Evaluation of mean glandular dose in a full-field digital mammography unit in Tabriz, Iran.

Hamed Alizadeh Riabi1, Parinaz Mehnati, Asghar Mesbahi.   

Abstract

This study was aimed at evaluating the mean glandular dose (MGD) and affecting factors during mammography examinations by a full-field digital mammography unit. An extensive quality control program was performed to assure that the unit is properly working. Required information including compressed breast thickness (CBT), breast parenchymal pattern and technical factors used for imaging were recorded. An entrance skin exposure measurement was also performed using slabs of polymethylmethacrylate with 2-8 cm thickness. On the basis of recorded information and measured data, the MGD was estimated for 1145 mammography examinations obtained from 298 patients. Mean CBTs of 4.9 and 5.8 cm and MGDs of 2 and 2.4 mGy were observed for craniocaudal and mediolateral oblique views, respectively. Significant correlation was seen between MGD and CBT, breast parenchymal pattern and applied kVp and mAs.

Entities:  

Mesh:

Year:  2010        PMID: 20823039     DOI: 10.1093/rpd/ncq218

Source DB:  PubMed          Journal:  Radiat Prot Dosimetry        ISSN: 0144-8420            Impact factor:   0.972


  3 in total

1.  Evaluation of mean glandular dose and modulation transfer function for different tube potentials and target-filter combinations in computed radiography mammography.

Authors:  Siti Aishah Abdul Aziz; Abdul Khaliq Mohd Saparudin; Ahmad Zaky Harun
Journal:  Malays J Med Sci       Date:  2013-05

2.  Patient investigation of average glandular dose and incident air kerma for digital mammography.

Authors:  Ai Kawaguchi; Yuta Matsunaga; Tomoko Otsuka; Shoichi Suzuki
Journal:  Radiol Phys Technol       Date:  2014-01

3.  Regression Analysis between the Different Breast Dose Quantities Reported in Digital Mammography and Patient Age, Breast Thickness, and Acquisition Parameters.

Authors:  Salam Dhou; Entesar Dalah; Reda AlGhafeer; Aisha Hamidu; Abdulmunhem Obaideen
Journal:  J Imaging       Date:  2022-07-31