Comparative study of two generations of the Complior device for aortic pulse wave velocity measurements.

Telmo Pereira1, João Maldonado.   

Abstract

INTRODUCTION: The purpose of this study was to evaluate intraobserver and interobserver reproducibility of carotid-femoral pulse wave velocity (PWV) with the Complior method, by comparing two generations of the device and the manual method.
METHODS: We studied 31 participants (9 women) with a mean age of 47.67±12.55 years, who were submitted to sequential measures of carotid-femoral PWV by two experienced operators. The evaluations were made by randomly switching between the manual method, the Complior first generation, and the Complior SP (third generation).
RESULTS: The analysis of concordance of measurements of PWV showed strong intraobserver and interobserver correlations. The intraobserver correlation coefficients were 0.984, 0.977, and 0.966, respectively, for manual evaluation, the Complior first generation, and the Complior SP (P<0.0001). In the case of interobserver reproducibility, we found correlations of 0.940, 0.979, and 0.974, respectively, for manual evaluation, the Complior first generation, and Complior SP (P<0.0001). Good indicators of reproducibility were also a withdrawal of the Bland-Altman analysis, with most values within 2 standard deviations. The mean difference±SD for intraobserver and interobserver agreement was -0.006±0.05 and -0.06±0.09 for manual determination; -0.02±0.06 and -0.04±0.06 for Complior first generation; and 0.16±0.09 and 0.15±0.07 for the Complior SP.
CONCLUSION: The data showed a high reproducibility of the Complior method, documented by excellent correlations and low mean differences in measurements within and between observers. These data showed unequivocally that this method met the quality requirements for its inclusion in integrated clinical follow-up programs.

Entities:  

Mesh:

Year:  2010        PMID: 20838328     DOI: 10.1097/MBP.0b013e32833f5685

Source DB:  PubMed          Journal:  Blood Press Monit        ISSN: 1359-5237            Impact factor:   1.444


  6 in total

1.  A region-matching method for pulse transit time estimation: potential for improving the accuracy in determining carotid femoral pulse wave velocity.

Authors:  F S Hu; Y L Zhang; Z C Ma; Q Q Cao; Y B Xu; Z J He; Y N Sun
Journal:  J Hum Hypertens       Date:  2015-02-19       Impact factor: 3.012

2.  Circadian hemodynamics in men and women with high blood pressure: dipper vs. nondipper and racial differences.

Authors:  Andrew Sherwood; LaBarron K Hill; James A Blumenthal; Alan L Hinderliter
Journal:  J Hypertens       Date:  2018-02       Impact factor: 4.844

3.  Validation of the new Complior sensor to record pressure signals non-invasively.

Authors:  Benjamin Sztrymf; Frédéric Jacobs; Denis Chemla; Christian Richard; Sandrine C Millasseau
Journal:  J Clin Monit Comput       Date:  2013-05-17       Impact factor: 2.502

4.  Recommendations for Improving and Standardizing Vascular Research on Arterial Stiffness: A Scientific Statement From the American Heart Association.

Authors:  Raymond R Townsend; Ian B Wilkinson; Ernesto L Schiffrin; Alberto P Avolio; Julio A Chirinos; John R Cockcroft; Kevin S Heffernan; Edward G Lakatta; Carmel M McEniery; Gary F Mitchell; Samer S Najjar; Wilmer W Nichols; Elaine M Urbina; Thomas Weber
Journal:  Hypertension       Date:  2015-07-09       Impact factor: 10.190

5.  Ideal Cardiovascular Health and Vascular Phenotype Associations in Mothers with Obesity and Their Six-Year-Old Children.

Authors:  Linda Litwin; Johnny K M Sundholm; Jelena Meinilä; Janne Kulmala; Tuija H Tammelin; Kristiina Rönö; Saila B Koivusalo; Johan G Eriksson; Taisto Sarkola
Journal:  Diabetes Metab Syndr Obes       Date:  2021-07-13       Impact factor: 3.168

6.  Aerobic training-induced improvements in arterial stiffness are not sustained in older adults with multiple cardiovascular risk factors.

Authors:  K M Madden; C Lockhart; D Cuff; T F Potter; G S Meneilly
Journal:  J Hum Hypertens       Date:  2012-09-06       Impact factor: 3.012

  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.