Comparison of central corneal thickness measurements using Pentacam and ultrasonic pachymetry in post-LASIK eyes for myopia.

Hani S Al-Mezaine1, Saleh A Al-Amro, Dustan Kangave, Saleh Al-Obeidan, Khaled M Al-Jubair.   

Abstract

PURPOSE: To compare central corneal thickness (CCT) measurements obtained using the Pentacam Scheimpflug system with those obtained using DGH ultrasound pachymetry (UP) in post-laser in situ keratomileusis (LASIK) eyes for myopia.
METHODS: In a prospective study, measurement agreement was assessed in 143 eyes of 72 post-LASIK patients using both the Pentacam and UP at the Eye Consultants Center in Riyadh, Saudi Arabia.
RESULTS: The mean CCT was 522 +/- 42.2 microm with the Oculus Pentacam and 516.2 +/- 40.6 microm with UP. The Bland-Altman plot showed that the mean +/- SD for the differences between the 2 devices was 5.8 +/- 13.6 microm, with 95% confidence interval limits ranging from -20.9 microm to 32.6 microm. A test of statistical significance indicated that the mean differences of 5.8 +/- 13.6 microm differed significantly from zero (p<0.001; Wilcoxon signed-rank test), thus indicating that the Pentacam measurements tended to overestimate CCT compared with UP. Analysis of regression showed a high correlation between the values obtained with both devices (r=0.947, p<0.001).
CONCLUSIONS: In post-LASIK myopic eyes, although a high correlation has been shown between Pentacam and UP measurements, Pentacam tends to overestimate CCT compared to UP. Pentacam probably cannot be used interchangeably with UP in post-LASIK eyes for myopia.

Entities:  

Mesh:

Year:  2010        PMID: 20815096     DOI: 10.1177/112067211002000507

Source DB:  PubMed          Journal:  Eur J Ophthalmol        ISSN: 1120-6721            Impact factor:   2.597


  6 in total

1.  Meta-analysis of Pentacam vs. ultrasound pachymetry in central corneal thickness measurement in normal, post-LASIK or PRK, and keratoconic or keratoconus-suspect eyes.

Authors:  Wenjing Wu; Yan Wang; Lulu Xu
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2013-11-12       Impact factor: 3.117

2.  Central Corneal Thickness in Highly Myopic Eyes: Inter-device Agreement of Ultrasonic Pachymetry, Pentacam and Orbscan II Before and After Photorefractive Keratectomy.

Authors:  Nader Nassiri; Kourosh Sheibani; Sare Safi; Saman Nassiri; Alireza Ziaei; Alireza Ziaee; Farnaz Haji; Shiva Mehravaran; Nariman Nassiri
Journal:  J Ophthalmic Vis Res       Date:  2014-01

3.  Comparison of Central Corneal Thickness Measurements by Pentacam and Ultrasound Pachymetry in Normal Myopic Patients.

Authors:  Sherif Emerah; Ehab ELZakzouk; Mohamed Farag
Journal:  Electron Physician       Date:  2016-12-25

4.  Comparison of a New Optical Biometer That Combines Scheimpflug Imaging With Partial Coherence Interferometry With That of an Optical Biometer Based on Swept-Source Optical Coherence Tomography and Placido-Disk Topography.

Authors:  Shihao Chen; Qiaoyue Zhang; Giacomo Savini; Shuangzhe Zhang; Xiaomin Huang; Jinjin Yu; Yirang Wang; Rui Ning; Jinhai Huang; Ruixue Tu
Journal:  Front Med (Lausanne)       Date:  2022-02-10

5.  Cataract Grading in Pure Senile Cataracts: Pentacam versus LOCS III.

Authors:  Mohammad Mirzaie; Erfan Bahremani; Nazli Taheri; Zhila Khamnian; Banafshe Kharrazi Ghadim
Journal:  J Ophthalmic Vis Res       Date:  2022-08-15

6.  Measurement of central corneal thickness by ultrasonic pachymeter and oculus pentacam in patients with well-controlled glaucoma: hospital-based comparative study.

Authors:  Riyam Faihan Rashid; Qasim K Farhood
Journal:  Clin Ophthalmol       Date:  2016-03-01
  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.