, and partial occlusion was achieved in 2 aneurysms. One patient had multiple stents placed. One patient had a ruptured aneurysm at the time of treatment. Technical problems included difficulty in deploying the stent (n = 2). Two periprocedural thromboembolic complications occurred. One patient had palsy after thrombolysis was attempted. The other patient made an excellent functional recovery after undergoing successful thrombolysis of a thrombosed internal carotid artery stent.
CONCLUSIONS: Preliminary data demonstrated that the Leo stent is useful device for the treatment of patients with wide-necked aneurysms. In cases with tortuous cerebral vasculature, delivery and deployment may be technically challenging. Clinically significant complications are uncommon.
Copyright © 2010 Elsevier Inc. All rights reserved.

Entities:  

Mesh:

Year:  2010        PMID: 20849796     DOI: 10.1016/j.wneu.2010.01.019

Source DB:  PubMed          Journal:  World Neurosurg        ISSN: 1878-8750            Impact factor:   2.104


  3 in total

1.  LEO Baby Stent Use following Balloon-Assisted Coiling: Single- and Dual-Stent Technique--Immediate and Midterm Results of 29 Consecutive Patients.

Authors:  P Machi; V Costalat; K Lobotesis; C Ruiz; Y B Cheikh; O Eker; G Gascou; F Danière; C Riquelme; A Bonafé
Journal:  AJNR Am J Neuroradiol       Date:  2015-08-13       Impact factor: 3.825

2.  Coiling after treatment with the woven EndoBridge cerebral aneurysm embolization device. A case report.

Authors:  A K Wallner; E Broussalis; T Hauser; E Trinka; M Killer-Oberpfalzer
Journal:  Interv Neuroradiol       Date:  2012-06-04       Impact factor: 1.610

3.  Alpha horizontal stent delivery for coil embolization of a broad-necked large basilar apex aneurysm: a case report.

Authors:  Tomotaka Ohshima; Masamune Nagakura; Toshihisa Nishizawa; Kyozo Kato
Journal:  Nagoya J Med Sci       Date:  2015-11       Impact factor: 1.131

  3 in total