Literature DB >> 20832393

Hemoglobin A1c: assessment of three POC analyzers relative to a central laboratory method.

John R Petersen1, Felix O Omoruyi, Amin A Mohammad, Thomas J Shea, Anthony O Okorodudu, Hyunsu Ju.   

Abstract

BACKGROUND: Glycosylated hemoglobin evaluation is very important for assessing the control of diabetes. Since the use of point-of-care (POC) devices for monitoring HbA1c is increasing, it is important to determine how these devices compare in relation to instrumentation used in the central laboratory (CL).
METHODS: Eighty-eight randomly selected samples previously analyzed using the Bio-Rad Variant™ II Hemoglobin Testing System were run on three POC Analyzers (Siemens DCA Vantage™ Analyzer, Axis-Shield Afinion™ AS100 Analyzer, and Bio-Rad In2it™ Analyzer).
RESULTS: All POC instruments showed good correlation to the CL method (R(2)>0.95 for all methods). HbA1c levels obtained using Variant II (mean=7.9; 95% CI=7.5-8.3%) and In2it (mean=7.9; 95% C.I.=7.5-8.2%) instruments were found to have no statistical mean difference (p=0.21), while the values obtained using DCA Vantage (mean=7.2% C.I.=6.9-7.5%) and Afinion (mean=7.3% C.I.=7.0-7.6%) instruments were different (p<0.001) from those of the CL method. The Afinion and DCA Vantage instruments increasingly underestimated the HbA1c compared to the CL as the HbA1c values increased. These differences were even more striking when the estimated average glucose is calculated.
CONCLUSIONS: Despite significant variation of results among the POC instruments evaluated relative to the CL method and pending resolution of HbA1c standardization issues, we conclude that all of the POC instruments can be used for HbA1c determination if clinicians are given instrument specific reference ranges.
Copyright © 2010 Elsevier B.V. All rights reserved.

Entities:  

Mesh:

Substances:

Year:  2010        PMID: 20832393     DOI: 10.1016/j.cca.2010.09.004

Source DB:  PubMed          Journal:  Clin Chim Acta        ISSN: 0009-8981            Impact factor:   3.786


  8 in total

1.  Accuracy and precision of the Axis-Shield Afinion hemoglobin A1c measurement device.

Authors:  Jamie R Wood; Brett M Kaminski; Craig Kollman; Roy W Beck; Callyn A Hall; Jason P Yun; Eda Cengiz; Michael J Haller; Krishna Hassan; Georgeanna J Klingensmith; William V Tamborlane
Journal:  J Diabetes Sci Technol       Date:  2012-03-01

2.  Hemoglobin A1c assay variations and implications for diabetes screening in obese youth.

Authors:  Christine L Chan; Kim McFann; Lindsey Newnes; Kristen J Nadeau; Philip S Zeitler; Megan Kelsey
Journal:  Pediatr Diabetes       Date:  2014-03-17       Impact factor: 4.866

3.  Long-Term Performance of Point-of-Care Hemoglobin A1c Assays.

Authors:  Sujaytha Paknikar; Rohan Sarmah; Losika Sivaganeshan; Adam Welke; Al Rizzo; Kirk Larson; Marc Rendell
Journal:  J Diabetes Sci Technol       Date:  2016-11-01

4.  Accuracy of verbal self-reported blood glucose in teenagers with type I diabetes at diabetes ski camp.

Authors:  Matthew Chae; David M Reith; Paul A Tomlinson; Jenny Rayns; Benjamin J Wheeler
Journal:  J Diabetes Metab Disord       Date:  2014-01-08

5.  Point-of-care capillary HbA1c measurement in the emergency department: a useful tool to detect unrecognized and uncontrolled diabetes.

Authors:  Fernando Gomez-Peralta; Cristina Abreu; Leonor Andreu-Urioste; Ana Cristina Antolí; Carmen Rico-Fontsaré; David Martín-Fernández; Rosa Resina-Rufes; Juan Jesús Pérez-García; Ángela Negrete-Muñoz; Daniel Muñoz-Álvarez; Guillermo E Umpierrez
Journal:  Int J Emerg Med       Date:  2016-02-19

6.  Study of the effect of altitude on the measurement of glycated haemoglobin using point-of-care instruments.

Authors:  Sandra W Veigne; Eugene Sobngwi; Brice E Nouthe; Joelle Sobngwi-Tambekou; Eric V Balti; Serge Limen; Mesmin Y Dehayem; Vicky Ama; Jean- Louis Nguewa; Maimouna Ndour-Mbaye; Alioune Camara; Naby M Balde; Jean- Claude Mbanya
Journal:  Cardiovasc J Afr       Date:  2015 Jan-Feb       Impact factor: 1.167

7.  Assessment of Two Diabetes Point-of-care Analyzers Measuring Hemoglobin A1c in the Peruvian Amazon.

Authors:  Anthony T Saxton; J Jaime Miranda; Ernesto J Ortiz; William Pan
Journal:  Ann Glob Health       Date:  2018-11-05       Impact factor: 2.462

8.  Haemoglobin A1c: comparing performance of two point of care devices with laboratory analyser.

Authors:  Ruziana Mona Wan Mohd Zin; Zati Iwani Ahmad Kamil; Tuan Rosidah Tuan Soh; Mustaffa Embong; Wan Nazaimoon Wan Mohamud
Journal:  BMC Res Notes       Date:  2013-12-18
  8 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.