Rapid urine drug screens: diphenhydramine and methadone cross-reactivity.

Steven C Rogers1, Charles W Pruitt, Dennis J Crouch, E Martin Caravati.   

Abstract

BACKGROUND: Rapid urine screens to detect drugs of abuse are often used in pediatric emergency departments (PEDs). A positive result may lead to further clinical testing, social evaluation, and increased stress/inconvenience. A PED patient with suspected diphenhydramine (DPH) ingestion had a positive methadone result on the rapid urine drug screen, One Step Multi-Drug, Multi-Line Screen Test Device (ACON Laboratories, San Diego, Calif). There was no history of methadone exposure so the patient was admitted while confirmatory testing was performed. Gas chromatography/mass spectroscopy testing of the urine failed to confirm the presence of methadone. We present this unreported false-positive methadone result and evaluation of the kit for cross-reactivity of DPH and methadone.
METHODS: The same One Step urine drug screen was tested at an independent laboratory for cross-reactivity between methadone and DPH including the DPH metabolites. Drug-free urine was fortified with DPH, nordiphenhydramine, or dinordiphenhydramine at 0, 10, 25, 50, and 100 μg/mL for each analyte. One hundred microliters of the solutions were added to each of the 4 wells on test cassettes. Urine was allowed to migrate according to manufacturer instructions. Each cassette was interpreted by 2 analysts to ensure consistent interpretation and accurate data recording.
RESULTS: In vitro laboratory testing results showed cross-reactivity between methadone and DPH but not for nordiphenhydramine or dinordiphenhydramine.
CONCLUSIONS: Rapid urine drug screens using immunoassays based on the principle of competitive binding may show false-positive methadone results for patients who have ingested DPH. Product information for urine drug screens may not include all cross-reacting agents and should be used with caution when interpreting drug screen results in PED patients.

Entities:  

Mesh:

Substances:

Year:  2010        PMID: 20838187     DOI: 10.1097/PEC.0b013e3181f05443

Source DB:  PubMed          Journal:  Pediatr Emerg Care        ISSN: 0749-5161            Impact factor:   1.454


  1 in total

1.  Buyer Beware: Pitfalls in Toxicology Laboratory Testing.

Authors:  D Adam Algren; Michael R Christian
Journal:  Mo Med       Date:  2015 May-Jun
  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.