Literature DB >> 20808700

Image quality improvement after implementation of a CT accreditation program.

You Sung Kim1, Seung Eun Jung, Byung Gil Choi, Yu Ri Shin, Seong Su Hwang, Young Mi Ku, Yeon Soo Lim, Jae Mun Lee.   

Abstract

OBJECTIVE: The purpose of this study was to evaluate any improvement in the quality of abdominal CTs after the utilization of the nationally based accreditation program.
MATERIALS AND METHODS: Approval was obtained from the Institutional Review Board, and informed consent was waived. We retrospectively analyzed 1,011 outside abdominal CTs, from 2003 to 2007. We evaluated images using a fill-up sheet form of the national accreditation program, and subjectively by grading for the overall CT image quality. CT scans were divided into two categories according to time periods; before and after the implementation of the accreditation program. We compared CT scans between two periods according to parameters pertaining to the evaluation of images. We determined whether there was a correlation between the results of a subjective assessment of the image quality and the evaluation scores of the clinical image.
RESULTS: The following parameters were significantly different after the implementation of the accreditation program: identifying data, display parameters, scan length, spatial and contrast resolution, window width and level, optimal contrast enhancement, slice thickness, and total score. The remaining parameters were not significantly different between scans obtained from the two different periods: scan parameters, film quality, and artifacts.
CONCLUSION: After performing the CT accreditation program, the quality of the outside abdominal CTs show marked improvement, especially for the parameters related to the scanning protocol.

Entities:  

Keywords:  Computed tomography (CT); Health policy and practice; QA/QC

Mesh:

Year:  2010        PMID: 20808700      PMCID: PMC2930165          DOI: 10.3348/kjr.2010.11.5.553

Source DB:  PubMed          Journal:  Korean J Radiol        ISSN: 1229-6929            Impact factor:   3.500


  2 in total

1.  Review of failed CT phantom image evaluations in 2005 and 2006 by the CT accreditation program of the Korean Institute for Accreditation of Medical Image.

Authors:  Hye Jung Park; Seung Eun Jung; Young Joon Lee; Woo Il Cho; Kyung Hyun Do; Seung Hoon Kim; Dong Gyu Na
Journal:  Korean J Radiol       Date:  2008 Jul-Aug       Impact factor: 3.500

2.  The relationship between subjective and objective parameters in CT phantom image evaluation.

Authors:  Hye Jung Park; Seung Eun Jung; Young Joon Lee; Woo Il Cho; Kyung Hyun Do; Seung Hyup Kim; Ki Hwang Kim
Journal:  Korean J Radiol       Date:  2009-08-25       Impact factor: 3.500

  2 in total
  1 in total

1.  Intra- and interobserver reliability of gray scale/dynamic range evaluation of ultrasonography using a standardized phantom.

Authors:  Song Lee; Joon-Il Choi; Michael Yong Park; Dong Myung Yeo; Jae Young Byun; Seung Eun Jung; Sung Eun Rha; Soon Nam Oh; Young Joon Lee
Journal:  Ultrasonography       Date:  2014-02-26
  1 in total