Literature DB >> 20838843

When staplers misfire: endoscopic rescue of low pelvic anastomoses.

R A Levine1, O Kadro.   

Abstract

The end-to-end stapler has made it possible for colorectal surgeons to construct deeper anastomoses. Although complications associated with the device are mainly postoperative, very serious intraoperative complications, such as stapler misfire, can occur. The authors report their experience with two cases of stapler misfire, describing their method for the extraction of the device and the entrapped tissue using a flexible sigmoidoscope and a hot biopsy forceps. There were no immediate or long-term problems. The technique appears to be safe and effective.

Entities:  

Mesh:

Year:  2010        PMID: 20838843     DOI: 10.1007/s10151-010-0641-9

Source DB:  PubMed          Journal:  Tech Coloproctol        ISSN: 1123-6337            Impact factor:   3.781


  3 in total

1.  New techniques of gastrointestinal anastomoses with the EEA stapler.

Authors:  F C Nance
Journal:  Ann Surg       Date:  1979-05       Impact factor: 12.969

Review 2.  An unusual intraoperative complication of stapled colorectal anastomosis: report of a case and review of the literature.

Authors:  V Bard; Z Cohen; T Ross
Journal:  Can J Surg       Date:  1993-04       Impact factor: 2.089

Review 3.  Experience with the use of the circular stapler in rectal surgery.

Authors:  S Kyzer; P H Gordon
Journal:  Dis Colon Rectum       Date:  1992-07       Impact factor: 4.585

  3 in total