Immersion B-guided versus contact A-mode biometry for accurate measurement of axial length and intraocular lens power calculation in siliconized eyes.

Khaled G Abu El Einen1, Mervat H Shalaby, Hoda T El Shiwy.   

Abstract

PURPOSE: To compare the accuracy of the immersion B-guided versus the contact A-mode biometry in measuring the axial length and intraocular lens power of the siliconized eyes.
METHODS: Sixty eyes with silicone cataract were divided into two groups. In Group A, the axial length of 30 eyes was measured by contact A-mode biometry. In Group B, the axial length of 30 eyes was measured by immersion B-guided biometry. The postoperative refraction was done after 3 months and was compared with the attempted values.
RESULTS: Twenty-five eyes of the B-guided group (83.3%) achieved a refraction deviation of ± 1 D or less versus 14 (46.6%) in the A-mode group. The mean refractive deviation was less with the B-mode (0.15 ± 0.94) versus contact A-mode (-0.95 ± 1.157); the difference is statistically significant (P = 0.024, Fisher test).
CONCLUSION: B-guided biometry was found to be more accurate than the contact method in measuring axial length and intraocular lens power calculation in siliconized eyes and spared the time and cost needed for a separate B-scan.

Entities:  

Mesh:

Substances:

Year:  2011        PMID: 20829737     DOI: 10.1097/IAE.0b013e3181e17f39

Source DB:  PubMed          Journal:  Retina        ISSN: 0275-004X            Impact factor:   4.256


  3 in total

1.  Accuracy and reliability of IOL master and A-scan immersion biometry in silicone oil-filled eyes.

Authors:  P Kunavisarut; P Poopattanakul; C Intarated; K Pathanapitoon
Journal:  Eye (Lond)       Date:  2012-08-10       Impact factor: 3.775

2.  Comparison of immersion ultrasound and low coherence reflectometry for ocular biometry in cataract patients.

Authors:  Yan Li; Hong-Xun Li; Yang-Chen Liu; Ya-Tu Guo; Jian-Min Gao; Bin Wu; Nan Zhang; Dong Liu; Xiao-Yong Yuan
Journal:  Int J Ophthalmol       Date:  2018-06-18       Impact factor: 1.779

3.  Comparison of ocular biometry measurements by applanation and immersion A-scan techniques.

Authors:  Dupe S Ademola-Popoola; Donald A Nzeh; Sadiat E Saka; Lateefat B Olokoba; Tokunbo S Obajolowo
Journal:  J Curr Ophthalmol       Date:  2016-02-09