| Literature DB >> 20820218 |
Yuuki Watanabe1, Seiya Maeno, Kenji Aoshima, Haruyuki Hasegawa, Hitoshi Koseki.
Abstract
The real-time display of full-range, 2048?axial pixelx1024?lateral pixel, Fourier-domain optical-coherence tomography (FD-OCT) images is demonstrated. The required speed was achieved by using dual graphic processing units (GPUs) with many stream processors to realize highly parallel processing. We used a zero-filling technique, including a forward Fourier transform, a zero padding to increase the axial data-array size to 8192, an inverse-Fourier transform back to the spectral domain, a linear interpolation from wavelength to wavenumber, a lateral Hilbert transform to obtain the complex spectrum, a Fourier transform to obtain the axial profiles, and a log scaling. The data-transfer time of the frame grabber was 15.73?ms, and the processing time, which includes the data transfer between the GPU memory and the host computer, was 14.75?ms, for a total time shorter than the 36.70?ms frame-interval time using a line-scan CCD camera operated at 27.9?kHz. That is, our OCT system achieved a processed-image display rate of 27.23 frames/s.Entities:
Mesh:
Year: 2010 PMID: 20820218 DOI: 10.1364/AO.49.004756
Source DB: PubMed Journal: Appl Opt ISSN: 1559-128X Impact factor: 1.980